Literature DB >> 22052018

Influence of examiner experience on the reproducibility of retinal nerve fiber thickness values using Cirrus and Stratus OCTs.

Javier Moreno-Montañés1, Natalia Olmo, Noelia García, Aurora Alvarez, Marta García-Granero.   

Abstract

PURPOSE: The purpose of the study was to evaluate the influence of the examiner's experience on the reproducibility of retinal nerve fiber layer (RNFL) measurements obtained with Cirrus optical coherence tomography (OCT) and Stratus.
METHODS: Fifty-one normal and glaucomatous eyes of 51 participants were included. Two examiners (1 novice, 1 experienced) obtained 2 scans using both OCTs. For quantitative measurements, Bland and Altman limits of agreement were evaluated. For qualitative classifications, κ coefficients were calculated.
RESULTS: Signal strength was higher with Cirrus than with Stratus (P<0.05). Signal strength was higher in scans performed by the experienced examiner than in those carried out by the inexperienced examiner in Stratus but not in Cirrus. RNFL measurement differences between Cirrus and Stratus were influenced by the examiner for the inferior (P=0.02), superior (P<0.001), and temporal quadrants (P=0.009). The RNFL quantitative agreement of examiners was higher in Cirrus than in Stratus. The qualitative agreement (κ coefficients) of both examiners in the RNFL classification were almost perfect with Cirrus (in the average, superior, and inferior quadrants), and moderate with Stratus (only in average and inferior quadrant).
CONCLUSIONS: The signal strength is independent of the examiner's experience in Cirrus but not in Stratus. RNFL measurements obtained by both examiners were more reproducible with Cirrus than with Stratus. The differences in RNFL measurements between both OCTs were related to the examiner's experience in all 3 quadrants. Agreement between operators in the RNFL classification was higher with Cirrus than Stratus.

Entities:  

Mesh:

Year:  2013        PMID: 22052018     DOI: 10.1097/IJG.0b013e318237c878

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  2 in total

1.  Retinal Nerve Fibre Layer Thickness Increases with Decreasing Spectralis OCT Signal Strength in Normal Eyes.

Authors:  Margaret R Strampe; Luai Eldweik; Benjamin C Chaon; Sarah Maki; Tyler Wieland; Celine Satija; Collin McClelland; Michael S Lee
Journal:  Neuroophthalmology       Date:  2019-10-29

2.  Determination of axial length requiring adjustment of measured circumpapillary retinal nerve fiber layer thickness for ocular magnification.

Authors:  Kazunori Hirasawa; Nobuyuki Shoji; Yukako Yoshii; Shota Haraguchi
Journal:  PLoS One       Date:  2014-09-12       Impact factor: 3.240

  2 in total