Literature DB >> 22146812

A comparative study on screw loosening in osteoporotic lumbar spine fusion between expandable and conventional pedicle screws.

Zi-xiang Wu1, Fu-tai Gong, Li Liu, Zhen-sheng Ma, Yang Zhang, Xiong Zhao, Min Yang, Wei Lei, Hong-xun Sang.   

Abstract

INTRODUCTION: The aim of this study is to compare the rate of screw loosening and clinical outcomes of expandable pedicle screws (EPS) with those of conventional pedicle screws (CPS) in patients treated for spinal stenosis (SS) combined with osteoporosis.
METHODS: One hundred and fifty-seven consecutive patients with SS received either EPS fixation (n = 80) or CPS fixation (n = 77) to obtain lumbosacral stabilization. Patients were observed for a minimum of 24 months. Outcome measures included screw loosening, fusion rate, Japanese Orthopaedic Association (JOA) scores and Oswestry disability index (ODI) scoring system, and complications.
RESULTS: In the EPS group, 20 screws became loose (4.1%) in 6 patients (7.5%), and two screws (0.4%) had broken. In the CPS group, 48 screws became loose (12.9%) in 15 patients (19.5%), but no screws were broken. The fusion rate in the EPS group (92.5%) was significantly higher than that of the CPS group (80.5%). The rate of screw loosening in the EPS group (4.1%) was significantly lower than that of the CPS group (12.9%). Six EPS (1.8%) screws were removed. In the EPS group, two screws had broken but without neural complications. Twelve months after surgeries, JOA and ODI scores in the EPS group were significantly improved. There were four cases of dural tears, which healed after corresponding treatment.
CONCLUSIONS: EPS can decrease the risk of screw loosening and achieve better fixation strength and clinical results in osteoporotic lumbar spine fusion.

Entities:  

Mesh:

Year:  2011        PMID: 22146812     DOI: 10.1007/s00402-011-1439-6

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  50 in total

1.  Development and validation of a quantitative method to assess pedicle screw loosening in posterior spine instrumentation on plain radiographs.

Authors:  Emin Aghayev; Nicolas Zullig; Peter Diel; Daniel Dietrich; Lorin M Benneker
Journal:  Eur Spine J       Date:  2013-10-31       Impact factor: 3.134

2.  The contribution of the cortical shell to pedicle screw fixation.

Authors:  Matthew Henry Pelletier; Nicky Bertollo; Darweesh Al-Khawaja; William Robert Walsh
Journal:  J Spine Surg       Date:  2017-06

3.  A pedicle screw system and a lamina hook system provide similar primary and long-term stability: a biomechanical in vitro study with quasi-static and dynamic loading conditions.

Authors:  Hans-Joachim Wilke; Dominik Kaiser; David Volkheimer; Carsten Hackenbroch; Klaus Püschel; Michael Rauschmann
Journal:  Eur Spine J       Date:  2016-07-12       Impact factor: 3.134

4.  Can cavity-based pedicle screw augmentation decrease screw loosening? A biomechanical in vitro study.

Authors:  Hans-Joachim Riesner; Thomas R Blattert; Renate Krezdorn; Simone Schädler; Hans-Joachim Wilke
Journal:  Eur Spine J       Date:  2020-12-23       Impact factor: 3.134

5.  Time to augment?! Impact of cement augmentation on pedicle screw fixation strength depending on bone mineral density.

Authors:  Lukas Weiser; Gerd Huber; Kay Sellenschloh; Lennart Viezens; Klaus Püschel; Michael M Morlock; Wolfgang Lehmann
Journal:  Eur Spine J       Date:  2018-06-09       Impact factor: 3.134

6.  Experimental validation of adaptive pedicle screws-a novel implant concept using shape memory alloys.

Authors:  Michael Werner; Niels Hammer; Christian Rotsch; Isabell Berthold; Mario Leimert
Journal:  Med Biol Eng Comput       Date:  2019-11-18       Impact factor: 2.602

7.  Anatomical research on strength of screw track fixation in novel cortical bone trajectory for osteoporosis lumbar spine.

Authors:  Paerhati Rexiti; Gulixian Aierken; Shuiquan Wang; Tuerhongjiang Abudurexiti; Nueraihemaiti Abuduwali; Qiang Deng; Hailong Guo; Weibin Sheng
Journal:  Am J Transl Res       Date:  2019-11-15       Impact factor: 4.060

8.  The Importance of Surface Technology in Spinal Fusion.

Authors:  Yoshihiro Katsuura; Joshua Wright-Chisem; Adam Wright-Chisem; Sohrab Virk; Steven McAnany
Journal:  HSS J       Date:  2020-02-12

9.  Cortical threaded pedicle screw improves fatigue strength in decreased bone quality.

Authors:  Lukas Weiser; Kay Sellenschloh; Klaus Püschel; Michael M Morlock; Lennart Viezens; Wolfgang Lehmann; Gerd Huber
Journal:  Eur Spine J       Date:  2020-09-17       Impact factor: 3.134

10.  Cement Augmentation in Sacroiliac Screw Fixation Offers Modest Biomechanical Advantages in a Cadaver Model.

Authors:  Georg Osterhoff; Andrew E Dodd; Florence Unno; Angus Wong; Shahram Amiri; Kelly A Lefaivre; Pierre Guy
Journal:  Clin Orthop Relat Res       Date:  2016-06-22       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.