Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments.

E S J Pereira1, I F C Peixoto, A C D Viana, I I Oliveira, B M Gonzalez, V T L Buono, M G A Bahia.   

Abstract

AIM: To compare physical and mechanical properties of one conventional and one thermomechanically treated nickel-titanium (NiTi) wire used to manufacture rotary endodontic instruments.
METHODOLOGY: Two NiTi wires 1.0 mm in diameter were characterized; one of them, C-wire (CW), was processed in the conventional manner, and the other, termed M-Wire (MW), received an additional heat treatment according to the manufacturer. Chemical composition was determined by energy-dispersive X-ray spectroscopy, phase constitution by XRD and the transformation temperatures by DSC. Tensile loading/unloading tests and Vickers microhardness measurements were performed to assess the mechanical behaviour. Data were analysed using analysis of variance (α = 0.05).
RESULTS: The two wires showed approximately the same chemical composition, close to the 1 : 1 atomic ratio, and the β-phase was the predominant phase present. B19' martensite and the R-phase were found in MW, in agreement with the higher transformation temperatures found in this wire compared with CW, whose transformation temperatures were below room temperature. Average Vickers microhardness values were similar for MW and CW (P = 0.91). The stress at the transformation plateau in the tensile load-unload curves was lower and more uniform in the M-Wire, which also showed the smallest stress hysteresis and apparent elastic modulus.
CONCLUSIONS: The M-Wire had physical and mechanical properties that can render endodontic instruments more flexible and fatigue resistant than those made with conventionally processed NiTi wires.
© 2011 International Endodontic Journal.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22188417     DOI: 10.1111/j.1365-2591.2011.01998.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  31 in total

1.  "Dentinal microcracks after root canal preparation" a comparative evaluation with hand, rotary and reciprocating instrumentation.

Authors:  N Tulasi Priya; Veeramachaneni Chandrasekhar; S Anita; Muralidhar Tummala; T B Phanindhar Raj; Vijetha Badami; Pradeep Kumar; E Soujanya
Journal:  J Clin Diagn Res       Date:  2014-12-05

2.  Shaping ability of reciprocating single-file systems in severely curved canals: WaveOne and Reciproc versus WaveOne Gold and Reciproc blue.

Authors:  Sebastian Bürklein; Stefanie Flüch; Edgar Schäfer
Journal:  Odontology       Date:  2018-05-18       Impact factor: 2.634

3.  Mechanism of fracture of NiTi superelastic endodontic rotary instruments.

Authors:  Javier Gil; Elisa Rupérez; Eugenio Velasco; Conrado Aparicio; José María Manero
Journal:  J Mater Sci Mater Med       Date:  2018-08-07       Impact factor: 3.896

4.  Both operator and heat treatment determine the centring ability of Reciproc® files in vitro.

Authors:  N Hofmann
Journal:  Clin Oral Investig       Date:  2018-05-12       Impact factor: 3.573

5.  Metallurgical analysis and fatigue resistance of WaveOne and ProTaper nickel-titanium instruments.

Authors:  Chiara Pirani; Oddone Ruggeri; Pier Paolo Cirulli; Gian Andrea Pelliccioni; Maria Giovanna Gandolfi; Carlo Prati
Journal:  Odontology       Date:  2013-04-09       Impact factor: 2.634

6.  A comparative study of cyclic fatigue of 6 endodontic systems. An in vitro study.

Authors:  Jorge Rubio; José-Ignacio Zarzosa; Susana Aranda; Alberto Casino; Antonio Pallarés
Journal:  J Clin Exp Dent       Date:  2022-07-01

Review 7.  Endodontic Rotary Files, What Should an Endodontist Know?

Authors:  Ana-Belén Dablanca-Blanco; Pablo Castelo-Baz; Ramón Miguéns-Vila; Pablo Álvarez-Novoa; Benjamín Martín-Biedma
Journal:  Medicina (Kaunas)       Date:  2022-05-27       Impact factor: 2.948

8.  Comparison of the cyclic fatigue resistance of One Curve, F6 Skytaper, Protaper Next, and Hyflex CM endodontic files.

Authors:  Charlotte Gouédard; Laurent Pino; Reza Arbab-Chirani; Shabnam Arbab-Chirani; Valérie Chevalier
Journal:  Restor Dent Endod       Date:  2022-03-04

9.  A comparative study of mechanical resistance of two reciprocating files.

Authors:  Dario Di Nardo; Massimo Galli; Antonio Morese; Marco Seracchiani; Valerio Ferri; Gabriele Miccoli; Gianluca Gambarini; Luca Testarelli
Journal:  J Clin Exp Dent       Date:  2019-03-01

10.  In vitro resistance to fracture of two nickel-titanium rotary instruments made with different thermal treatments.

Authors:  Gabriele Miccoli; Gianfranco Gaimari; Marco Seracchiani; Antonio Morese; Tatyana Khrenova; Dario Di Nardo
Journal:  Ann Stomatol (Roma)       Date:  2017-11-08
View more