| Literature DB >> 22187837 |
Bobin Varghese1, Nagarajan Muthukumar, Muthukumar Balasubramaniam, Andrew Scally.
Abstract
The purpose of this study was to assess the accuracy of digital radiographs and hence their effectiveness in templating. The methodology involved a retrospective study of post operative radiographs of patients with hemiarthroplasty of the hip. Three observers made observations blinded to each other's measurements. A statistical analysis of the data highlights magnification varying from 6 to 31 percent. There is a statistically significant relationship between the size of the error (size measured on radiograph minus implant size, i.e. magnification) and the implant size (p = 0.005) but the percentage error (error/implant size x 100) is independent of implant size (p = 0.505). It is our impression that digital radiographs and templating on the digital radiographs should not be considered a precise process.Entities:
Mesh:
Year: 2011 PMID: 22187837
Source DB: PubMed Journal: Acta Orthop Belg ISSN: 0001-6462 Impact factor: 0.500