Literature DB >> 22284333

The application of a computer-assisted thermoplastic membrane navigation system in screw fixation of the sacroiliac joint--a clinical study.

Zhanle Zheng1, Yingze Zhang, Zhiyong Hou, Jiandong Hao, Fushan Zhai, Yanling Su, Jinshe Pan.   

Abstract

BACKGROUND: We inserted iliosacral screws with the aid of a computer-assisted thermoplastic membrane navigation (CATMN) system which is widely used for the accurate and repeatable location tumour in radiation therapy. We hypothesised that application of the CATMN system on IS screws' insertion will provide a superior result to conventional fluoroscopic imaging with less operative time, more accuracy and lower complication rates.
METHODS: We prospectively evaluated 26 consecutive patients who suffered from sacroiliac joint fractures and dislocations (type C, Tile classification) from April 2007 to June 2010 in our hospital. Patients were randomised into two groups: 13 patients in control group and 13 patients in CATMN groups. After operation, inlet and outlet X-ray views and computed tomography (CT) scanning were performed to confirm and compare the screw positions. The operative time, blood loss and accuracy (measured with postoperative CT) were analysed between groups.
RESULTS: In the control group, 18 screws were placed in 13 patients with conventional fluoroscopic technique; two of 18 (11.1%) screws were misplaced. The average intra-operative blood loss was 145.4±112.0ml, and operation time was 619.2±199.5s. In the CATMN group, 21 screws were placed in 13 patients with the application of the CATMN system. All 21 screws were in safe zones. The average intra-operative blood loss was 46.2±24.3ml and the operation time was 353.8±111.2s. Operative time and blood loss were reduced significantly with the CATMN system (p<0.05).
CONCLUSION: Application of CATMN system has high accuracy in treating sacroiliac joint dislocations and provides a new alternative method for guidance of the IS screw placement.
Copyright © 2011 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22284333     DOI: 10.1016/j.injury.2011.12.022

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

1.  A novel patient-specific three-dimensional-printed external template to guide iliosacral screw insertion: a retrospective study.

Authors:  Fan Yang; Sheng Yao; Kai-Fang Chen; Feng-Zhao Zhu; Ze-Kang Xiong; Yan-Hui Ji; Ting-Fang Sun; Xiao-Dong Guo
Journal:  BMC Musculoskelet Disord       Date:  2018-11-13       Impact factor: 2.362

2.  Preoperative planning and safe intraoperative placement of iliosacral screws under fluoroscopic control.

Authors:  Dietmar Krappinger; Richard A Lindtner; Stefan Benedikt
Journal:  Oper Orthop Traumatol       Date:  2019-06-03       Impact factor: 1.154

3.  A novel rabbit fixator made of a thermoplastic mask for awake imaging experiments.

Authors:  Rencai Lu; Li Hou; Siyu Wang; Bo She; Hong He; Wentao Gao; Sidang Wang; Dongdong Xv; Yunhai Ji; Shasha Yang; Zhaohui Yang; Shaobo Wang
Journal:  Sci Rep       Date:  2021-01-15       Impact factor: 4.379

4.  An experimental study on the safe placement of sacroiliac screws using a 3D printing navigation module.

Authors:  Xuanhuang Chen; Feng Zheng; Guodong Zhang; Xiaoqiang Gao; Ya Wang; Wenhua Huang; Haibin Lin
Journal:  Ann Transl Med       Date:  2020-11
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.