Repeatability and reproducibility of anterior chamber volume measurements using 3-dimensional corneal and anterior segment optical coherence tomography.

Shinichi Fukuda1, Keisuke Kawana, Yoshiaki Yasuno, Tetsuro Oshika.   

Abstract

PURPOSE: To evaluate the repeatability and reproducibility of anterior chamber volume (ACV) measurements using swept-source 3-dimensional corneal and anterior segment optical coherence tomography (CAS-OCT) and dual Scheimpflug imaging.
SETTING: Department of Ophthalmology, Institute of Clinical Medicine, University of Tsukuba, Ibaraki, Japan.
DESIGN: Nonrandomized clinical trial.
METHODS: Measurements were taken in normal eyes (subject group) and in eyes with primary angle closure (PAC) (patient group). In the subject group, the entire ACV and the central 8.0 mm diameter ACV were measured using CAS-OCT and dual Scheimpflug imaging. In the patient group, the entire ACV and 8.0 mm ACV were measured using CAS-OCT. The coefficient of variation and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility, and the correlation between the 2 devices was assessed.
RESULTS: In the subject group, the mean 8.0 mm ACV was 110.14 mm(3) ± 12.57 (SD) using CAS-OCT and 114.51 ± 14.69 mm(3) using Scheimpflug imaging; there was a significant linear correlation (r = 0.878, P < .0001). The mean entire ACV on CAS-OCT was 165.15 ± 29.29 mm(3). The ICCs of the 8.0 mm and entire ACV measurements were greater than 0.94. The coefficients of repeatability and reproducibility of the 8.0 mm ACV and entire ACV measurements were less than 5%. In the patient group, the 8.0 mm and entire ACV measurements showed good reproducibility and repeatability.
CONCLUSIONS: The CAS-OCT method allowed noninvasive measurement of the entire ACV with sufficient repeatability and reproducibility. The 8.0 mm ACV measurements with CAS-OCT and Scheimpflug imaging were comparable.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21333870     DOI: 10.1016/j.jcrs.2010.08.053

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  11 in total

1.  Ocular biocompatibility of nitinol intraocular clips.

Authors:  Jeffrey L Olson; Raul Velez-Montoya; Michael Erlanger
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-01-25       Impact factor: 4.799

2.  Quantitative analysis of the intraretinal layers and optic nerve head using ultra-high resolution optical coherence tomography.

Authors:  Yuhong Wang; Hong Jiang; Meixiao Shen; Byron L Lam; Delia Cabrera DeBuc; Yufeng Ye; Ming Li; Aizhu Tao; Yilei Shao; Jianhua Wang
Journal:  J Biomed Opt       Date:  2012-06       Impact factor: 3.170

3.  Long-term results of treatment with diquafosol ophthalmic solution for aqueous-deficient dry eye.

Authors:  Shizuka Koh; Chikako Ikeda; Yoshihiro Takai; Hitoshi Watanabe; Naoyuki Maeda; Kohji Nishida
Journal:  Jpn J Ophthalmol       Date:  2013-06-06       Impact factor: 2.447

4.  Dynamic imaging of accommodation by swept-source anterior segment optical coherence tomography.

Authors:  Alberto Neri; Marco Ruggeri; Alessandra Protti; Rosachiara Leaci; Stefano A Gandolfi; Claudio Macaluso
Journal:  J Cataract Refract Surg       Date:  2015-02-20       Impact factor: 3.351

5.  Repeatability of ocular measurements with a dual-Scheimpflug analyzer in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Lorena Fuentes-Broto; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-04-24       Impact factor: 3.411

6.  Comparison of anterior segment measurements obtained by three different devices in healthy eyes.

Authors:  Carmen Lopez de la Fuente; Ana Sanchez-Cano; Francisco Segura; Isabel Pinilla
Journal:  Biomed Res Int       Date:  2014-06-02       Impact factor: 3.411

7.  Repeatability and Reproducibility of Anterior Segment Measurements in Normal Eyes Using Dual Scheimpflug Analyzer.

Authors:  Zeynep Altıparmak; Ramazan Yağcı; Emre Güler; Zeynel Arslanyılmaz; Metin Canbal; İbrahim F Hepşen
Journal:  Turk J Ophthalmol       Date:  2015-12-05

8.  Reproducibility of scleral spur identification and angle measurements using fourier domain anterior segment optical coherence tomography.

Authors:  Ricardo J Cumba; Sunita Radhakrishnan; Nicholas P Bell; Kundandeep S Nagi; Alice Z Chuang; Shan C Lin; Kimberly A Mankiewicz; Robert M Feldman
Journal:  J Ophthalmol       Date:  2012-11-01       Impact factor: 1.909

Review 9.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

10.  Topographic characteristics after Descemet's membrane endothelial keratoplasty and Descemet's stripping automated endothelial keratoplasty.

Authors:  Takahiko Hayashi; Takefumi Yamaguchi; Kentaro Yuda; Naoko Kato; Yoshiyuki Satake; Jun Shimazaki
Journal:  PLoS One       Date:  2017-11-30       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.