A comparative in vitro evaluation of two different magnetic devices detecting the stability of osseo-integrated implants.

O Geckili1, H Bilhan, A Cilingir, E Mumcu, C Bural.   

Abstract

BACKGROUND AND
OBJECTIVE: It is unknown whether the resonance frequency analysis (RFA) measurements made by two different magnetic resonance frequency analysers are comparable. This in vitro study was designed to compare the RFA measurements made by the two magnetic resonance frequency analysers and to evaluate the intra- and interobserver reliability of the magnetic devices.
MATERIAL AND METHODS: Thirty-two implants were placed in four cow ribs. The RFA value of each implant was measured by five different examiners. The measurements were repeated five times, in both the buccal and mesial directions, for each implant at 2 h intervals, and the averages of registered implant stability quotient (ISQ) units were recorded as the buccal ISQ value and the mesial ISQ value for every implant.
RESULTS: No statistically significant differences (p>0.05) were observed between the RFA measurements made by the two magnetic devices. The intra-observer reliability of both devices was excellent, whereas the interobserver reliability of the devices was poor.
CONCLUSION: The results of the RFA measurements of both tested devices overlap. Although both devices show excellent intra-observer reliability, there are variations between the measurements of different examiners.
© 2012 John Wiley & Sons A/S.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22251236     DOI: 10.1111/j.1600-0765.2011.01462.x

Source DB:  PubMed          Journal:  J Periodontal Res        ISSN: 0022-3484            Impact factor:   4.419


  7 in total

1.  Biomechanical effects of the implant material and implant-abutment interface in immediately loaded small-diameter implants.

Authors:  Aaron Yu-Jen Wu; Heng-Li Huang; Jui-Ting Hsu; Winston Chee
Journal:  Clin Oral Investig       Date:  2013-09-03       Impact factor: 3.573

2.  Stability, Survival, and Tolerability of an Auditory Osseointegrated Implant for Bone Conduction Hearing: Long-Term Follow-Up of a Randomized Controlled Trial.

Authors:  Christine A den Besten; Joacim Stalfors; Stina Wigren; Johan Ivarsson Blechert; Mark Flynn; Måns Eeg-Olofsson; Rohini Aggarwal; Kevin Green; Rik C Nelissen; Emmanuel A M Mylanus; Myrthe K S Hol
Journal:  Otol Neurotol       Date:  2016-09       Impact factor: 2.311

Review 3.  Clinical evaluation of osseointegration using resonance frequency analysis.

Authors:  Parth Satwalekar; Sandeep Nalla; Ramaswamy Reddy; Sheeba Glory Chowdary
Journal:  J Indian Prosthodont Soc       Date:  2015 Jul-Sep

4.  In vitro comparison of primary stability of two implant designs in D3 bone.

Authors:  J González-Serrano; R Ortega-Aranegui; J López-Quiles
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2017-07-01

5.  Repeatability of the resonance frequency analysis values in implants with a new technology.

Authors:  M-C Díaz-Castro; A Falcao; P López-Jarana; C Falcao; J-V Ríos-Santos; A Fernández-Palacín; M Herrero-Climent
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2019-09-01

6.  Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study.

Authors:  Jungwon Lee; Se-Wook Pyo; Hyun-Jae Cho; Jung-Sub An; Jae-Hyun Lee; Ki-Tae Koo; Yong-Moo Lee
Journal:  J Periodontal Implant Sci       Date:  2020-02-18       Impact factor: 2.614

7.  An In Vitro Evaluation of Primary Stability Values for Two Differently Designed Implants to Suit Immediate Loading in Very Soft Bone.

Authors:  Dirk Krischik; Selen Ergin Tokgöz; Andreas van Orten; Anton Friedmann; Hakan Bilhan
Journal:  Dent J (Basel)       Date:  2021-01-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.