A new water-equivalent 2D plastic scintillation detectors array for the dosimetry of megavoltage energy photon beams in radiation therapy.

Mathieu Guillot1, Luc Beaulieu, Louis Archambault, Sam Beddar, Luc Gingras.   

Abstract

PURPOSE: The objective of this work is to present a new 2D plastic scintillation detectors array (2D-PSDA) designed for the dosimetry of megavoltage (MV) energy photon beams in radiation therapy and to characterize its basic performance.
METHODS: We developed a 2D detector array consisting of 781 plastic scintillation detectors (PSDs) inserted into a plane of a water-equivalent phantom. The PSDs were distributed on a 26 × 26 cm(2) grid, with an interdetector spacing of 10 mm, except for two perpendicular lines centered on the detection plane, where the spacing was 5 mm. Each PSD was made of a 1 mm diameter by 3 mm long cylindrical polystyrene scintillating fiber coupled to a clear nonscintillating plastic optical fiber. All of the light signals emitted by the PSDs were read simultaneously with an optical system at a rate of one measurement per second. We characterized the performance of the optical system, the angular dependency of the device, and the perturbation of dose distributions caused by the hundreds of PSDs inserted into the phantom. We also evaluated the capacity of the system to monitor complex multileaf collimator (MLC) sequences such as those encountered in step-and-shoot intensity modulated radiation therapy (IMRT) plans. We compared our results with calculations performed by a treatment planning system and with measurements taken with a 2D ionization chamber array and with a radiochromic film.
RESULTS: The detector array that we developed allowed us to measure doses with an average precision of better than 1% for cumulated doses equal to or greater than 6.3 cGy. Our results showed that the dose distributions produced by the 6-MV photon beam are not perturbed (within ±1.1%) by the presence of the hundreds of PSDs located into the phantom. The results also showed that the variations in the beam incidences have little effect on the dose response of the device. For all incidences tested, the passing rates of the gamma tests between the 2D-PSDA and the treatment planning system were higher than 97.5% when the standard clinical tolerances of 3% or 3 mm were used. Excellent agreement was obtained between the doses measured and calculated when we used the 2D-PSDA for monitoring a MLC sequence from a step-and-shoot IMRT plan.
CONCLUSIONS: We demonstrated the feasibility of using a large number of PSDs in a new 2D-PSDA for the dosimetry of MV energy photon beams in radiation therapy. The excellent precision, accuracy, and low angular dependence of the device indicate that such a prototype could potentially be used as a high-accuracy quality assurance tool for IMRT and arc therapy patient plan verification. The homogeneity and water-equivalence of the prototype we built suggest that this technology could be extended to multiple detection planes by arranging the fibers into more complex orientations, opening the possibility for 3D dosimetry with PSDs.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22149858     DOI: 10.1118/1.3664007

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Deep learning-augmented radiotherapy visualization with a cylindrical radioluminescence system.

Authors:  Mengyu Jia; Xiaomeng Li; Yan Wu; Yong Yang; Priya Kasimbeg; Lawrie Skinner; Lei Wang; Lei Xing
Journal:  Phys Med Biol       Date:  2021-02-09       Impact factor: 3.609

Review 2.  A review of recent advances in optical fibre sensors for in vivo dosimetry during radiotherapy.

Authors:  S O'Keeffe; D McCarthy; P Woulfe; M W D Grattan; A R Hounsell; D Sporea; L Mihai; I Vata; G Leen; E Lewis
Journal:  Br J Radiol       Date:  2015-03-11       Impact factor: 3.039

3.  Projection imaging of photon beams by the Čerenkov effect.

Authors:  Adam K Glaser; Scott C Davis; David M McClatchy; Rongxiao Zhang; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

4.  Video-rate optical dosimetry and dynamic visualization of IMRT and VMAT treatment plans in water using Cherenkov radiation.

Authors:  Adam K Glaser; Jacqueline M Andreozzi; Scott C Davis; Rongxiao Zhang; Brian W Pogue; Colleen J Fox; David J Gladstone
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

5.  Polyethylene Naphthalate Scintillator: A Novel Detector for the Dosimetry of Radioactive Ophthalmic Applicators.

Authors:  Dirk Flühs; Andrea Flühs; Melanie Ebenau; Marion Eichmann
Journal:  Ocul Oncol Pathol       Date:  2015-06-06

6.  Development of a novel multi-point plastic scintillation detector with a single optical transmission line for radiation dose measurement.

Authors:  François Therriault-Proulx; Louis Archambault; Luc Beaulieu; Sam Beddar
Journal:  Phys Med Biol       Date:  2012-10-12       Impact factor: 3.609

7.  Projection imaging of photon beams using Čerenkov-excited fluorescence.

Authors:  Adam K Glaser; Scott C Davis; William H A Voigt; Rongxiao Zhang; Brian W Pogue; David J Gladstone
Journal:  Phys Med Biol       Date:  2013-01-14       Impact factor: 3.609

8.  Quality assurance in proton beam therapy using a plastic scintillator and a commercially available digital camera.

Authors:  Mansour Almurayshid; Yusuf Helo; Andrzej Kacperek; Jennifer Griffiths; Jem Hebden; Adam Gibson
Journal:  J Appl Clin Med Phys       Date:  2017-07-29       Impact factor: 2.102

9.  Feasibility of Ultra-Thin Fiber-Optic Dosimeters for Radiotherapy Dosimetry.

Authors:  Bongsoo Lee; Guwon Kwon; Sang Hun Shin; Jaeseok Kim; Wook Jae Yoo; Young Hoon Ji; Kyoung Won Jang
Journal:  Sensors (Basel)       Date:  2015-11-17       Impact factor: 3.576

10.  Characterization of a Low-Cost Plastic Fiber Array Detector for Proton Beam Dosimetry.

Authors:  Cigdem Ozkan Loch; Michael Alexander Eichenberger; Michele Togno; Simon Pascal Zinsli; Martina Egloff; Angela Papa; Rasmus Ischebeck; Antony John Lomax; Peter Peier; Sairos Safai
Journal:  Sensors (Basel)       Date:  2020-10-09       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.