Multistage evaluation and commissioning of a pre-calibrated, single-use OneDosePlus MOSFET system for in vivo dosimetry in a radiotherapy department.

A P M Lonsdale1.   

Abstract

OBJECTIVES: The aim was to determine the agreement between absorbed dose measurements using the OneDosePlus and treatment planning system calculations in ideal circumstances, minimising patient-related uncertainties, before deciding upon action levels.
METHODS: A OneDose metal oxide semiconductor field-effect transistor (Sicel Technologies, Morrisville, NC) in vivo dosimetry system was subjected to a multistage evaluation that tested standard and non-standard field conditions related to treatment planning system calculations for an anthropomorphic phantom.
RESULTS: The system was found to perform within manufacturer specifications. Batch uniformity was found to be within specification when measured using a method described by Halvorsen. A modification used to assess statistical distribution of response showed an increase in the value of two standard deviations (2 σ) from ± 2.3 to ± 2.9%, which was still within the manufacturer-stated value of ± 5%. The tests using the anthropomorphic phantom also emphasised the fact that patient density inhomogeneities in the region of the D(max) point will affect the dose calculated by the treatment planning system and delivered to the patient.
CONCLUSION: The OneDose system does not account for these inhomogeneities, leading to dependence in the deviation between expected and reported dose on inhomogeneity and choice of calculation algorithm.

Entities:  

Mesh:

Year:  2011        PMID: 22045952      PMCID: PMC3486649          DOI: 10.1259/bjr/21460681

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  2 in total

1.  Dosimetric evaluation of a new design MOSFET in vivo dosimeter.

Authors:  Per H Halvorsen
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

2.  Clinical application of the OneDose Patient Dosimetry System for total body irradiation.

Authors:  S Best; A Ralston; N Suchowerska
Journal:  Phys Med Biol       Date:  2005-12-06       Impact factor: 3.609

  2 in total
  1 in total

1.  Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams.

Authors:  A Sathish Kumar; S D Sharma; B Paul Ravindran
Journal:  J Med Phys       Date:  2014-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.