Use of ethylene glycol to evaluate gradient performance in gradient-intensive diffusion MR sequences.

William M Spees1, Sheng-Kwei Song, Joel R Garbow, Jeffrey J Neil, Joseph J H Ackerman.   

Abstract

Imaging a phantom of known dimensions is a widely used and simple method for calibrating MRI gradient strength. However, full-range characterization of gradient response is not achievable using this approach. Measurement of the apparent diffusion coefficient of a liquid with known diffusivity allows for calibration of gradient amplitudes across a wider dynamic range. An important caveat is that the temperature dependence of the liquid's diffusion characteristics must be known, and the temperature of the calibration phantom must be recorded. In this report, we demonstrate that the diffusion coefficient of ethylene glycol is well described by Arrhenius-type behavior across the typical range of ambient MRI magnet temperatures. Because of ethylene glycol's utility as an NMR chemical-shift thermometer, the same (1)H MR spectroscopy measurements that are used for gradient calibration also simultaneously "report" the sample temperature. The high viscosity of ethylene glycol makes it well-suited for assessing gradient performance in demanding diffusion-weighted imaging and spectroscopy sequences.
Copyright © 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22127787      PMCID: PMC3296827          DOI: 10.1002/mrm.23201

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  10 in total

Review 1.  Diffusion NMR spectroscopy.

Authors:  K Nicolay; K P Braun; R A Graaf; R M Dijkhuizen; M J Kruiskamp
Journal:  NMR Biomed       Date:  2001-04       Impact factor: 4.044

2.  Strategies for diagnosing and alleviating artifactual attenuation associated with large gradient pulses in PGSE NMR diffusion measurements.

Authors:  W S Price; K Hayamizu; H Ide; Y Arata
Journal:  J Magn Reson       Date:  1999-08       Impact factor: 2.229

3.  Oscillating gradient measurements of water diffusion in normal and globally ischemic rat brain.

Authors:  Mark D Does; Edward C Parsons; John C Gore
Journal:  Magn Reson Med       Date:  2003-02       Impact factor: 4.668

4.  Acceptance testing of magnetic resonance imaging systems: report of AAPM Nuclear Magnetic Resonance Task Group No. 6.

Authors:  J G Och; G D Clarke; W T Sobol; C W Rosen; S K Mun
Journal:  Med Phys       Date:  1992 Jan-Feb       Impact factor: 4.071

Review 5.  Characterization of tissue structure at varying length scales using temporal diffusion spectroscopy.

Authors:  John C Gore; Junzhong Xu; Daniel C Colvin; Thomas E Yankeelov; Edward C Parsons; Mark D Does
Journal:  NMR Biomed       Date:  2010-08       Impact factor: 4.044

6.  On the nature of the NAA diffusion attenuated MR signal in the central nervous system.

Authors:  Christopher D Kroenke; Joseph J H Ackerman; Dmitriy A Yablonskiy
Journal:  Magn Reson Med       Date:  2004-11       Impact factor: 4.668

Review 7.  Theoretical models of the diffusion weighted MR signal.

Authors:  Dmitriy A Yablonskiy; Alexander L Sukstanskii
Journal:  NMR Biomed       Date:  2010-08       Impact factor: 4.044

8.  Non-mono-exponential attenuation of water and N-acetyl aspartate signals due to diffusion in brain tissue.

Authors:  Y Assaf; Y Cohen
Journal:  J Magn Reson       Date:  1998-03       Impact factor: 2.229

9.  Test liquids for quantitative MRI measurements of self-diffusion coefficient in vivo.

Authors:  P S Tofts; D Lloyd; C A Clark; G J Barker; G J Parker; P McConville; C Baldock; J M Pope
Journal:  Magn Reson Med       Date:  2000-03       Impact factor: 4.668

10.  A method for improving the performance of gradient systems for diffusion-weighted MRI.

Authors:  Zoltan Nagy; Nikolaus Weiskopf; Daniel C Alexander; Ralf Deichmann
Journal:  Magn Reson Med       Date:  2007-10       Impact factor: 4.668

  10 in total
  5 in total

1.  Multi-system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice-water phantom.

Authors:  Dariya Malyarenko; Craig J Galbán; Frank J Londy; Charles R Meyer; Timothy D Johnson; Alnawaz Rehemtulla; Brian D Ross; Thomas L Chenevert
Journal:  J Magn Reson Imaging       Date:  2012-09-28       Impact factor: 4.813

Review 2.  Physical and numerical phantoms for the validation of brain microstructural MRI: A cookbook.

Authors:  Els Fieremans; Hong-Hsi Lee
Journal:  Neuroimage       Date:  2018-06-18       Impact factor: 6.556

3.  Temperature and concentration calibration of aqueous polyvinylpyrrolidone (PVP) solutions for isotropic diffusion MRI phantoms.

Authors:  Friedrich Wagner; Frederik B Laun; Tristan A Kuder; Anna Mlynarska; Florian Maier; Jonas Faust; Kerstin Demberg; Linus Lindemann; Boris Rivkin; Armin M Nagel; Mark E Ladd; Klaus Maier-Hein; Sebastian Bickelhaupt; Michael Bach
Journal:  PLoS One       Date:  2017-06-19       Impact factor: 3.240

4.  The role of structural dynamics in the thermal adaptation of hyperthermophilic enzymes.

Authors:  Giuliana Fusco; Francesco Bemporad; Fabrizio Chiti; Christopher M Dobson; Alfonso De Simone
Journal:  Front Mol Biosci       Date:  2022-09-07

5.  Efficient gradient calibration based on diffusion MRI.

Authors:  Irvin Teh; Mahon L Maguire; Jürgen E Schneider
Journal:  Magn Reson Med       Date:  2016-01-08       Impact factor: 4.668

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.