High-resolution esophageal manometry: addressing thermal drift of the manoscan system.

E V Robertson1, Y Y Lee, M H Derakhshan, A A Wirz, J R H Whiting, J P Seenan, P Connolly, K E L McColl.   

Abstract

BACKGROUND: The high resolution esophageal manometry system manufactured by Sierra Scientific Instruments is widely used. The technology is liable to 'thermal drift', a change in measured pressure due to change in temperature. This study aims to characterize 'thermal drift' and minimize its impact.
METHODS: Response of the system to immediate temperature change (20 °C to 37 °C) was tested. Accuracy of pressure measurement over two hours at 37 °C was examined. Six repetitions were performed and median pressure change calculated for each sensor. Sensors were compared using Kruskal-Wallis test. Current correction processes were tested. KEY
RESULTS: There was a biphasic response of the system to body temperature: an immediate change in recorded pressure, 'thermal effect' and an ongoing pressure change with time, 'baseline drift'. Median thermal effect for all 36 sensors was 7 mmHg (IQR 3.8 mmHg). Median baseline drift was 11.1 mmHg (IQR 9.9 mmHg). Baseline drift varied between sensors but for a given sensor was linear. Interpolated thermal compensation, recommended for prolonged studies, corrects data assuming a linear drift of pressures. When pressures were corrected in this way, baseline pressure was almost restored to zero (Median 0.3 mmHg, IQR 0.3). The standard thermal compensation process did not address the error associated with baseline drift. CONCLUSIONS & INFERENCES: Thermal effect is well compensated in the current operation of the system but baseline drift is not well recognized or addressed. Incorporation of a linear correction into current software would improve accuracy without impact on ease of use.
© 2011 Blackwell Publishing Ltd.

Entities:  

Mesh:

Year:  2012        PMID: 22188326     DOI: 10.1111/j.1365-2982.2011.01817.x

Source DB:  PubMed          Journal:  Neurogastroenterol Motil        ISSN: 1350-1925            Impact factor:   3.598


  15 in total

Review 1.  High resolution and high definition anorectal manometry and pressure topography: diagnostic advance or a new kid on the block?

Authors:  Yeong Yeh Lee; Askin Erdogan; Satish S C Rao
Journal:  Curr Gastroenterol Rep       Date:  2013-12

2.  Pharyngeal Swallowing During Wake and Sleep.

Authors:  Esther Guiu Hernandez; Kristin Gozdzikowska; Richard D Jones; Maggie-Lee Huckabee
Journal:  Dysphagia       Date:  2019-02-26       Impact factor: 3.438

3.  Pharyngeal peristaltic pressure variability, operational range, and functional reserve.

Authors:  Mark K Kern; Gokulakrishnan Balasubramanian; Patrick Sanvanson; Dilpesh Agrawal; Ashley Wuerl; Reza Shaker
Journal:  Am J Physiol Gastrointest Liver Physiol       Date:  2017-03-02       Impact factor: 4.052

4.  Comparison of unidirectional and circumferential manometric measures within the pharyngoesophageal segment: an exploratory study.

Authors:  Esther Guiu Hernandez; Kristin Gozdzikowska; Richard Jones; Maggie-Lee Huckabee
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-06-26       Impact factor: 2.503

5.  Roux Limb Motility in Gastric Bypass Patients with Chronic Abdominal Pain-Is There an Association to Prescribed Opioids?

Authors:  Per Björklund; Almantas Maleckas; Hans Lönroth; Niclas Björnfot; Sven Egron Thörn; Lars Fändriks
Journal:  Obes Surg       Date:  2019-12       Impact factor: 4.129

Review 6.  High-resolution anorectal manometry: An expensive hobby or worth every penny?

Authors:  G Basilisco; A E Bharucha
Journal:  Neurogastroenterol Motil       Date:  2017-08       Impact factor: 3.598

7.  Determinants of pressure drift in Manoscan(™) esophageal high-resolution manometry system.

Authors:  A Babaei; E C Lin; A Szabo; B T Massey
Journal:  Neurogastroenterol Motil       Date:  2014-12-29       Impact factor: 3.598

8.  Determinants and clinical impact of pressure drift in manoscan anorectal high resolution manometry system.

Authors:  G Parthasarathy; J McMaster; K Feuerhak; A R Zinsmeister; A E Bharucha
Journal:  Neurogastroenterol Motil       Date:  2016-04-08       Impact factor: 3.598

9.  Pressure exposure and catheter impingement affect the recorded pressure in the Manoscan 360™ system.

Authors:  A Babaei; A Szabo; S D Yorio; B T Massey
Journal:  Neurogastroenterol Motil       Date:  2018-03-09       Impact factor: 3.598

10.  How to Perform and Interpret a High-resolution Anorectal Manometry Test.

Authors:  Tae Hee Lee; Adil E Bharucha
Journal:  J Neurogastroenterol Motil       Date:  2016-01-31       Impact factor: 4.924

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.