Location of glaucoma drainage devices relative to the optic nerve.

M Y Kahook1, R J Noecker, M B Pantcheva, J S Schuman.   

Abstract

BACKGROUND: Limited data are available to guide optimal positioning of glaucoma drainage devices (GDD) in relation to the limbus and optic nerve. The authors aim to provide guidelines for appropriate and safe GDD implantation.
METHOD: The optimal positioning of five different GDD were evaluated using necropsy eyes of varying axial lengths. The dependent variable that was measured was the maximum distance that a GDD could be placed posterior to the limbus while remaining 2 mm away from the optic nerve.
RESULTS: The average maximum distance posterior to the limbus of the anterior plate edge ranged between 9.0-15.0 mm in the superotemporal quadrant for the GDD tested. The distances for superonasal, inferonasal, and inferotemporal quadrants ranged between 8.0-14.0 mm, 9.0-14.0 mm, and 11.0-17.0 mm, respectively. The Molteno device could be placed most posteriorly while remaining 2 mm away from the nerve. The Ahmed FP7 and S2 were the least amenable to posterior placement before encroaching on the 2 mm limit.
CONCLUSION: The maximum distance that a GDD can be placed posterior to the limbus, before encroachment around the optic nerve, varies between different devices and quadrants of placement. Taking a measurement of the exact distance of the plate from the limbus during GDD surgery is recommended.

Entities:  

Mesh:

Year:  2006        PMID: 16613923      PMCID: PMC1857219          DOI: 10.1136/bjo.2006.091272

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  8 in total

1.  Histopathologic study of the Molteno glaucoma implant in three patients.

Authors:  B Rubin; C C Chan; M Burnier; L Munion; J Freedman
Journal:  Am J Ophthalmol       Date:  1990-10-15       Impact factor: 5.258

2.  Inferonasal placement of aqueous shunts.

Authors:  Paul A Sidoti
Journal:  J Glaucoma       Date:  2004-12       Impact factor: 2.503

3.  Experimental studies of aqueous filtration using the Molteno implant.

Authors:  D S Minckler; A Shammas; M Wilcox; T E Ogden
Journal:  Trans Am Ophthalmol Soc       Date:  1987

4.  New implant for drainage in glaucoma. Clinical trial.

Authors:  A C Molteno
Journal:  Br J Ophthalmol       Date:  1969-09       Impact factor: 4.638

5.  New implant for drainage in glaucoma. Animal trial.

Authors:  A C Molteno
Journal:  Br J Ophthalmol       Date:  1969-03       Impact factor: 4.638

6.  Anatomic considerations in the implantation of the Ahmed glaucoma valve.

Authors:  M M Leen; G S Witkop; D P George
Journal:  Arch Ophthalmol       Date:  1996-02

7.  Anatomic and histopathologic findings following a failed ahmed glaucoma valve device.

Authors:  R S Ayyala; W E Layden; C B Slonim; C E Margo
Journal:  Ophthalmic Surg Lasers       Date:  2001 May-Jun

8.  Optic nerve changes following posterior insertion of glaucoma drainage device in rabbit model.

Authors:  Ramesh S Ayyala; Scott E Parma; Zeynel A Karcioglu
Journal:  J Glaucoma       Date:  2004-04       Impact factor: 2.503

  8 in total
  2 in total

Review 1.  Modern concepts in antiglaucomatous implant surgery.

Authors:  T S Dietlein; J Jordan; C Lueke; G K Krieglstein
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-08-06       Impact factor: 3.117

Review 2.  Aqueous shunt implantation in glaucoma.

Authors:  Jing Wang; Keith Barton
Journal:  Taiwan J Ophthalmol       Date:  2017 Jul-Sep
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.