Examination of ten fractured Oxford unicompartmental knee bearings.

E Pegg1, H Pandit, H S Gill, G W Keys, U G C Svard, J J O'Connor, D W Murray.   

Abstract

Since the Oxford knee was first used unicompartmentally in 1982, a small number of bearings have fractured. Of 14 retrieved bearings, we examined ten samples with known durations in situ (four Phase 1, four Phase 2 and two Phase 3). Evidence of impingement and associated abnormally high wear (> 0.05 mm per year) as well as oxidation was observed in all bearings. In four samples the fracture was associated with the posterior radio-opaque wire. Fracture surfaces indicated fatigue failure, and scanning electron microscopy suggested that the crack initiated in the thinnest region. The estimated incidence of fracture was 3.20% for Phase 1, 0.74% for Phase 2, 0.35% for Phase 3, and 0% for Phase 3 without the posterior marker wire. The important aetiological factors for bearing fracture are impingement leading to high wear, oxidation, and the posterior marker wire. With improved surgical technique, impingement and high wear should be prevented and modern polyethylene may reduce the oxidation risk. A posterior marker wire is no longer used in the polyethylene meniscus. Therefore, the rate of fracture, which is now very low, should be reduced to a negligible level.

Entities:  

Mesh:

Year:  2011        PMID: 22161922     DOI: 10.1302/0301-620X.93B12.27408

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  5 in total

1.  An uncommon cause of cemented unicompartmental knee arthroplasty failure: fracture of metallic components.

Authors:  Alfonso Manzotti; Cesare Chemello; Chris Pullen; Pietro Cerveri; Norberto Confalonieri
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-05-26       Impact factor: 4.342

2.  Minimally invasive Oxford medial unicompartmental knee arthroplasty in young patients.

Authors:  Marcus R Streit; Julia Streit; Tilman Walker; Thomas Bruckner; J Philippe Kretzer; Volker Ewerbeck; Christian Merle; Peter R Aldinger; Tobias Gotterbarm
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-05-10       Impact factor: 4.342

3.  Fracture and Posterior Dislocation of Meniscal Bearing Insert in Mobile Bearing Unicompartmental Knee Arthroplasty: A Case Report.

Authors:  A Munjal
Journal:  Malays Orthop J       Date:  2018-07

4.  Long-term in vivo wear of different bearing types used for the Oxford Unicompartmental Knee Replacement.

Authors:  Hasan R Mohammad; Stefano Campi; James A Kennedy; Andrew Judge; David W Murray; Stephen J Mellon
Journal:  Bone Joint Res       Date:  2019-12-03       Impact factor: 5.853

5.  Low polyethylene creep and wear following mobile-bearing unicompartmental knee replacement.

Authors:  Priyanka Ghosh; Hasan R Mohammad; Benjamin Martin; Stefano Campi; David W Murray; Stephen J Mellon
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2020-09-17       Impact factor: 4.342

  5 in total