Qian-Feng Xia1, Yang-An Wen, Ping Liu, Pu Li, Jin-Bo Liu, Xi Qin, Shi-Yun Qian, Zhi-Guang Tu.
Abstract
BACKGROUND: The duplex mutation primers offer many advantages over other multi-labeled probes for real-time detection of amplification products.Entities:
Keywords: Duplex mutation primers; HCV; Hepatitis; PCR; Quantification
Year: 2011 PMID: 22087189 PMCID: PMC3212762
Source DB: PubMed Journal: Hepat Mon ISSN: 1735-143X Impact factor: 0.660
Figure 1Fluorescent curves of the standard dilution series.
Limit of the HCV RNA real-time PCR assay
| 100 (8/8) | |
| 100 (8/8) | |
| 100 (8/8) | |
| 87.5 (7/8) | |
| 50 (4/8) | |
| 37.5 (3/8) | |
| 12.5 (1/8) | |
| 0 | |
| 0 |
Study of intra and inter-run variability of the HCV RNA real-time PCR assay to determine its precision
| 2.14×102 | 2.36×102 | 2.24×102 | 2.01×102 | 1.95×102 | 2.14×102 | 1.67×101 | 7.80 | |
| 1.39×103 | 1.21×103 | 1.24×103 | 1.31×103 | 1.19×103 | 1.27×103 | 8.2×101 | 6.46 | |
| 5.17×104 | 5.36×104 | 5.25×104 | 5.19×104 | 5.12×104 | 5.22×104 | 9.2×102 | 1.76 | |
| 7.45×107 | 7.37×107 | 7.39×107 | 7.41×107 | 7.31×107 | 7.39×107 | 5.2×105 | 0.70 | |
| N | N | N | N | N | - | - | - | |
| 2.44×102 | 2.31×102 | 2.12×102 | 2.19×102 | 1.85×102 | 2.18×102 | 2.22×101 | 10.18 | |
| 1.39×103 | 1.32×103 | 1.20×103 | 1.27×103 | 1.15×103 | 1.27×103 | 9.50×101 | 7.48 | |
| 5.17×104 | 5.24×104 | 5.09×104 | 5.31×104 | 5.37×104 | 5.24×104 | 1.11×103 | 2.12 | |
| 7.51×107 | 7.37×107 | 7.30×107 | 7.21×107 | 7.47×107 | 7.37×107 | 1.23×105 | 1.67 | |
| N | N | N | N | N | - | - | - | |
a N: negative
Figure 2Graph shin correlation of the HCV real-time PCR ass av results with the new method results for 92 HCV-positive clinical spcimens (numerical values e not listed). Thre was excellent correlation (r 0. 95) between the two assays for all samples.