| Literature DB >> 22231213 |
Nathan Whitmore1, Adrian Nalichowski, Jay Burmeister.
Abstract
Delivery quality assurance (DQA) of tomotherapy plans is routinely performed with silver halide film which has a limited range due to the effects of saturation. DQA plans with dose values exceeding this limit require the dose of the entire plan to be scaled downward if film is used, to evaluate the dose distribution in two dimensions. The potential loss of fidelity between scaled and unscaled DQA plans as a function of dose scaling is investigated. Three treatment plans for 12 Gy fractions designed for SBRT of the lung were used to create DQA procedures that were scaled between 100% and 10%. The dose was measured with an ionization chamber array and compared to values from the tomotherapy treatment planning system. Film and cylindrical ion chamber measurements were also made for one patient for scaling factors of 50% to 10% to compare with the ionization chamber array measurements. The array results show the average gamma pass rate is ≥ 99% from 100% to 30% scaling. The average gamma pass rate falls to 93.6% and 51.1% at 20% and 10% scaling, respectively. Film analysis yields similar pass rates. Cylindrical ion chambers did not exhibit significant variation with dose scaling, but only represent points in the low gradient region of the dose distribution. Scaling the dose changes the mechanics of the radiation delivery, as well as the signal-to-noise ratio. Treatment plans which exhibit parameters that differ significantly from those common to DQA plans studied in this paper may exhibit different behavior. Dose scaling should be limited to the smallest degree possible. Planar information, such as that from film or a detector array, is required. The results show that it is not necessary to perform both a scaled and unscaled DQA plan for the treatment plans considered here.Entities:
Mesh:
Year: 2012 PMID: 22231213 PMCID: PMC5716129 DOI: 10.1120/jacmp.v13i1.3621
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Percent gamma pass rates for the seven29 data for numbered patient (Pt) plans.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| 100 | 100 | 99.6 | 98.9 | 99.5 | 0.560 |
| 80 | 100 | ||||
| 60 | 100 | ||||
| 50 | 100 | 99.6 | 100 | 99.9 | 0.231 |
| 40 | 100 | 100 | 100 | 100.0 | 0.000 |
| 30 | 98.3 | 100 | 99.5 | 99.3 | 0.880 |
| 20 | 90.3 | 100 | 90.4 | 93.6 | 5.95 |
| 10 | 46.9 | 51.4 | 55 | 51.1 | 7.94 |
The Standard Deviation (SD) is given as a percentage of the mean.
Quality factor for the seven29 high‐dose, low‐gradient region for numbered patient (Pt) plans.
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|
| 100 | 0.997 | 0.999 | 0.447 | 1.01 | 0.998 | 1.04 | 0.984 | 0.993 | 0.749 |
| 50 | 0.970 | 0.972 | 0.234 | 0.996 | 0.991 | 0.768 | 0.981 | 0.989 | 0.867 |
| 40 | 0.972 | 0.974 | 0.281 | 0.994 | 0.992 | 0.825 | 0.983 | 0.989 | 0.891 |
| 30 | 0.975 | 0.975 | 0.387 | 0.992 | 0.991 | 0.881 | 0.986 | 0.990 | 0.950 |
| 20 | 0.978 | 0.976 | 0.620 | 0.996 | 0.994 | 0.792 | 0.990 | 0.990 | 1.16 |
| 10 | 0.953 | 0.950 | 1.45 | 1.01 | 1.01 | 2.73 | 0.961 | 0.973 | 2.19 |
The Standard Deviation (SD) is given as a percentage of the mean.
Quality factor mean for all patients combined for the seven29 data.
|
|
|
|
|---|---|---|
| 100 | 0.997 | 0.829 |
| 50 | 0.985 | 1.09 |
| 40 | 0.985 | 1.08 |
| 30 | 0.986 | 1.06 |
| 20 | 0.987 | 1.17 |
| 10 | 0.978 | 3.22 |
The data is for ion chambers in the high‐dose, low‐gradient region. The standard deviation is given as a percentage of the mean.
A1SL ion chamber quality factors.
|
|
|
|
|---|---|---|
| 50 | 1.01 | 1.00 |
| 40 | 1.00 | 0.994 |
| 30 | 1.01 | 0.991 |
| 20 | 1.02 | 0.995 |
| 10 | 1.01 | 1.02 |
Film gamma pass rates.
|
|
|
|
|---|---|---|
| 50% | 97% | NA |
| 40% | 99% | 100% |
| 30% | 99% | 100% |
| 20% | 93% | 98.85% |
| 10% | 42% | 29.71% |
Figure 1Images from the High•Art TPS software: the panels compare the 10% (A, C) and 50% (B, D) scaled DQA plan vertical profiles and gamma maps for the film analysis for Patient 1. Red and yellow on the gamma maps show points failing gamma analysis ().
Figure 2Images from VeriSoft software: the panels compare the 10% (A, C) and 50% (B, D) scaled DQA plan vertical profiles and gamma maps for seven29 analysis for Patient 1. Red and blue on the gamma maps show ion chambers that fail gamma analysis ().
Figure 4Images from VeriSoft software: the panels compare the 10% (A, C) and 50% (B, D) scaled DQA plan vertical profiles and gamma maps for seven29 analysis for Patient 3. Red and blue on the gamma maps show ion chambers that fail gamma analysis ().