Localising rectus muscle insertions using high frequency wide-field ultrasound biomicroscopy.

Hayat Ahmad Khan1, David R Smith, Stephen P Kraft.   

Abstract

AIM: The ultrasound biomicroscope (UBM) can accurately locate an extraocular muscle (EOM) insertion. The authors compared the accuracy of the Sonomed UBM (SUBM), a new 'wide-field ultrasound biomicroscope', with the older model Humphrey UBM (HUBM) in localising EOM insertions and compared their ranges of detection of muscle insertions.
METHODS: Prospective, double-masked, observational study of 27 patients undergoing primary (n=40 muscles) or repeat (n=10 muscles) horizontal or vertical rectus muscle surgery. EOM insertional distances were measured with SUBM, and then intraoperatively with callipers. A Bland-Altman analysis and intraclass correlation coefficient were used to compare the SUBM and surgical data.
RESULTS: For all muscles, the differences between SUBM and surgery measurements were less than 1.0 mm. The mean of the SUBM insertion distances was 6.67 mm (SD 1.65 mm) versus 6.7 mm (SD 1.6 mm) at surgery. The intraclass correlation coefficient showed 'excellent' correlation between the two sets of data and was higher than that reported with HUBM. The image quality with the SUBM was superior to the HUBM, and its range of field was much larger (14×18 mm vs 5×6 mm).
CONCLUSION: The SUBM with its smaller, more manoeuvrable probe handpiece and a wider scanning field was more accurate in detecting muscle insertions compared with HUBM.

Entities:  

Mesh:

Year:  2012        PMID: 22223747     DOI: 10.1136/bjophthalmol-2011-300960

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  4 in total

1.  One Year of Pediatric Ophthalmology and Strabismus Research in Review.

Authors:  Iris S Kassem; Marilyn T Miller; Steven M Archer
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2013 Nov-Dec

Review 2.  Reliability of Time Domain AS-OCT in Measuring the Extraocular Rectus Muscle Insertion-Limbus Distances: A Systematic Review and Meta-Analysis.

Authors:  Eric J Kim; Arjun Ganga; Viren K Rana; Joshua R Tanzer; Yasmyne C Ronquillo; Majid Moshirfar
Journal:  Clin Ophthalmol       Date:  2022-08-25

Review 3.  A systematic review of ultrasound biomicroscopy use in pediatric ophthalmology.

Authors:  Janet L Alexander; Libby Wei; Jamie Palmer; Alex Darras; Moran R Levin; Jesse L Berry; Emilie Ludeman
Journal:  Eye (Lond)       Date:  2020-09-22       Impact factor: 3.775

4.  Pre-, Intra-, and Post-Operative Evaluation of Extraocular Muscle Insertions Using Optical Coherence Tomography: A Comparison of Four Devices.

Authors:  Matthew S Pihlblad; Andrew Troia; Sapna Tibrewal; Parth R Shah
Journal:  J Clin Med       Date:  2019-10-19       Impact factor: 4.241

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.