Jamal J Derakhshan1, Sherif G Nour, Jeffrey L Sunshine, Mark A Griswold, Jeffrey L Duerk.
Abstract
Resolution enhanced T(1)-insensitive steady-state imaging (RE-TOSSI) is a new MRI pulse sequence for the generation of rapid T(2) contrast with high spatial resolution. TOSSI provides T(2) contrast by using nonequally spaced inversion pulses throughout a balanced steady-state free precession (SSFP) acquisition. In RE-TOSSI, these energy and time intensive adiabatic inversion pulses and associated magnetization preparation are removed from TOSSI after acquisition of the data around the center of k-space. Magnetization evolution simulations demonstrate T(2) contrast in TOSSI as well as reduction in the widening of the point spread function width (by up to a factor of 4) to a near ideal case for RE-TOSSI. Phantom experimentation is used to characterize and compare the contrast and spatial resolution properties of TOSSI, RE-TOSSI, balanced SSFP, Half-Fourier Acquisition Single-Shot Turbo Spin Echo (HASTE), and turbo spin echo and to optimize the fraction of k-space acquired using TOSSI. Comparison images in the abdomen and brain demonstrate similar contrast and improved spatial resolution in RE-TOSSI compared with TOSSI; comparison balanced SSFP, HASTE, and turbo spin echo images are provided. RE-TOSSI is capable of providing high spatial resolution T(2)-weighted images in 1 s or less per image.Entities: