Literature DB >> 22299984

A torsion balance for impulse and thrust measurements of micro-Newton thrusters.

Yuan-Xia Yang1, Liang-Cheng Tu, Shan-Qing Yang, Jun Luo.   

Abstract

This paper reports the performance of a torsion-type thrust stand suitable for studies of micro-Newton thrusters, which is developed for ground testing the micro-Newton thruster in Chinese Test of the Equivalence Principle with Optical readout space mission. By virtue of specially suspending design and precise assembly of torsion balance configuration, the thrust stand with load capacity up to several kilograms is able to measure the impulse bit up to 1350 μNs with a resolution of 0.47 μNs, and the average thrust up to 264 μN with a resolution of 0.09 μN in both open and close loop operation. A pulsed plasma thruster, the preliminary prototype developed for Chinese TEPO space mission, is tested by the thrust stand, and the results reveal that the average impulse bit per pulse is measured to be 58.4 μNs with a repeatability of about 5%.

Year:  2012        PMID: 22299984     DOI: 10.1063/1.3675576

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Recommended Practice for Thrust Measurement in Electric Propulsion Testing.

Authors:  James E Polk; Anthony Pancotti; Thomas Haag; Scott King; Mitchell Walker; Joseph Blakely; John Ziemer
Journal:  J Propuls Power       Date:  2017-04-24       Impact factor: 1.500

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.