Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™ 500) during general anaesthesia.

R Hahn1, H Rinösl, M Neuner, S C Kettner.   

Abstract

BACKGROUND: We conducted a prospective study to test the validity of a new continuous non-invasive blood pressure (NIBP) monitor (CNAP) (CNAP™ Monitor 500).
METHODS: One hundred patients undergoing elective surgery under general anaesthesia were included in the study after informed written consent. The CNAP finger cuffs were placed on the fingers of one arm, an arterial catheter was inserted into the same arm and data were recorded simultaneously. Agreement between invasive arterial pressure (IAP) and blood pressure obtained by CNAP was compared using the Bland-Altman method for repeated measurements. The data from the first 50 patients (software V3.0) were used to improve the software of the CNAP (software V3.5), which was then evaluated in another 50 patients. We defined a clinically acceptable agreement according to the standards of the American Association for the Advancement of Medical Instrumentation for NIBP measurements [limits of agreement (LOA) ± 15 mm Hg].
RESULTS: We analysed 524 878 paired measurements in 100 patients. The mean bias of the mean arterial pressure in the first 50 patients was -2.9 mm Hg (sd 10.6 mm Hg, LOA -23.7 to 17.9 mm Hg), and in the consecutive 50 patients (using software V3.5) the bias was -3.1 mm Hg (sd 9.5 mm Hg, LOA -21.6 to 15.4 mm Hg).
CONCLUSIONS: The new CNAP monitor showed an agreement with the IAP that is promising but did not match our predefined criteria.

Entities:  

Mesh:

Year:  2012        PMID: 22307242     DOI: 10.1093/bja/aer499

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  17 in total

1.  The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients.

Authors:  Bernd Saugel; Florian Fassio; Alexander Hapfelmeier; Agnes S Meidert; Roland M Schmid; Wolfgang Huber
Journal:  Intensive Care Med       Date:  2012-06-29       Impact factor: 17.440

2.  Continuous noninvasive arterial pressure measurement using the volume clamp method: an evaluation of the CNAP device in intensive care unit patients.

Authors:  Julia Y Wagner; Ileana Negulescu; Miriam Schöfthaler; Alexander Hapfelmeier; Agnes S Meidert; Wolfgang Huber; Roland M Schmid; Bernd Saugel
Journal:  J Clin Monit Comput       Date:  2015-03-01       Impact factor: 2.502

3.  [Principles and pitfalls of arterial blood pressure measurement].

Authors:  A S Meidert; J Briegel; B Saugel
Journal:  Anaesthesist       Date:  2019-09       Impact factor: 1.041

Review 4.  Non-invasive continuous blood pressure monitoring: a review of current applications.

Authors:  Elena Chung; Guo Chen; Brenton Alexander; Maxime Cannesson
Journal:  Front Med       Date:  2013-01-23       Impact factor: 4.592

5.  Continuous non-invasive monitoring improves blood pressure stability in upright position: randomized controlled trial.

Authors:  Jan Benes; Alena Simanova; Tereza Tovarnicka; Silvie Sevcikova; Jakub Kletecka; Jan Zatloukal; Richard Pradl; Ivan Chytra; Eduard Kasal
Journal:  J Clin Monit Comput       Date:  2014-05-20       Impact factor: 2.502

6.  Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure measurements during bariatric surgery in severely obese adolescents and young adults.

Authors:  Joseph D Tobias; Chris McKee; Dan Herz; Steve Teich; Paul Sohner; Julie Rice; N'Diris Barry; Marc Michalsky
Journal:  J Anesth       Date:  2014-05-01       Impact factor: 2.078

7.  Significant postictal hypotension: expanding the spectrum of seizure-induced autonomic dysregulation.

Authors:  Alireza Bozorgi; Stephanie Chung; Farhad Kaffashi; Kenneth A Loparo; Satya Sahoo; G Q Zhang; Kitti Kaiboriboon; Samden D Lhatoo
Journal:  Epilepsia       Date:  2013-06-12       Impact factor: 5.864

8.  Automated, continuous and non-invasive assessment of pulse pressure variations using CNAP® system.

Authors:  Matthieu Biais; Laurent Stecken; Aurélie Martin; Stéphanie Roullet; Alice Quinart; François Sztark
Journal:  J Clin Monit Comput       Date:  2016-06-16       Impact factor: 2.502

9.  Blood pressure and heart rate variability during yoga-based alternate nostril breathing practice and breath awareness.

Authors:  Shirley Telles; Sachin Kumar Sharma; Acharya Balkrishna
Journal:  Med Sci Monit Basic Res       Date:  2014-11-19

10.  Noninvasive continuous versus intermittent arterial pressure monitoring: evaluation of the vascular unloading technique (CNAP device) in the emergency department.

Authors:  Julia Y Wagner; Julia S Prantner; Agnes S Meidert; Alexander Hapfelmeier; Roland M Schmid; Bernd Saugel
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2014-01-29       Impact factor: 2.953

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.