Oscillometric sphygmomanometers: a critical appraisal of current technology.

John Nicholas Amoore1.   

Abstract

AIM: Blood pressure (BP), a key vital sign, monitors general health. Oscillometric devices are increasingly used for measurement, although their accuracy continues to be critically debated. A functional block diagram is used to review the components that affect accuracy.
METHODS: A block diagram is presented covering the components from cuff to algorithm. The oscillometric waveform is described, considering factors that can alter its shape. Methods used to assess accuracy, including the potential use of simulators, are described. RESULTS AND DISCUSSION: The block diagram focuses attention on cuff, amplifier, signal processing and algorithm. The importance of correct cuff size is emphasized. Accuracy can be affected by the extraction of the oscillometric pulses and the interpolation to compensate for higher deflation rates. Modern electronic amplifiers are assumed to be stable and do not drift, an assumption largely untested. Crucial to accuracy is the algorithm, but there is no standard algorithm and limited theoretical basis, leading to significant measurement errors in groups of patients, even by approved devices. The causes are not well understood, but differences in oscillometric waveform shape between patient groups have been observed and may explain auscultatory-oscillometric differences. The ability of theoretical models to explain the effects of arterial stiffness on BP measurements is discussed. Validation remains statistical though steps have been taken to improve it.
CONCLUSION: The indirect nature of BP measurement poses particularly problems for ensuring accuracy. Critical assessment has done much to improve standards, but a solid theoretical understanding of the technique has not been formulated and further work is required.

Entities:  

Mesh:

Year:  2012        PMID: 22228040     DOI: 10.1097/MBP.0b013e32835026b0

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  9 in total

1.  Even Small Decreases in Blood Pressure during Conscious Sedation Affect Clinical Outcome after Stroke Thrombectomy: An Analysis of Hemodynamic Thresholds.

Authors:  M K Whalin; K M Halenda; D C Haussen; L C Rebello; M R Frankel; R Y Gershon; R G Nogueira
Journal:  AJNR Am J Neuroradiol       Date:  2016-11-03       Impact factor: 3.825

Review 2.  Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure.

Authors:  James E Sharman; Isabella Tan; George S Stergiou; Carolina Lombardi; Francesca Saladini; Mark Butlin; Raj Padwal; Kei Asayama; Alberto Avolio; Tammy M Brady; Alan Murray; Gianfranco Parati
Journal:  J Hum Hypertens       Date:  2022-05-30       Impact factor: 3.012

3.  Discrepancy of blood pressure between the brachial artery and radial artery.

Authors:  Wen-Yuan Li; Xiao-Hai Wang; Li-Chong Lu; Hao Li
Journal:  World J Emerg Med       Date:  2013

4.  Update: ambulatory blood pressure monitoring in children and adolescents: a scientific statement from the American Heart Association.

Authors:  Joseph T Flynn; Stephen R Daniels; Laura L Hayman; David M Maahs; Brian W McCrindle; Mark Mitsnefes; Justin P Zachariah; Elaine M Urbina
Journal:  Hypertension       Date:  2014-03-03       Impact factor: 10.190

5.  The value of a BP determination method using a novel non-invasive BP device against the invasive catheter measurement.

Authors:  Jinsong Xu; Yanqing Wu; Hai Su; Weitong Hu; Juxiang Li; Wenying Wang; Xin Liu; Xiaoshu Cheng
Journal:  PLoS One       Date:  2014-06-23       Impact factor: 3.240

6.  Neighborhood perceptions and hypertension among low-income black women: a qualitative study.

Authors:  Maliyhah Al-Bayan; Nadia Islam; Shawneaqua Edwards; Dustin T Duncan
Journal:  BMC Public Health       Date:  2016-10-12       Impact factor: 3.295

7.  Blood pressure measurements and hypertension in infants, children, and adolescents: from the postmercury to mobile devices.

Authors:  Seon Hee Lim; Seong Heon Kim
Journal:  Clin Exp Pediatr       Date:  2021-09-15

8.  Implementation and Validation of a Novel Noninvasive blood Pressure Measurement Algorithm Based on Oscillometric Method with Obtaining International Hypertension Standards.

Authors:  Sahar Karimian; Mohammadreza Yazdchi; Reza Hajian
Journal:  J Med Signals Sens       Date:  2022-05-12

9.  Assessment of blood pressure in children: it's all in the details.

Authors:  Joseph T Flynn
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-09-09       Impact factor: 3.738

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.