Histopathologic study of the Molteno glaucoma implant in three patients.

B Rubin1, C C Chan, M Burnier, L Munion, J Freedman.   

Abstract

Three eyes with the Molteno glaucoma implant (one eye with epithelial downgrowth, one eye with iridocorneal endothelial syndrome, and one eye with aphakia and glaucoma) were enucleated two to six years after implantation. Histopathologic examinations disclosed no evidence of erosion of sclera or conjunctiva of the eye by the glaucoma implant device. In the outer layers of the bleb wall, few and mostly degenerated inflammatory cells were present, which represented a minimal inflammatory reaction. Scanning electron microscopy of the tubes in these three patients showed that the tube was intact, patent, and without signs of degradation. The tube entering into the anterior chamber caused no appreciable inflammation and maintained its patency even when downgrowth epithelial cells lined the anterior chamber. The Molteno plate induced little or no inflammatory reaction. Therefore, the Molteno glaucoma implant is a useful device for patients with high risk for failure after surgery for glaucoma.

Entities:  

Mesh:

Year:  1990        PMID: 2220971     DOI: 10.1016/s0002-9394(14)77017-5

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  6 in total

1.  Survival analysis for success of Molteno tube implants.

Authors:  D C Broadway; M Iester; M Schulzer; G R Douglas
Journal:  Br J Ophthalmol       Date:  2001-06       Impact factor: 4.638

2.  Location of glaucoma drainage devices relative to the optic nerve.

Authors:  M Y Kahook; R J Noecker; M B Pantcheva; J S Schuman
Journal:  Br J Ophthalmol       Date:  2006-04-13       Impact factor: 4.638

3.  Anti-proliferation effects of Sirolimus sustained delivery film in rabbit glaucoma filtration surgery.

Authors:  Zhi-chao Yan; Yu-jing Bai; Zhen Tian; Hai-yan Hu; Xiu-hua You; Jian-xian Lin; Shao-rui Liu; Ye-hong Zhuo; Rong-jiang Luo
Journal:  Mol Vis       Date:  2011-09-27       Impact factor: 2.367

4.  Altered expression of fibrosis genes in capsules of failed Ahmed glaucoma valve implants.

Authors:  Alka Mahale; Maha W Othman; Sami Al Shahwan; Ibrahim Al Jadaan; Ohood Owaydha; Zahid Khan; Deepak P Edward
Journal:  PLoS One       Date:  2015-04-16       Impact factor: 3.240

5.  The Effect of Dry Eye Disease on Scar Formation in Rabbit Glaucoma Filtration Surgery.

Authors:  Hong Ji; Yingting Zhu; Yingying Zhang; Yu Jia; Yiqing Li; Jian Ge; Yehong Zhuo
Journal:  Int J Mol Sci       Date:  2017-05-28       Impact factor: 5.923

6.  Correlation between Tissue Characterization and Dynamic Expression of Matrix Metalloproteinase-2 and Its Tissue Inhibitor in Conjunctival Filtering Bleb of Rats.

Authors:  Ling Wang; Da-Bo Wang; Meng-Ying Liu; Ru-Yong Yao
Journal:  Biomed Res Int       Date:  2017-09-18       Impact factor: 3.411