Literature DB >> 22047202

Comparison of long-term biocompability of PVDF and PP meshes.

C D Klink1, K Junge, M Binnebösel, H P Alizai, J Otto, U P Neumann, U Klinge.   

Abstract

BACKGROUND: Abdominal hernia repair is the most frequently performed operation in surgery. Mesh repair in hernia surgery has become an integral component. Although meshes made of PVDF are already in clinical use, so far no data of long-term biocompability are available.
METHODS: In this study a PVDF mesh was compared to a polypropylene mesh with regard to its long-term biocompatibility. A total of 28 rats were randomized to two groups. Mesh material was implanted subcutaneously; animals were euthanized seven days and six months postoperatively. The quantity of inflammatory tissue response was characterized by measuring the diameter of the foreign body granuloma. Furthermore quality of cellular immune response (T-lymphocytes, macrophages, and neutrophils), and inflammation (COX-2) was analyzed by immunohistochemistry. Furthermore the collagen type I/III ratio was determined.
RESULTS: Macrophages, T-lymphocytes, neutrophiles, and COX-2 declined significantly up to six months postoperatively in comparison to day 7 for both PVDF and PP meshes, and in both groups the collagen ratio increased significantly in the course of time. PVDF meshes showed a foreign body granuloma size significantly reduced compared to PP (7 days: 20 ± 2 μm vs. 27 ± 2 μm; 6 months 15 ± 2 μm vs. 22 ± 3 μm; p < .001). However no significant differences were found analyzing cellular response six months postoperatively.
CONCLUSIONS: Our current data suggest that even in the long-term course after six months and despite a higher effective surface of the PVDF samples it showed a smaller foreign body granuloma than with PP whereas the cellular response was similar.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22047202     DOI: 10.3109/08941939.2011.589883

Source DB:  PubMed          Journal:  J Invest Surg        ISSN: 0894-1939            Impact factor:   2.533


  15 in total

1.  Mesh implants: An overview of crucial mesh parameters.

Authors:  Lei-Ming Zhu; Philipp Schuster; Uwe Klinge
Journal:  World J Gastrointest Surg       Date:  2015-10-27

2.  In vivo MRI visualization of parastomal mesh in a porcine model.

Authors:  J Otto; D Busch; C Klink; A Ciritsis; A Woitok; C Kuhl; U Klinge; U P Neumann; N A Kraemer; J Conze
Journal:  Hernia       Date:  2014-06-10       Impact factor: 4.739

Review 3.  Evidence-Based Hernia Treatment in Adults.

Authors:  Dieter Berger
Journal:  Dtsch Arztebl Int       Date:  2016-03-04       Impact factor: 5.594

4.  Effectiveness of Prophylactic Intraperitoneal Mesh Implantation for Prevention of Incisional Hernia in Patients Undergoing Open Abdominal Surgery: A Randomized Clinical Trial.

Authors:  Andreas Kohler; Joel L Lavanchy; Ursina Lenoir; Anita Kurmann; Daniel Candinas; Guido Beldi
Journal:  JAMA Surg       Date:  2019-02-01       Impact factor: 14.766

5.  Coated meshes for hernia repair provide comparable intraperitoneal adhesion prevention.

Authors:  Marc H F Schreinemacher; Kevin W Y van Barneveld; Rieky E G Dikmans; Marion J J Gijbels; Jan-Willem M Greve; Nicole D Bouvy
Journal:  Surg Endosc       Date:  2013-06-08       Impact factor: 4.584

6.  The early outcomes of complex abdominal wall reconstruction with polyvinylidene (PVDF) mesh in the setting of active infection: a prospective series.

Authors:  Claudio Birolini; Eduardo Yassushi Tanaka; Jocielle Santos de Miranda; Abel Hiroshi Murakami; Sergio Henrique Bastos Damous; Edivaldo Massazo Utiyama
Journal:  Langenbecks Arch Surg       Date:  2022-07-29       Impact factor: 2.895

7.  Influence of gentamicin-coded PVDF suture material on the healing of intestinal anastomosis in a rat model.

Authors:  Dominik S Schoeb; Christian D Klink; Andreas Lambertz; Roman Eickhoff; Daniel Busch; Tom F Ulmer; Ulf P Neumann; Marcel Binnebösel
Journal:  Int J Colorectal Dis       Date:  2015-08-12       Impact factor: 2.571

8.  International guidelines for groin hernia management.

Authors: 
Journal:  Hernia       Date:  2018-01-12       Impact factor: 4.739

9.  TOT 8/4: A Way to Standardize the Surgical Procedure of a Transobturator Tape.

Authors:  Sebastian Ludwig; Martin Stumm; Peter Mallmann; Wolfram Jager
Journal:  Biomed Res Int       Date:  2016-02-15       Impact factor: 3.411

10.  3D non-woven polyvinylidene fluoride scaffolds: fibre cross section and texturizing patterns have impact on growth of mesenchymal stromal cells.

Authors:  Anne Schellenberg; Robin Ross; Giulio Abagnale; Sylvia Joussen; Philipp Schuster; Annahit Arshi; Norbert Pallua; Stefan Jockenhoevel; Thomas Gries; Wolfgang Wagner
Journal:  PLoS One       Date:  2014-04-11       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.