A feasibility study of using couch-based real time dosimetric device in external beam radiotherapy.

Ramachandran Prabhakar1, Jim Cramb, Tomas Kron.   

Abstract

PURPOSE: Measurement of actual dose delivered during radiotherapy treatment aids in checking the accuracy of dose delivered to the patient. In this study, a couch-based real time dosimetric device has been proposed to measure the exit or entrance dose to a patient during external beam radiotherapy. The utility and feasibility of such a device using a 2D array of diodes has been demonstrated.
METHODS: Two MAPCHECK devices: MAPCHECK (1175) and MAPCHECK 2 (both SunNuclear) were embedded in a foam block in the treatment couch of a Varian 21iX linear accelerator. The angular dependence of the detector response for both devices was studied before implementing the MAPCHECKs for experimental purposes. An Alderson Rando head phantom was scanned with the MAPCHECK and MAPCHECK 2 devices separately and four different treatment plans were generated with target volumes at three different positions simulating typical clinical situations. The analytical anisotropic algorithm (AAA) was used to compute the doses in an Eclipse treatment planning system (Varian Medical Systems). The Rando phantom with the MAPCHECK device was exposed in Clinac 21iX linear accelerator. The measured dose distribution was compared with the calculated dose distribution to check for the accuracy in dose delivery.
RESULTS: Measured and computed dose distribution were found to agree with more than 93% of pixels passing at 3% and 3 mm gamma criteria for all the treatment plans. The couch-based real time dosimetry system may also be applied for noncoplanar beams where electronic portal imaging device (EPID) is not practical to measure the dose. Other advantages include checking the beam stability during the patient treatment, performing routine morning quality assurance (QA) tests in the linear accelerator, and to perform pretreatment verification of intensity modulated radiation therapy (IMRT). One of the drawbacks of this system is that it cannot be used for measuring the dose at 90° or 270° gantry angles.
CONCLUSIONS: This preliminary study shows that a 2D array of detectors may be used as part of the treatment couch for real time patient dosimetry in studying the dose delivered to the patient in real time and also for performing routine quality assurance.

Entities:  

Mesh:

Year:  2011        PMID: 22149836     DOI: 10.1118/1.3660773

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Real-time dosimetry in external beam radiation therapy.

Authors:  Ramachandran Prabhakar
Journal:  World J Radiol       Date:  2013-10-28

2.  The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Arno J Mundt; Sasa Mutic; Jatinder R Palta; Frank Rath; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

3.  Noncoplanar verification: a feasibility study using Philips' Pinnacle3 treatment planning system.

Authors:  Indra Yohannes; Heru Prasetio; Christoph Bert
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.