Accuracy of a portable glucose meter and of a Continuous Glucose Monitoring device used at home by patients with type 1 diabetes.

M P Francescato1, M Geat, G Stel, S Cauci.   

Abstract

BACKGROUND: Patients with diabetes are recommended to self-monitor their blood glucose levels also at home. Accuracy of a hand-held glucometer and a Continuous Glucose Monitoring (CGM) device were comparatively evaluated.
METHODS: Venous blood samples (for reference laboratory determinations; n=428) were collected from 18 type 1 patients (35-65 years old), immediately followed by capillary measurement (Bayer ContourLink meter) and CGM readings (Medtronic Paradigm).
RESULTS: Laboratory values did not differ statistically from ContourLink and CGM readings, mean difference (±SD) being -0.05±1.06 mmol/L and 0.10±1.84 mmol/L glucose, respectively. A bias ((value-reference)/reference×100) ≥15% was observed in 27.7% and 54.9% of cases, respectively. Notably, below 3.9 mmol/L glucose (hypoglycemic threshold), an absolute error>0.8 mmol/L was found in 78.9% and 94.1% of cases. The absolute errors of the CGM device were inversely related to the rate of glucose change (r=0.598, p<0.001).
CONCLUSIONS: A very large error was observed at the extreme glycemic values, which may lead to erroneous therapy. Consequently, performance of future portable glucometers should be focused in particular under hypo- and hyper-glycemia. Moreover, integrated CGM devices should not disregard the effect of the rate of blood glucose change on the sensor readings.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22032826     DOI: 10.1016/j.cca.2011.10.012

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  5 in total

1.  Referrals for Hypoglycemia to the Pediatric Endocrine Clinic: Is It For Real?

Authors:  David W Hansen; Erica A Eugster
Journal:  Clin Pediatr (Phila)       Date:  2018-09-10       Impact factor: 1.168

2.  Pilot study of a prototype minimally invasive intradermal continuous glucose monitor.

Authors:  Ruth S Weinstock; Suzan Bristol; Andrew Armenia; A Chris Gesswein; B Wayne Bequette; John P Willis
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

3.  Effect of Continuous Glucose Monitoring Accuracy on Clinicians' Retrospective Decision Making in Diabetes: A Pilot Study.

Authors:  Zeinab Mahmoudi; Mette Dencker Johansen; Hanne Holdflod Nørgaard; Steen Andersen; Ulrik Pedersen-Bjergaard; Lise Tarnow; Jens Sandahl Christiansen; Ole Hejlesen
Journal:  J Diabetes Sci Technol       Date:  2015-06-08

4.  Comparative Accuracy of 17 Point-of-Care Glucose Meters.

Authors:  Laya Ekhlaspour; Debbie Mondesir; Norman Lautsch; Courtney Balliro; Mallory Hillard; Kendra Magyar; Laura Goergen Radocchia; Aryan Esmaeili; Manasi Sinha; Steven J Russell
Journal:  J Diabetes Sci Technol       Date:  2016-10-03

5.  The role of continuous glucose monitoring in the care of children with type 1 diabetes.

Authors:  Noelle S Larson; Jordan E Pinsker
Journal:  Int J Pediatr Endocrinol       Date:  2013-03-26
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.