| Literature DB >> 22110977 |
Liyun Guo1, Andrew M Kwarciak, Russell Rodriguez, Nilanjan Sarkar, W Mark Richter.
Abstract
This paper describes the design and validation of the OptiPush Biofeedback System, a commercially available, instrumented wheel system that records handrim biomechanics and provides stroke-by-stroke biofeedback and targeting for 11 propulsion variables. Testing of the system revealed accurate measurement of wheel angle (0.02% error), wheel speed (0.06% error), and handrim loads. The maximum errors in static force and torque measurements were 3.80% and 2.05%, respectively. Measured forces were also found to be highly linear (0.985 < slope < 1.011) and highly correlated to the reference forces (r(2) > .998). Dynamic measurements of planar forces (F(x) and F(y)) and axle torque also had low error (-0.96 N to 0.83 N for force and 0.10 Nm to 0.14 Nm for torque) and were highly correlated (r > .986) with expected force and torque values. Overall, the OptiPush Biofeedback System provides accurate measurement of wheel dynamics and handrim biomechanics and may be a useful tool for improving manual wheelchair propulsion.Entities:
Year: 2011 PMID: 22110977 PMCID: PMC3196933 DOI: 10.1155/2011/590780
Source DB: PubMed Journal: Rehabil Res Pract ISSN: 2090-2867
Figure 1The components (a) and assembly (b) of the OptiPush wheel.
Figure 2Definition of the stroke cycle, push phase, and coast phase.
Description of biofeedback variables.
| Variable | Description (Units) |
|---|---|
| Braking torque | Maximum negative (braking) torque about the axle within the stroke (Nm) |
| Cadence | Push frequency (pushes/min) |
| Coast time | Time between handrim release and subsequent handrim contact (s) |
| Contact angle | Angle through which wheel rotates when hand is contact with handrim (degrees) |
| Impact | Maximum rate of force applied to the handrim (N/s) |
| Peak force | Maximum total force during the stroke, where total force = √( |
| Peak torque | Maximum torque about the axle during the stroke (Nm) |
| Power output | Power generated by applying a torque about the axle for a given contact angle (W) |
| Push distance | Distance travelled during the stroke (m) |
| Smoothness | Mean force divided by peak force, unit-less |
| Speed | Mean speed during the stroke (m/s) |
N: Newton; m: meter; W: Watts.
Figure 3Example of the push-by-push biofeedback display.
Figure 4Centrifugal force (F ) and gravitation (F ) force of the test mass applied to the handrim during dynamic testing, where θ is the wheel angle, ω is the angular velocity of the wheel, r is the radius of the handrim, and α is the angle of the mass with respect to the load cell coordinate system.
Validation of F measurements.
| Wheel angle (degrees) | Actual force = 23.28 N | Actual force = 68.04 N | Actual force = 109.99 N | |||
|---|---|---|---|---|---|---|
| Measured force (N) | Error (%) | Measured force (N) | Error (%) | Measured force (N) | Error (%) | |
| 0 | 23.11 ± 1.59 | −0.74 | 66.55 ± 1.65 | −2.19 | 108.68 ± 1.47 | −1.19 |
| 45 | 23.75 ± 1.66 | 2.02 | 67.05 ± 1.74 | −1.45 | 109.17 ± 1.47 | −0.75 |
| 90 | 22.94 ± 1.21 | −1.48 | 66.26 ± 1.53 | −2.62 | 108.25 ± 1.35 | −1.59 |
| 135 | 24.04 ± 1.63 | 3.24 | 67.39 ± 1.35 | −0.96 | 109.45 ± 1.44 | −0.49 |
| 180 | 23.85 ± 1.47 | 2.43 | 67.08 ± 1.50 | −1.42 | 109.23 ± 1.70 | −0.69 |
| 225 | 23.38 ± 1.68 | 0.41 | 66.65 ± 1.75 | −2.05 | 108.92 ± 1.59 | −0.97 |
| 270 | 22.75 ± 1.34 | −2.28 | 65.84 ± 1.42 | −3.23 | 107.78 ± 1.41 | −2.01 |
| 315 | 22.40 ± 1.89 | −3.80 | 65.51 ± 1.64 | −3.72 | 107.43 ± 1.57 | −2.33 |
Measured values are mean ± 1 standard deviation.
Validation of F measurements.
| Actual force (N) | Measured force (N) | Error (%) |
|---|---|---|
| 23.28 | 23.31 ± 3.20 | 0.10 |
| 68.04 | 66.49 ± 3.74 | −2.28 |
| 109.99 | 112.66 ± 3.16 | 2.42 |
Measured values are mean ± 1 standard deviation.
Validation of torque measurements.
| Variable | Actual torque (Nm) | Measured torque (Nm) | Error (%) |
|---|---|---|---|
| √( | 6.03 | 5.99 ± 0.09 | −0.56 |
| 17.61 | 17.45 ± 0.10 | −0.89 | |
|
| 6.03 | 6.08 ± 0.07 | 0.87 |
| 17.61 | 17.97 ± 0.07 | 2.04 |
Measured values are mean ± 1 standard deviation.
Results of dynamic testing.
| Test Mass = 1.17 kg | Test Mass = 2.30 kg | ||||
|---|---|---|---|---|---|
| Variable | Speed (m/s) | Correlation coefficient | Measured value-actual value | Correlation coefficient | Measured value-actual value |
|
| 0.5 | 0.989 | −0.04 ± 1.20 | 0.998 | 0.11 ± 1.31 |
| 1.0 | 0.994 | 0.31 ± 0.93 | 0.998 | 0.83 ± 1.22 | |
| 1.5 | 0.994 | 0.42 ± 0.92 | 0.997 | 0.69 ± 1.36 | |
|
| 0.5 | 0.988 | −0.20 ± 1.32 | 0.997 | −0.28 ± 1.25 |
| 1.0 | 0.993 | −0.60 ± 1.03 | 0.998 | −0.96 ± 1.03 | |
| 1.5 | 0.993 | −0.81 ± 1.16 | 0.995 | −0.60 ± 1.55 | |
|
| 0.5 | 0.999 | 0.13 ± 0.08 | 0.999 | 0.14 ± 0.11 |
| 1.0 | 0.997 | 0.13 ± 0.17 | 0.998 | 0.14 ± 0.26 | |
| 1.5 | 0.986 | 0.11 ± 0.35 | 0.998 | 0.10 ± 0.25 | |
Differences in the values are mean ± 1 standard deviation.