Evaluation of radiation during EVAR performed on a mobile C-arm.

B Maurel1, J Sobocinski, P Perini, M Guillou, M Midulla, R Azzaoui, S Haulon.   

Abstract

BACKGROUND: The aim of this study was to evaluate radiation exposure during aortic endovascular aneurysm repair (EVAR) on a mobile C-arm using a low dose and pulse mode.
METHODS: We performed a retrospective analysis of a prospectively maintained database on patients undergoing EVAR. Indirect dose measurements of dose area product (DAP, mGy m²) calculated by the C-arm (OEC 9900MD), fluoroscopic time (FT), type of procedure, contrast media volume and body mass index were analysed. To confirm the correlation between direct and indirect DAP measurements, direct dose was measured with radiochromic films on a sample of 15 patients. Film grey level response was calibrated according to a reference dose measurement performed with a calibrated dosimeter. DAP and peak skin dose (PSD, Gy) were measured on each film. Correlation between DAP from direct and indirect measures, and between DAP and PSD, were analysed.
RESULTS: From January 2009 to April 2011, 335 patients underwent EVAR. Complete data were available on 301 procedures including 188 bifurcated, 54 fenestrated, 28 thoracic, 20 branched and 11 aorto-uni-iliac endografts implantation. The respective median FT and DAP was 9.36 min (1.8-67) and 3 mGy m(2) (0.4-28); 27.2 min (2-69) and 7.3 mGy m(2) (1.2-29); 7.75 min (1.2-19.1) and 2 mGy m(2) (0.3-11); 42.98 min (2.4-95.4) and 15.95 mGy m(2) (2.98-77.7); 6.2 min (0.5-36.3) and 2 mGy m(2) (0.3-11). Direct DAP measurement on radiochromic films was strongly correlated with DAP values provided by the C-arm (r = 0.98). PSD correlated weakly with DAP. DAP was significantly increased (p < 0.001) in patients with a body mass index >30. Contrast media volume was significantly increased in the branched endograft group.
CONCLUSION: Indirect DAP values measured by the C-arm are accurate to evaluate radiation exposure. Compared to the literature, our values for standard procedures are significantly decreased by the usage of low dose and pulse mode. DAP for fenestrated and branched procedures was comparable to published DAP values with standard procedures using a regular fluoroscopic mode.
Copyright © 2011 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22047911     DOI: 10.1016/j.ejvs.2011.09.017

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  9 in total

1.  Radiation Exposure in Endovascular Infra-Renal Aortic Aneurysm Repair and Factors that Influence It.

Authors:  Rui Machado; Vitor Miguel Dias Ferreira; Luis Loureiro; João Gonçalves; Pedro Oliveira; Rui Almeida
Journal:  Braz J Cardiovasc Surg       Date:  2016 Nov-Dec

Review 2.  Physician and Patient Radiation Exposure During Endovascular Procedures.

Authors:  Andrew M Goldsweig; J Dawn Abbott; Herbert D Aronow
Journal:  Curr Treat Options Cardiovasc Med       Date:  2017-02

3.  Investigation of reference levels and radiation dose associated with abdominal EVAR (endovascular aneurysm repair) procedures across several European Centres.

Authors:  E Tuthill; L O'Hora; M O'Donohoe; S Panci; P Gilligan; D Campion; R Trenti; E Fox; D Catania; L Rainford
Journal:  Eur Radiol       Date:  2017-05-18       Impact factor: 5.315

4.  Cumulative radiation dose and radiation risk from medical imaging in patients subjected to endovascular aortic aneurysm repair.

Authors:  Marco Brambilla; Paolo Cerini; Domenico Lizio; Luca Vigna; Alessandro Carriero; Rita Fossaceca
Journal:  Radiol Med       Date:  2015-01-09       Impact factor: 3.469

5.  Three-dimensional fusion computed tomography decreases radiation exposure, procedure time, and contrast use during fenestrated endovascular aortic repair.

Authors:  Michael M McNally; Salvatore T Scali; Robert J Feezor; Daniel Neal; Thomas S Huber; Adam W Beck
Journal:  J Vasc Surg       Date:  2014-08-28       Impact factor: 4.268

6.  Efficacy Analysis of a Script-based Guide for EVAR Execution: is it Possible to Reduce Patient Exposure to Contrast, Operative Time and Blood Loss even when Advanced Technologies are not Available?

Authors:  Giovani José Dal Poggetto Molinari; Ana Terezinha Guillaumon; Andréia Marques de Oliveira Dalbem
Journal:  Braz J Cardiovasc Surg       Date:  2015 Nov-Dec

7.  Mechanical thrombectomy using a combined CT/C-arm X-ray system.

Authors:  Johannes Pfaff; Christian Herweh; Mirko Pham; Silvia Schönenberger; Julian Bösel; Peter A Ringleb; Sabine Heiland; Martin Bendszus; Markus Möhlenbruch
Journal:  J Neurointerv Surg       Date:  2015-05-02       Impact factor: 5.836

8.  Hybrid operating room applications in the increasingly complex endovascular era: the trump card of modern vascular surgery.

Authors:  Davide Esposito; Francesco Gonfiantini; Aaron Thomas Fargion; Walter Dorigo; Flavio Villani; Rossella Di Domenico; Sara Speziali; Carlo Pratesi
Journal:  Ann Surg Treat Res       Date:  2020-12-30       Impact factor: 1.859

9.  Radiation-Induced DNA Damage in Operators Performing Endovascular Aortic Repair.

Authors:  Tamer El-Sayed; Ashish S Patel; Jun S Cho; James A Kelly; Francesca E Ludwinski; Prakash Saha; Oliver T Lyons; Alberto Smith; Bijan Modarai
Journal:  Circulation       Date:  2017-10-20       Impact factor: 39.918

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.