Literature DB >> 22027339

Comparison of automated multiplexed bead-based ANA screening assay with ELISA for detecting five common anti-extractable nuclear antigens and anti-dsDNA in systemic rheumatic diseases.

Yoonjung Kim1, Yongjung Park, Eun Young Lee, Hyon-Suk Kim.   

Abstract

BACKGROUND: A newly developed and totally automated Luminex-based assay, the BioPlex™ 2200 system, is able to detect various autoantibodies simultaneously from a single sample. We compared the BioPlex™ 2200 system with ELISA for the detection of six autoantibodies.
METHODS: A total of 127 serum samples from the patients with systemic rheumatic diseases were collected and assayed with the BioPlex™ 2200 system (Bio-Rad, USA) and conventional ELISA (INOVA Diagnostics, USA) for 5 anti-extractable nuclear antigens. Additionally, relative sensitivity of the BioPlex™ 2200 system for detecting anti-dsDNA was evaluated with 79 specimens from SLE patients, which were positive for anti-dsDNA by ELISA.
RESULTS: The concordance rates between ELISA and the BioPlex ranged from 88.1% for anti-RNP to 95.2% for anti-Scl-70, and the kappa coefficients between the results by the two assays were from 0.48 to 0.67. Among the 79 anti-dsDNA positive specimens by ELISA, seventy-eight (98.7%) showed positive results for anti-dsDNA by the BioPlex.
CONCLUSIONS: The BioPlex™ 2200 system showed comparable results with those by conventional ELISA for detecting autoantibodies, and this automated assay could measure multifarious autoantibodies concurrently in a single sample. It could be effectively used in clinical laboratories for screening autoimmune diseases.
Copyright © 2011 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22027339     DOI: 10.1016/j.cca.2011.10.017

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  The clinical utility of anti-chromatin antibodies as measured by BioPlex 2200 in the diagnosis of systemic lupus erythematosus versus other rheumatic diseases.

Authors:  Nilanjana Bose; Xiaofeng Wang; Majula Gupta; Qingping Yao
Journal:  Int J Clin Exp Med       Date:  2012-08-22

2.  A comparison of anti-nuclear antibody quantification using automated enzyme immunoassays and immunofluorescence assays.

Authors:  Renata Baronaite; Merete Engelhart; Troels Mørk Hansen; Gorm Thamsborg; Hanne Slott Jensen; Steen Stender; Pal Bela Szecsi
Journal:  Autoimmune Dis       Date:  2014-01-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.