Water-silicone separated volumetric MR acquisition for rapid assessment of breast implants.

Ananth J Madhuranthakam1, Martin P Smith, Huanzhou Yu, Ann Shimakawa, Scott B Reeder, Neil M Rofsky, Charles A McKenzie, Jean H Brittain.   

Abstract

PURPOSE: To develop a robust T(2) -weighted volumetric imaging technique with uniform water-silicone separation and simultaneous fat suppression for rapid assessment of breast implants in a single acquisition.
MATERIALS AND METHODS: A three-dimensional (3D) fast spin echo sequence that uses variable refocusing flip angles was combined with a three-point chemical-shift technique (IDEAL) and short tau inversion recovery (STIR). Phase shifts of -π/6, +π/2, and +7π/6 between water and silicone were used for IDEAL processing. For comparison, two-dimensional images using 2D-FSE-IDEAL with STIR were also acquired in axial, coronal, and sagittal orientations.
RESULTS: Near-isotropic (true spatial resolution-0.9 × 1.3 × 2.0 mm(3) ) volumetric breast images with uniform water-silicone separation and simultaneous fat suppression were acquired successfully in clinically feasible scan times (7:00-10:00 min). The 2D images were acquired with the same in-plane resolution (0.9 × 1.3 mm(2) ), but the slice thickness was increased to 6 mm with a slice gap of 1 mm for complete coverage of the implants in a reasonable scan time, which varied between 18:00 and 22:30 min.
CONCLUSION: The single volumetric acquisition with uniform water and silicone separation enables images to be reformatted into any orientation. This allows comprehensive assessment of breast implant integrity in less than 10 min of total examination time.
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22247050     DOI: 10.1002/jmri.22872

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  3 in total

1.  Dual-energy CT for the evaluation of silicone breast implants.

Authors:  Thorsten R C Johnson; Isabelle Himsl; Karin Hellerhoff; Doris Mayr; Dorothea Rjosk-Dendorfer; Nina Ditsch; Bernhard Krauss; Klaus Friese; Maximilian F Reiser; Miriam S Lenhard
Journal:  Eur Radiol       Date:  2012-10-13       Impact factor: 5.315

2.  Subject-specific models of susceptibility-induced B0 field variations in breast MRI.

Authors:  Caroline D Jordan; Bruce L Daniel; Kevin M Koch; Huanzhou Yu; Steve Conolly; Brian A Hargreaves
Journal:  J Magn Reson Imaging       Date:  2012-08-03       Impact factor: 4.813

3.  Dual-Energy CT for Evaluation of Intra- and Extracapsular Silicone Implant Rupture.

Authors:  Katrina N Glazebrook; Shuai Leng; Steven R Jacobson; Cynthia M McCollough
Journal:  Case Rep Radiol       Date:  2016-01-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.