Framework for estimating tumour parameters using thermal imaging.

V Umadevi1, S V Raghavan, Sandeep Jaipurkar.   

Abstract

BACKGROUND &
OBJECTIVES: Non-invasive and non-ionizing medical imaging techniques are safe as these can be repeatedly used on as individual and are applicable across all age groups. Breast thermography is a non-invasive and non-ionizing medical imaging that can be potentially used in breast cancer detection and diagnosis. In this study, we used breast thermography to estimate the tumour contour from the breast skin surface temperature.
METHODS: We proposed a framework called infrared thermography based image construction (ITBIC) to estimate tumour parameters such as size and depth from cancerous breast skin surface temperature data. Markov Chain Monte Carlo method was used to enhance the accuracy of estimation in order to reflect clearly realistic situation.
RESULTS: We validated our method experimentally using Watermelon and Agar models. For the Watermelon experiment error in estimation of size and depth parameters was 1.5 and 3.8 per cent respectively. For the Agar model it was 0 and 8 per cent respectively. Further, thermal breast screening was done on female volunteers and compared it with the magnetic resonance imaging. The results were positive and encouraging. INTERPRETATION &
CONCLUSIONS: ITBIC is computationally fast thermal imaging system and is perhaps affordable. Such a system will be useful for doctors or radiologists for breast cancer diagnosis.

Entities:  

Mesh:

Year:  2011        PMID: 22199114      PMCID: PMC3249973          DOI: 10.4103/0971-5916.91012

Source DB:  PubMed          Journal:  Indian J Med Res        ISSN: 0971-5916            Impact factor:   2.375


Breast thermography started during 19561, but the results of Breast Cancer Detection and Demonstration Project (BCDDP) done between 1973 and 1981 by American Cancer Society and National Cancer Institute of United States were not promising due to lack of trained professionals and also primitive thermal imaging techniques2. Recent advances in sensor technology and awareness of using non-invasive and non-ionizing medical imaging techniques has made breast thermography a choice for breast cancer diagnosis. Inference of tumour parameters from surface temperature distribution of cancerous skin is an inverse heat transfer problem which is “ill-posed”3. Methods have been proposed to find parameters of embedded tumours using surface temperature distribution4–6. Thermal texture method4 finds depth of tumour, but its principle is hard to comprehend7. Other methods56 are based on artificial neural network and genetic algorithm. These methods are time consuming and error prone in the presence of the noisy data8. More than 1.2 million people are diagnosed with breast cancer each year worldwide and over 500,000 die from this disease9. Predictions of breast cancer development for the next 10 years9 based on cases diagnosed during 2000-2002 world wide has been done10. On January 29, 1982 United States Food and Drug Administration (FDA) has approved breast thermography for breast cancer detection and diagnosis11. Earlier studies12–14 on breast thermography showed both positive and negative aspects of the technology. Intensities of infra red (IR) radiations emitted by human body vary in accordance to the temperature of the body. The process of capturing, converting, and recording IR radiation emissions from human body is referred as thermal imaging or thermography which is used in medicine for vascular disorder detection, chronic pain detection, assessment of skin burn, dentistry, skin cancer detection, breast cancer detection and many others15–19. Biologically, metabolic rate of cancerous cells is comparatively higher than normal cells. Hence, tumours act as heat source increasing surface temperatures around the cancerous area which could be seen as a hot spot in a thermal image. Here we propose the infrared thermography based image construction (ITBIC) framework to find parameters i.e. size and depth, of a tumour. We used Markov Chain Monte Carlo (MCMC) numerical method20 to enhance the accuracy in parameter estimation. Results were obtained experimentally using synthetic tissue models (watermelon and agar) as well as direct clinical trials using volunteer subjects.

Material & Methods

Two different infrared cameras namely Ti40FT from M/s Fluke Corp., USA and Varioscan 3021 ST from Jenoptik Laser, Germany, were used for breast thermal scanning. Specifications of these two infrared cameras are given in Table I. Fifty women of various age groups were screened for breast thermal scanning at Vijaya scans, Chennai, after informed consent was taken from the patients during May 2007 - March 2010. The study protocol was approved by the ethics committee of the Paterson Cancer Center, Chennai. For each breast thermal images were captured of front, left and right side views. The first 27 sets were screened using Ti40FT and second 23 sets were screened using Varioscan 3021 ST. Procedure followed for screening of first and second sets were different. First 27 sets were screened in a room which was air conditioned and with fluorescent bulb on. To normalize, subjects were made to sit in the examination room for 15 min with upper body part undressed. Then the thermal image of the breast part were captured. Second 23 sets were screened in a room which was not air conditioned and with fluorescent bulb off (i.e. images were captured in a dark room). For this set of subjects images were obtained immediately (i.e. without normalizing) with upper body part undressed. Subject were found to be more comfortable to thermographic procedure when carried out in a dark room.
Table I

Specifications of infrared camera

Specifications of infrared camera Direct estimation of tumour parameters from breast thermal image is a tedious task. We proposed a framework named ITBIC to estimate tumour parameters from surface temperature data (Fig. 1).
Fig. 1

Framework for tumour parameter estimation.

Framework for tumour parameter estimation. In the initial ITBIC framework2122, case of radiative heat transfer model was considered tissue (breast) as cube and tumour as a 2D model i.e., a disc. Subsequently, breast tissue was considered as a 3D hemisphere model and tumour as a sphere model to improve the accuracy of ITBIC framework. Pennes bio-heat transfer equation23 was used to model heat transfer process within the tissue. Table II gives the values of the tissue and tumour considered by our estimation system for solving Pennes heat transfer equation. COMSOL24 software was used for solving Pennes heat transfer equation.
Table II

Pennes bio-heat transfer equation parameters values

Pennes bio-heat transfer equation parameters values ITBIC framework works in two stages, first it estimates actual size & location of tumour and second stage calculates depth. Estimation of size and location: ITBIC frame work is supplied with input of cancerous surface temperature measurements, Tmeas and then initial assumption for tumor size (r0) and location (x0 y0 z0) within the hemisphere (breast model). Using COMSOL software, Pennes bio-heat transfer equation is solved to generate surface temperature measurement, Tgen, for the initial assumed tumour size (r0) and location (x0 y0 z0). Then using Metropolis-Hastings algorithm, a MCMC method20, next estimate of the tumour size (ri) and location (xi yi zi) is generated depending on the comparison result between Tmeas and Tgen. This procedure is repeated in a loop until a stable estimation for tumor size I and location (x y z) is found. Calculation of depth: Using the estimated values of size I and location (x y z) of the tumour inside the breast, depth of the tumour is calculated. Considering the center co-ordinates of the hemisphere as (xc yc zc), estimated location co-ordinates of tumour inside the breast as (x y z) and coordinates on the border surface of the breast as (xb yb zb), value of xb is equivalent to x and value of yb is equivalent to y. Depth of the tumor from the top surface of the breast to be calculated is D, which is as follows: Study with phantom tissue: Framework proposed was validated first with synthetically designed phantom tissue and tumour materials and then clinical studies were carried out to validate. Laboratory experiments were conducted for the study of homogeneous and heterogeneous nature of tissue i.e. Agar model experiment to represent homogeneous and Watermelon experiment to represent heterogeneous nature. The experimental apparatus and set up are shown in Fig. 3.
Fig. 3

Laboratory experimental setup with phantom tissue and tumour material.

(a). Normal and (2b). abnormal breast thermogram. Laboratory experimental setup with phantom tissue and tumour material. Electric bulb was used with phantom tissues to mimic tumour. Diameter of the electric bulb was 1.28 cm. Phantom tissue with bulb inserted was kept on a thermocol sheet. Electric bulb was connected to power supply unit. Agar model experiment: Agar powder was boiled in water around 30 min and poured into a plastic jar which contained 4.5 watt electric bulb. The solution was then allowed to cool down to room temperature. At the steady state of room temperature the size of cast model obtained was 10 × 10 × 9 cm (length × width × height), which was shaped like a cube to mimic tissue. Watermelon experiment: Watermelon pulp of size 10 × 10 × 10 cm (length × width × height) was used. Pulp was shaped like a cube to mimic tissue. An electric bulb of 4.5 watts was introduced into the cube to mimic tumour.

Results

Example of an normal and abnormal breast thermal image obtained from women volunteers are shown in Figs 2a and 2b as captured by the Varioscan 3021 ST camera at a temperature range of 32-38°C.
Fig. 2

(a). Normal and (2b). abnormal breast thermogram.

Temperature distribution between left and right parts of the breast was symmetrical for a normal case (45 yr old healthy woman) (Fig. 2a) (as the colour pattern is symmetrical). There will be rise in temperature of the breast skin surface area above the existence of cancerous mass. Thermogram of an abnormal case (female aged 54 yr) (Fig. 2b) showed asymmetrical colour pattern. Her self breast examination showed lump in her left breast part, so she reported to clinic for further examination. MRI scan report of this woman confirmed having cancerous mass in left part of the breast. In the thermal image, left part of the breast showed the increased surface temperature when compared to right part of the breast. Further, biopsy test conducted on this patient also proved lump as malignant neoplasm. With the experimental set up, the electric bulb was switched on for around 30 min and then thermal images were captured using the camera Fluke Ti40FT for Agar and watermelon experiments (Figs 4, 5). Twenty five surface temperatures, from the top and from a side of agar model were measured. This measurement values were given as an input to ITBIC framework to estimate radius and depth of the tumour. Estimated parameter values against actual values are given in Table III. The estimated values were very close to the actual values.
Fig. 4

Infrared images of agar model with surface temperature distribution.

Fig. 5

Watermelon model surface temperature distribution.

Table III

ITBIC framework result for Agar model and watermalon experiment

Infrared images of agar model with surface temperature distribution. Watermelon model surface temperature distribution. ITBIC framework result for Agar model and watermalon experiment Among 50 women, two had a focal area of raised temperature in left and right breasts respectively. The same subjects were scanned by MRI and also biopsy was carried out, both tests proved the area to be malignant neoplasm. The temperature difference between normal breast surface and the neoplastic zone was greater than 2°C. Thermal image and MRI of the first case are shown in Fig 6a and Fig 6b. She was 54 yr old and had menopause at the age of 44. In the upper part of the left breast (which is marked by a block) tumour was present. Temperature measurements of this skin surface were used as an input to our ITBIC framework to estimate tumour size and depth. Estimated tumour size and depth are given in Table IV. Results of our framework correlated closely with the MRI findings.
Fig. 6

(a). Breast thermal scan image of Case 1 showing the abnormality. (b) MRI scan image of Case 1showing the same abnormality.

Table IV

ITBIC framework results for case 1 and 2 breast thermogram

(a). Breast thermal scan image of Case 1 showing the abnormality. (b) MRI scan image of Case 1showing the same abnormality. ITBIC framework results for case 1 and 2 breast thermogram Thermal image and MRI of the second case shown in Fig. 7a and Fig. 7b. She was around 37 yr old. In the lower part of the right breast (which is marked by a block) the tumour was present. Estimated tumour size and depth are given in Table IV.
Fig. 7

(a) Breast thermal scan image of Case 2 showing the abnormality. (b) MRI scan image of Case 2 showing the same abnormality.

(a) Breast thermal scan image of Case 2 showing the abnormality. (b) MRI scan image of Case 2 showing the same abnormality. Of the 50 female subjects, 44 were normal, two had malignancy, three had fibrocystic disease and one was abscess case. With thermography, temperature difference between left and right parts of the breast in each of these cases were observed (Table V).
Table V

Observed breast temperature variation

Observed breast temperature variation

Discussion

Presently commonly used medical imaging techniques for breast cancer are mammography, MRI and ultrasound, each having their own advantages and limitations. Our improved ITBIC framework surpassed all the previous vogue for methods in tumour parameter estimation5625 in terms of accuracy and simulation time. The results of ITBIC framework in comparison with conventional method i.e. MRI showed a promising relation in estimation of tumour parameters from breast thermograms. Further study of ITBIC framework on large data set need to be done to judge its performance. In conclusion, a framework was developed to identify parameters of tumour in human female breasts from cancerous skin surface temperature. Information on the size and depth of a tumour may provide critical information to medical professionals for diagnosis. Tumour size and depth parameter values were estimated by ITBIC framework from surface temperature measurements of breast. Estimated values matched with actual values given by MRI scanning method. This shows practicability of using ITBIC framework by medical professionals to diagnose the breast tumour. Currently, the procedure of breast thermography is not practiced or less in use. Progressively, the use of breast thermography for scanning of subjects may supercede the other scanning methods due to its advantages.
  10 in total

Review 1.  Infrared imaging: making progress in fulfilling its medical promise.

Authors:  Jonathan F Head; Robert L Elliott
Journal:  IEEE Eng Med Biol Mag       Date:  2002 Nov-Dec

2.  Implications of surface temperatures in the diagnosis of breast cancer.

Authors:  R LAWSON
Journal:  Can Med Assoc J       Date:  1956-08-15       Impact factor: 8.262

3.  Application of thermography in dentistry--visualization of temperature distribution on oral tissues.

Authors:  Manabu Komoriyama; Rie Nomoto; Rin Tanaka; Noriyasu Hosoya; Kazuhiro Gomi; Fumiaki Iino; Akihiro Yashima; Yasuko Takayama; Masahiko Tsuruta; Hajime Tokiwa; Kenzo Kawasaki; Takashi Arai; Toshio Hosoi; Ayao Hirashita; Susumu Hirano
Journal:  Dent Mater J       Date:  2003-12       Impact factor: 2.102

4.  Fundamental theoretic research of thermal texture maps I--simulation and analysis of the relation between the depth of inner heat source and surface temperature distribution in isotropy tissue.

Authors:  Zhigang Shang; Guotai Jiang
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2004

5.  Thermal parametric imaging in the evaluation of skin burn depth.

Authors:  Jacek Rumiński; Mariusz Kaczmarek; Alicja Renkielska; Antoni Nowakowski
Journal:  IEEE Trans Biomed Eng       Date:  2007-02       Impact factor: 4.538

6.  Improved infrared thermography based image construction for biomedical applications using Markov Chain Monte Carlo method.

Authors:  V Umadevi; S Suresh; S V Raghavan
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009

Review 7.  Long-term survival of cancer patients compared to heart failure and stroke: a systematic review.

Authors:  Vasileios Askoxylakis; Christian Thieke; Sven T Pleger; Patrick Most; Judith Tanner; Katja Lindel; Hugo A Katus; Jürgen Debus; Marc Bischof
Journal:  BMC Cancer       Date:  2010-03-22       Impact factor: 4.430

8.  Functional infrared imaging for skin-cancer screening.

Authors:  Thorsten M Buzug; Steffen Schumann; Lucas Pfaffmann; Uwe Reinhold; Jürgen Ruhlmann
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2006

9.  Thermography in screening for breast cancer.

Authors:  K L Williams; B H Phillips; P A Jones; S A Beaman; P J Fleming
Journal:  J Epidemiol Community Health       Date:  1990-06       Impact factor: 3.710

Review 10.  An overview of recent application of medical infrared thermography in sports medicine in Austria.

Authors:  Carolin Hildebrandt; Christian Raschner; Kurt Ammer
Journal:  Sensors (Basel)       Date:  2010-05-07       Impact factor: 3.576

  10 in total
  7 in total

1.  Exhaled breath temperature in NSCLC: could be a new non-invasive marker?

Authors:  Giovanna E Carpagnano; Donato Lacedonia; Antonio Spanevello; Domenico Martinelli; Valerio Saliani; Cinzia Ruggieri; Maria P Foschino-Barbaro
Journal:  Med Oncol       Date:  2014-04-11       Impact factor: 3.064

2.  Thermal distribution analysis of three-dimensional tumor-embedded breast models with different breast density compositions.

Authors:  Asnida Abd Wahab; Maheza Irna Mohamad Salim; Mohamad Asmidzam Ahamat; Noraida Abd Manaf; Jasmy Yunus; Khin Wee Lai
Journal:  Med Biol Eng Comput       Date:  2015-10-13       Impact factor: 2.602

3.  Detection of breast abnormality from thermograms using curvelet transform based feature extraction.

Authors:  Sheeja V Francis; M Sasikala; S Saranya
Journal:  J Med Syst       Date:  2014-03-23       Impact factor: 4.460

4.  Determining the thermal characteristics of breast cancer based on high-resolution infrared imaging, 3D breast scans, and magnetic resonance imaging.

Authors:  Adolfo Lozano; Jody C Hayes; Lindsay M Compton; Jamasp Azarnoosh; Fatemeh Hassanipour
Journal:  Sci Rep       Date:  2020-06-22       Impact factor: 4.379

5.  Assessment of Computer Regulation Thermography (CRT) as a Complemetrary Diagnostic tool for Breast Cancer Patient.

Authors:  B Hashemi; F Hasanaj; M E Akbari; H R Mirzaei; M Mojtahed; M Bakhshandeh
Journal:  J Biomed Phys Eng       Date:  2019-12-01

6.  Exhaled Breath Temperature Home Monitoring to Detect NSCLC Relapse: Results from a Pilot Study.

Authors:  Giovanna Elisiana Carpagnano; Todor A Popov; Giulia Scioscia; Nicoletta Pia Ardò; Donato Lacedonia; Mario Malerba; Pasquale Tondo; Piera Soccio; Domenico Loizzi; Maria Pia Foschino Barbaro; Francesco Sollitto
Journal:  Biomed Res Int       Date:  2022-02-21       Impact factor: 3.411

7.  Feasibility for Using Thermography Throughout an Exercise Program in Mastectomized Patients.

Authors:  Maria Jane das Virgens Aquino; Paula Michele Dos Santos Leite; Ingrid Kyelli Lima Rodrigues; Josimari Melo DeSantana
Journal:  Front Oncol       Date:  2022-04-14       Impact factor: 5.738

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.