Improved keratoconus detection by ocular response analyzer testing after consideration of corneal thickness as a confounding factor.

Jeremías Gastón Galletti1, Tomás Pförtner, Fernando Fuentes Bonthoux.   

Abstract

PURPOSE: To compare corneal hysteresis (CH) and corneal resistance factor (CRF) between normal eyes and eyes with keratoconus correcting for the effect of central corneal thickness (CCT) and to estimate keratoconus detection sensitivity and specificity of these parameters.
METHODS: Observational case series of 102 normal eyes (control group) and 77 eyes with keratoconus (keratoconus group). Examination included corneal topography, tomography, and biomechanical testing with the Ocular Response Analyzer (Reichert Technologies). The confounding effect of CCT was controlled by stratification (20-μm CCT intervals) and linear transformation. Receiver operating characteristic curves were used to identify optimal CH and CRF cutoff points for keratoconus detection. Main outcome measures were CCT, CH, CRF, and diagnostic performance of CH and CRF.
RESULTS: Corneal hysteresis and CRF were positively correlated to CCT in both groups. In the control versus keratoconus group, CH was 9.79 ± 1.51 vs 8.49 ± 1.48 (P<.0001) and CRF was 9.55 ± 1.64 vs 7.24 ± 1.43 (P<.0001), respectively. Only CRF remained significantly lower in eyes with keratoconus after CCT stratification in 20-μm intervals, and keratoconus grade influenced CH and CRF. Transformed CH and CRF data confirmed these results. Sensitivity and specificity of CRF cutoff points ranged between 83% and 94% and 69% and 83%, respectively. True positive rate for CRF in eyes with keratoconus with normal topography was 84%.
CONCLUSIONS: Corneal resistance factor was better than CH for detecting keratoconic corneas once the effect of CCT on ORA measurements was considered, even for topographically unaffected fellow eyes of patients with keratoconus. The CCT-corrected CRF cutoff values and transformed indices may be of clinical use. Copyright 2012, SLACK Incorporated.

Entities:  

Mesh:

Year:  2012        PMID: 22230059     DOI: 10.3928/1081597X-20120103-03

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  18 in total

1.  Alterations in biomechanical properties of the cornea among patients with polycystic kidney disease.

Authors:  Kubra Serefoglu Cabuk; Bennur Esen; Kursat Atalay; Ahmet Kirgiz; Rukiye Aydin
Journal:  Int Ophthalmol       Date:  2017-06-29       Impact factor: 2.031

2.  Evaluation of Intraocular Pressure and Other Biomechanical Parameters to Distinguish between Subclinical Keratoconus and Healthy Corneas.

Authors:  Cristina Peris-Martínez; María Amparo Díez-Ajenjo; María Carmen García-Domene; María Dolores Pinazo-Durán; María José Luque-Cobija; María Ángeles Del Buey-Sayas; Susana Ortí-Navarro
Journal:  J Clin Med       Date:  2021-04-28       Impact factor: 4.241

3.  Multivariate Analysis of the Ocular Response Analyzer's Corneal Deformation Response Curve for Early Keratoconus Detection.

Authors:  Jonatán D Galletti; Pablo R Ruiseñor Vázquez; Fernando Fuentes Bonthoux; Tomás Pförtner; Jeremías G Galletti
Journal:  J Ophthalmol       Date:  2015-05-14       Impact factor: 1.909

4.  Detection of the early keratoconus based on corneal biomechanical properties in the refractive surgery candidates.

Authors:  Zofia Pniakowska; Piotr Jurowski
Journal:  Indian J Ophthalmol       Date:  2016-02       Impact factor: 1.848

5.  Ocular response analyzer parameters in healthy, keratoconus suspect and manifest keratoconus eyes.

Authors:  Mehrdad Mohammadpour; Ifa Etesami; Zahra Yavari; Mohammad Naderan; Fatemeh Abdollahinia; Mahmoud Jabbarvand
Journal:  Oman J Ophthalmol       Date:  2015 May-Aug

6.  Keratoconus diagnosis using Corvis ST measured biomechanical parameters.

Authors:  Roghiyeh Elham; Ebrahim Jafarzadehpur; Hassan Hashemi; Kazem Amanzadeh; Fereshteh Shokrollahzadeh; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  J Curr Ophthalmol       Date:  2017-05-22

7.  Corneal ectasia after myopic laser in situ keratomileusis: a long-term study.

Authors:  Leopoldo Spadea; Emilia Cantera; Magdalena Cortes; Nicole Evangelista Conocchia; Charles Wm Stewart
Journal:  Clin Ophthalmol       Date:  2012-11-02

8.  Corneal biomechanics in iatrogenic ectasia and keratoconus: A review of the literature.

Authors:  Majid Moshirfar; Jason N Edmonds; Nicholas L Behunin; Steven M Christiansen
Journal:  Oman J Ophthalmol       Date:  2013-01

9.  Distribution of Young's modulus in porcine corneas after riboflavin/UVA-induced collagen cross-linking as measured by atomic force microscopy.

Authors:  Jan Seifert; Christian M Hammer; Johannes Rheinlaender; Saadettin Sel; Michael Scholz; Friedrich Paulsen; Tilman E Schäffer
Journal:  PLoS One       Date:  2014-01-31       Impact factor: 3.240

Review 10.  Corneal biomechanical properties in different ocular conditions and new measurement techniques.

Authors:  Nery Garcia-Porta; Paulo Fernandes; Antonio Queiros; Jose Salgado-Borges; Manuel Parafita-Mato; Jose Manuel González-Méijome
Journal:  ISRN Ophthalmol       Date:  2014-03-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.