Literature DB >> 22074705

A comparison of the Ex-PRESS™ mini glaucoma shunt with standard trabeculectomy in the surgical treatment of glaucoma.

Lionel Marzette1, Leon W Herndon.   

Abstract

BACKGROUND AND
OBJECTIVE: To compare the success and complication rates of patients with glaucoma who had an Ex-PRESS mini glaucoma shunt device implantation (Optonol, Ltd., Neve Ilan, Israel) to those who had conventional trabeculectomy. PATIENTS AND METHODS: The records of 76 eyes of 69 consecutive subjects who had Ex-PRESS implants and 77 eyes of 65 consecutive controls who had trabeculectomy procedures were reviewed. All surgeries were performed by one of the authors (LWH). Success was defined as an intraocular pressure (IOP) between 5 and 21 mm Hg in patients who did not require further glaucoma surgery in the eye of note.
RESULTS: The difference in the percentage of cases of postoperative hypotony between the standard trabeculectomy group (16%) and the Ex-PRESS group (4%) was statistically significant (P = .023).
CONCLUSION: The Ex-PRESS device is at least as effective as the standard trabeculectomy in lowering the IOP of patients with glaucoma, with a significantly lower risk of postoperative hypotony. The data further suggest that the Ex-PRESS device results in an overall greater percentage reduction in IOP than with trabeculectomy, although this did not reach statistical significance. Copyright 2011, SLACK Incorporated.

Entities:  

Mesh:

Year:  2011        PMID: 22074705     DOI: 10.3928/15428877-20111017-03

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  24 in total

1.  Imaging of filtering blebs after implantation of the Ex-PRESS shunt with the use of the Visante optical coherence tomography.

Authors:  Aristeidis Konstantinidis; Georgios D Panos; Magdalini Triantafylla; Georgios Labiris; Efthimia Tsaragli; Zisis Gatzioufas; Vassilios Kozobolis
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

2.  Ex-PRESS implantation versus trabeculectomy in Chinese patients with POAG: fellow eye pilot study.

Authors:  Wei Wang; Min-Wen Zhou; Wen-Bin Huang; Xin-Bo Gao; Xiu-Lan Zhang
Journal:  Int J Ophthalmol       Date:  2017-01-18       Impact factor: 1.779

3.  Combined phacoemulsification with ExPRESS shunt in angle closure glaucoma.

Authors:  Aparna Rao
Journal:  Digit J Ophthalmol       Date:  2015-03-21

Review 4.  Long-term outcomes and risk factors for failure with the EX-press glaucoma drainage device.

Authors:  C Mariotti; E Dahan; M Nicolai; L Levitz; S Bouee
Journal:  Eye (Lond)       Date:  2013-11-15       Impact factor: 3.775

5.  Efficacy and economic analysis of Ex-PRESS implantation versus trabeculectomy in uncontrolled glaucoma: a systematic review and Meta-analysis.

Authors:  Ling Wang; Fang Sha; Da-Dong Guo; Hong-Sheng Bi; Jun-Kang Si; Yu-Xiang Du; Kai Tang
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

6.  Ex-PRESS implantation versus trabeculectomy in uncontrolled glaucoma: a meta-analysis.

Authors:  Wei Wang; Minwen Zhou; Wenbin Huang; Xiulan Zhang
Journal:  PLoS One       Date:  2013-05-31       Impact factor: 3.240

Review 7.  Standard Trabeculectomy and Ex-PRESS Miniature Glaucoma Shunt: A Comparative Study and Literature Review.

Authors:  Elad Moisseiev; Eran Zunz; Rotem Tzur; Shimon Kurtz; Gabi Shemesh
Journal:  J Glaucoma       Date:  2015-08       Impact factor: 2.503

8.  Evaluation of ExPress glaucoma filtration device in Indian patients with advanced glaucoma.

Authors:  Dewang Angmo; Reetika Sharma; Shreyas Temkar; Tanuj Dada
Journal:  Indian J Ophthalmol       Date:  2015-05       Impact factor: 1.848

9.  Postoperative Changes in Aqueous Monocyte Chemotactic Protein-1 Levels and Bleb Morphology after Trabeculectomy vs. Ex-PRESS Shunt Surgery.

Authors:  Kohei Shobayashi; Toshihiro Inoue; Motofumi Kawai; Keiichiro Iwao; Saori Ohira; Sachi Kojima; Utako Kuroda; Kei-Ichi Nakashima; Hidenobu Tanihara
Journal:  PLoS One       Date:  2015-10-01       Impact factor: 3.240

Review 10.  Device-modified trabeculectomy for glaucoma.

Authors:  Xue Wang; Rabeea Khan; Anne Coleman
Journal:  Cochrane Database Syst Rev       Date:  2015-12-01
View more