Literature DB >> 22074115

Sensitive detection and accurate monitoring of Plasmodium vivax parasites on routine complete blood count using automatic blood cell analyzer (DxH800(TM)).

H K Lee1, S I Kim, H Chae, M Kim, J Lim, E J Oh, Y Kim, Y J Park, W Lee, K Han.   

Abstract

INTRODUCTION: Plasmodium vivax malaria is one of the most important infectious diseases plaguing humanity and causes significant mortality and morbidity worldwide. The gold standard of P. vivax malaria diagnosis is the microscopy of blood smears. Although microscopy is a rapid, cost-effective, and readily applicable method, it has many disadvantages, including low sensitivity, specificity, and precision. Therefore, there is a clear need for an effective screening test for P. vivax malaria detection both in high-prevalence areas and developed countries.
METHODS: A total of 1761 complete blood count (CBC) samples generated by the automated hematology analyzer (DxH 800™; Beckman Coulter Inc., Miami, FL, USA) were retrospectively analyzed. The sample pool contained 123 samples from 52 P. vivax malaria patients and 1504 nonmalarial samples including 509 patients with leukopenia (white blood cell <2000/μL) and 134 normal subjects.
RESULTS: The P. vivax malaria samples exhibited easily recognizable typical malaria signals on the nucleated red blood cell (nRBC) plots (sensitivity 100%) in DxH 800™. All 1504 samples without P. vivax infection were negative for malaria signal (specificity 100%). The size of P. vivax malaria signals correlated roughly with the parasite burden.
CONCLUSION: DxH800™ provides very sensitive and specific, easily recognizable P. vivax malaria signals on routine CBC without need for the additional reagents or special procedures.
© 2011 Blackwell Publishing Ltd.

Entities:  

Mesh:

Year:  2011        PMID: 22074115     DOI: 10.1111/j.1751-553X.2011.01383.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  5 in total

1.  Detection of intracellular parasites by use of the CellaVision DM96 analyzer during routine screening of peripheral blood smears.

Authors:  Lori D Racsa; Rita M Gander; Paul M Southern; Erin McElvania TeKippe; Christopher Doern; Hung S Luu
Journal:  J Clin Microbiol       Date:  2014-11-05       Impact factor: 5.948

2.  Automated hematology analyzers: Recent trends and applications.

Authors:  Gaurav Chhabra
Journal:  J Lab Physicians       Date:  2018 Jan-Mar

3.  The Leukocyte VCS Parameters Compared with Procalcitonin, Interleukin-6, and Soluble Hemoglobin Scavenger Receptor sCD163 for Prediction of Sepsis in Patients with Cirrhosis.

Authors:  Fan Guo; Yang-Chun Feng; Gang Zhao; Hui-Li Wu; Ling Xu; Jing Zhao; Jie Lv; Song-Tao Han; Yan-Chun Huang; Xiu-Min Ma
Journal:  Dis Markers       Date:  2019-12-12       Impact factor: 3.434

4.  Novel markers of early neutrophilic and monocytic engraftment after hematopoietic stem cell transplantation.

Authors:  Jimin Kahng; Seung-Ah Yahng; Jae Wook Lee; Yonggoo Kim; Myungshin Kim; Eun-Jee Oh; Yeon-Joon Park; Jong Wook Lee; Bin Cho; Kyungja Han
Journal:  Ann Lab Med       Date:  2014-02-13       Impact factor: 3.464

5.  Incidental Identification of Plasmodium vivax During Routine Complete Blood Count Analysis Using the UniCel DxH 800.

Authors:  Soyoung Shin; Sun Hee Park; Joonhong Park
Journal:  Ann Lab Med       Date:  2018-03       Impact factor: 3.464

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.