Experience with more than 100 total artificial heart implants.

Jack G Copeland1, Hannah Copeland, Monica Gustafson, Nicole Mineburg, Diane Covington, Richard G Smith, Mark Friedman.   

Abstract

OBJECTIVE: The SynCardia Total Artificial Heart (SynCardia Systems Inc, Tucson, Ariz) has been used as a bridge to cardiac transplantation in 930 patients worldwide and in 101 patients in our program. Our experience with SynCardia Total Artificial Heart implantation documents its indications, safety, and efficacy.
METHODS: Data regarding preoperative condition, mortality, and morbidity have been reviewed and analyzed.
RESULTS: From January 1993 to December 2009, 101 patients had bridge to transplant procedures with the SynCardia Total Artificial Heart. Ninety-one percent of cases were Interagency Registry for Mechanically Assisted Circulatory Support profile 1, and the remaining 9% of cases were failing medical therapy on multiple inotropic medications. The mean support time was 87 days (median, 53 days; range, 1-441 days). Pump outputs during support were 7 to 9 L/min. Adverse events included strokes in 7.9% of cases and take-back for hemorrhage in 24.7% of cases. Survival to transplantation was 68.3%. Causes of death of 32 patients on device support included multiple organ failure (13), pulmonary failure (6), and neurologic injury (4). Survival after transplantation at 1, 5, and 10 years was 76.8%, 60.5%, and 41.2%, respectively. The longest-term survivor is currently alive 16.4 years postimplantation.
CONCLUSIONS: These patients were not candidates for left ventricular assist device therapy and were expected to die. The SynCardia Total Artificial Heart offers a real alternative for survival with a reasonable complication rate in appropriate candidates who otherwise might have been assigned to hospice care. Copyright Â
© 2012 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22245242     DOI: 10.1016/j.jtcvs.2011.12.002

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  27 in total

Review 1.  Adult and pediatric mechanical circulation: a guide for the hematologist.

Authors:  Lisa Baumann Kreuziger; M Patricia Massicotte
Journal:  Hematology Am Soc Hematol Educ Program       Date:  2018-11-30

Review 2.  SynCardia Total Artificial Heart: update and future.

Authors:  Jack G Copeland
Journal:  Tex Heart Inst J       Date:  2013

3.  Chugging to silent machines: development of mechanical cardiac support.

Authors:  Cumaraswamy Sivathasan
Journal:  Indian J Thorac Cardiovasc Surg       Date:  2020-08-25

4.  Thromboresistance comparison of the HeartMate II ventricular assist device with the device thrombogenicity emulation- optimized HeartAssist 5 VAD.

Authors:  Wei-Che Chiu; Gaurav Girdhar; Michalis Xenos; Yared Alemu; Jõao S Soares; Shmuel Einav; Marvin Slepian; Danny Bluestein
Journal:  J Biomech Eng       Date:  2014-02       Impact factor: 2.097

5.  Role of percutaneous veno-arterial extracorporeal membrane oxygenation as bridge to left ventricular assist device.

Authors:  Koichi Toda; Tomoyuki Fujita; Osamu Seguchi; Masanobu Yanase; Takeshi Nakatani
Journal:  J Artif Organs       Date:  2017-09-04       Impact factor: 1.731

Review 6.  How New Support Devices Change Critical Care Delivery.

Authors:  Asma Zainab; Divina Tuazon; Faisal Uddin; Iqbal Ratnani
Journal:  Methodist Debakey Cardiovasc J       Date:  2018 Apr-Jun

7.  Use of pulmonary arteriovenous extracorporeal membrane oxygenation in conjunction with the total artificial heart.

Authors:  L G Behrens; N L Goodale; J W Turek; M J Bates
Journal:  Perfusion       Date:  2015-04-24       Impact factor: 1.972

8.  Total Artificial Heart as Bridge to Transplantation for Severe Culture-Negative Prosthetic Valve Endocarditis Due to Gemella haemolysans.

Authors:  Meena S Ramchandani; Robert M Rakita; Rosario V Freeman; Wayne C Levy; Peter Von Homeyer; Nahush A Mokadam
Journal:  ASAIO J       Date:  2014 Jul-Aug       Impact factor: 2.872

Review 9.  The total artificial heart.

Authors:  Jason A Cook; Keyur B Shah; Mohammed A Quader; Richard H Cooke; Vigneshwar Kasirajan; Kris K Rao; Melissa C Smallfield; Inna Tchoukina; Daniel G Tang
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

Review 10.  An overview of mechanical circulatory support in single-ventricle patients.

Authors:  Jacob R Miller; Timothy S Lancaster; Connor Callahan; Aaron M Abarbanell; Pirooz Eghtesady
Journal:  Transl Pediatr       Date:  2018-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.