Literature DB >> 22127375

Stereotactic core biopsy: Comparison of 11 gauge with 8 gauge vacuum assisted breast biopsy.

Shambhavi Venkataraman1, Vandana Dialani, Hannah L Gilmore, Tejas S Mehta.   

Abstract

PURPOSE: The compare the performance and ability to obtain a correct diagnosis on needle biopsy between 11 gauge and 8 gauge vacuum assisted biopsy devices.
MATERIALS AND METHODS: Hospital records of all consecutive stereotactic core biopsies performed over five years were retrospectively reviewed in compliance Health Insurance Portability and Accountability Act (HIPPA) policy and with approval from the hospital institutional review board (IRB). Pathology from core biopsy was compared with surgical pathology and/or imaging follow-up. A histological underestimation was defined if the surgical excision yielded a higher grade on pathology which changed management.
RESULTS: 828 needle core biopsies (47.5%, 393/828 with 11 gauge and 52.5%, 435/828 with 8 gauge) yielded 471 benign, 153 high risk and 204 malignant lesions. 30/193 (15.5%) 11 gauge lesions and 16/185 (8.6%) 8 gauge lesions demonstrated higher grade pathology on surgical excision. The difference in the rates of the number of correct diagnoses on core needle biopsy between 11 gauge (363/393, 92.4%) and 8 gauge (419/435, 96.3%) based on either surgical or clinical/imaging follow up and the difference in the number of discordant benign core biopsies between 11 (17/217, 7.8%) and 8 gauge (4/254, 1.6%) necessitating a surgical biopsy was significant (P=0.013; P=0.001). Although there were more underestimations with the 11 gauge (25/193, 13.0%) than 8 gauge (15/185, 8.1%) needle, this was not significant.
CONCLUSION: Our study demonstrates improved performance and increased diagnostic ability of 8 gauge needle over 11 gauge in obtaining a correct diagnosis on needle biopsy.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Mesh:

Year:  2011        PMID: 22127375     DOI: 10.1016/j.ejrad.2011.10.027

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  9 in total

1.  A variable stiffness mechanism for steerable percutaneous instruments: integration in a needle.

Authors:  Iris De Falco; Costanza Culmone; Arianna Menciassi; Jenny Dankelman; John J van den Dobbelsteen
Journal:  Med Biol Eng Comput       Date:  2018-06-04       Impact factor: 2.602

2.  Optimizing Design With Extensive Simulation Data: A Case Study of Designing a Vacuum-Assisted Biopsy Tool.

Authors:  Chi-Lun Lin; Dane Coffey; Daniel Keefe; Arthur Erdman
Journal:  J Med Device       Date:  2018-05-04       Impact factor: 0.582

3.  Mammotome(®) and EnCor (®): comparison of two systems for stereotactic vacuum-assisted core biopsy in the characterisation of suspicious mammographic microcalcifications alone.

Authors:  Giovanna Mariscotti; Manuela Durando; Mattia Robella; Francesca Angelino; Elisa Regini; Pier Paolo Campanino; Marco Belletti; Silvia Osano; Laura Bergamasco; Paolo Fonio; Giovanni Gandini
Journal:  Radiol Med       Date:  2014-08-15       Impact factor: 3.469

4.  A System for Optimizing Medical Device Development Using Finite Element Analysis Predictions.

Authors:  Chi-Lun Lin; Ashutosh Srivastava; Dane Coffey; Daniel Keefe; Marc Horner; Mark Swenson; Arthur Erdman
Journal:  J Med Device       Date:  2014-04-28       Impact factor: 0.582

5.  Value of stereotactic 11-gauge vacuum-assisted breast biopsy in non-palpable suspicious calcifications: an eight-year single institution experience with 587 patients.

Authors:  Shiping Li; Feilin Qu; Yinlong Yang; Lei Wang; Juping Shen; Zhiming Shao
Journal:  Gland Surg       Date:  2020-10

6.  Subsequent Marking under Ultrasound Guidance of Vacuum-Assisted Breast Biopsy Areas after Receipt of Histology: A Feasibility Study of a New Technique.

Authors:  Clara Park; Frauke Chevalier; Volker Möbus; Petra Hoedl; Kerstin Engelmann; Stephan Falk; Doris Leithner; Benjamin Kaltenbach; Thomas J Vogl; Markus Müller-Schimpfle
Journal:  Breast Care (Basel)       Date:  2020-02-21       Impact factor: 2.860

7.  Core needle biopsy of breast cancer tumors increases distant metastases in a mouse model.

Authors:  Edward Gitau Mathenge; Cheryl Ann Dean; Derek Clements; Ahmad Vaghar-Kashani; Steffany Photopoulos; Krysta Mila Coyle; Michael Giacomantonio; Benjamin Malueth; Anna Nunokawa; Julie Jordan; John D Lewis; Shashi Ashok Gujar; Paola Marcato; Patrick W K Lee; Carman Anthony Giacomantonio
Journal:  Neoplasia       Date:  2014-11-20       Impact factor: 5.715

8.  A ten-year, single-center experience: Concordance between breast core needle biopsy/vacuum-assisted biopsy and postoperative histopathology in B3 and B5a cases.

Authors:  Mohamed Elsharkawy; Thomas Vestring; Hans-Juergen Raatschen
Journal:  PLoS One       Date:  2020-05-21       Impact factor: 3.240

9.  Stereotactic breast biopsies: Radiological-pathological concordance in a South African referral unit.

Authors:  Natasha Alexander; Ilana Viljoen; Susan Lucas
Journal:  SA J Radiol       Date:  2022-08-26
  9 in total