Variation of noise in multi-run functional MRI using generalized autocalibrating partially parallel acquisition (GRAPPA).

Hu Cheng1.   

Abstract

PURPOSE: To investigate the noise variation in multi-run functional MRI (fMRI) scans using generalized autocalibrating partially parallel acquisition (GRAPPA), with a focus on the cause of this variation.
MATERIALS AND METHODS: A phantom was continuously scanned for 10 runs using echo-planar imaging (EPI) combined with GRAPPA to simulate a multi-run fMRI exam. The variation of noise between runs was examined for different GRAPPA acceleration factors. The noise variation was also evaluated in a real fMRI experiment with human subjects at an acceleration factor of two. The cause of noise variation was explored by offline reconstruction using different GRAPPA weights and numerical simulation of GRAPPA reference scans.
RESULTS: It was found that the noise distribution in the image is stable within a run but may vary randomly from run to run. The variation of noise was also observed in fMRI experiments with human subjects. The variation can be significantly reduced if all the images from individual runs are reconstructed using the same reference scan data.
CONCLUSION: Both phantom experiments and human data showed that the noise pattern may change in different fMRI runs. The variation is mainly caused by the random noise in separate reference scans for GRAPPA in each run.
Copyright © 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 22069162     DOI: 10.1002/jmri.22891

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  The renaissance of functional 18F-FDG PET brain activation imaging.

Authors:  Antoine Verger; Eric Guedj
Journal:  Eur J Nucl Med Mol Imaging       Date:  2018-09-29       Impact factor: 9.236

2.  Reduction of across-run variability of temporal SNR in accelerated EPI time-series data through FLEET-based robust autocalibration.

Authors:  Anna I Blazejewska; Himanshu Bhat; Lawrence L Wald; Jonathan R Polimeni
Journal:  Neuroimage       Date:  2017-02-20       Impact factor: 6.556

3.  Improvement of temporal signal-to-noise ratio of GRAPPA accelerated echo planar imaging using a FLASH based calibration scan.

Authors:  S Lalith Talagala; Joelle E Sarlls; Siyuan Liu; Souheil J Inati
Journal:  Magn Reson Med       Date:  2015-07-20       Impact factor: 4.668

4.  Metabolic connectivity of resting-state networks in alpha synucleinopathies, from prodromal to dementia phase.

Authors:  Cecilia Boccalini; Elisa Bortolin; Giulia Carli; Andrea Pilotto; Andrea Galbiati; Alessandro Padovani; Luigi Ferini-Strambi; Daniela Perani
Journal:  Front Neurosci       Date:  2022-08-08       Impact factor: 5.152

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.