In vitro study of the Norwood palliation: a patient-specific mock circulatory system.

Giovanni Biglino1, Alessandro Giardini, Catriona Baker, Richard S Figliola, Tain-Yen Hsia, Andrew M Taylor, Silvia Schievano.   

Abstract

The aim of this study was to build a mock circulatory system replicating in vitro the hemodynamics following the Norwood procedure and testing patient-specific anatomies focusing on the effect of aortic coarctation. Three anatomies were reconstructed from magnetic resonance images and rapid prototyped with transparent rigid resin. The models presented varying degrees of coarctation (none, moderate, and severe). A Blalock-Taussing (BT) shunt was modeled in all phantoms, which were inserted into a mock circulation. The single ventricle was simulated using a Berlin Heart driven with a PC-controlled piston. Resistive and compliant elements were implemented, creating a lumped parameter network. Pressure was measured at three locations: the transverse aortic arch, just after the aortic isthmus, and further downstream in the thoracic aorta. Volume distribution was derived from the instantaneous flow measurements at three outlets: upper body, lower body, and BT shunt. The combination of three-dimensional (3D) detailed anatomy and lumped parameter network effectively renders the circuit a multiscale in vitro model that successfully reproduces physiologic pressure signals. The pressure results highlight the larger pressure drop caused by coarctation and show the effect of pressure recovery. Results also suggest a reduction of flow to the lower body with increasing severity of coarctation, to the advantage of upper body and pulmonary circulation.

Entities:  

Mesh:

Year:  2012        PMID: 22210648     DOI: 10.1097/MAT.0b013e3182396847

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  7 in total

Review 1.  Computer modeling to tailor therapy for congenital heart disease.

Authors:  Michael A Quail; Andrew M Taylor
Journal:  Curr Cardiol Rep       Date:  2013-09       Impact factor: 2.931

2.  Right ventricular afterload in repaired D-TGA is associated with inefficient flow patterns, rather than stenosis alone.

Authors:  Marc Delaney; Vincent Cleveland; Paige Mass; Francesco Capuano; Jason G Mandell; Yue-Hin Loke; Laura Olivieri
Journal:  Int J Cardiovasc Imaging       Date:  2021-11-02       Impact factor: 2.357

3.  Simulating Radial Pressure Waveforms with a Mock Circulatory Flow Loop to Characterize Hemodynamic Monitoring Systems.

Authors:  Anna Packy; Gavin A D'Souza; Masoud Farahmand; Luke Herbertson; Christopher G Scully
Journal:  Cardiovasc Eng Technol       Date:  2021-09-01       Impact factor: 2.305

4.  A Mock Circulatory System Incorporating a Compliant 3D-Printed Anatomical Model to Investigate Pulmonary Hemodynamics.

Authors:  Paul G M Knoops; Giovanni Biglino; Alun D Hughes; Kim H Parker; Linzhang Xu; Silvia Schievano; Ryo Torii
Journal:  Artif Organs       Date:  2016-12-07       Impact factor: 3.094

5.  Rapid prototyping compliant arterial phantoms for in-vitro studies and device testing.

Authors:  Giovanni Biglino; Peter Verschueren; Raf Zegels; Andrew M Taylor; Silvia Schievano
Journal:  J Cardiovasc Magn Reson       Date:  2013-01-16       Impact factor: 5.364

Review 6.  Modeling single ventricle physiology: review of engineering tools to study first stage palliation of hypoplastic left heart syndrome.

Authors:  Giovanni Biglino; Alessandro Giardini; Tain-Yen Hsia; Richard Figliola; Andrew M Taylor; Silvia Schievano
Journal:  Front Pediatr       Date:  2013-10-30       Impact factor: 3.418

7.  Distensibility of Deformable Aortic Replicas Assessed by an Integrated In-Vitro and In-Silico Approach.

Authors:  Luigi Di Micco; Giulia Comunale; Stefano Bonvini; Paolo Peruzzo; Francesca Maria Susin
Journal:  Bioengineering (Basel)       Date:  2022-02-26
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.