Usage and effectiveness of the low glucose suspend feature of the Medtronic Paradigm Veo insulin pump.

Pratik Agrawal1, John B Welsh, Brian Kannard, Sina Askari, Qingqing Yang, Francine R Kaufman.   

Abstract

BACKGROUND: Sensor-augmented insulin pumps may be programmed to suspend insulin delivery in response to hypoglycemia. The Medtronic Paradigm® Veo™ pump with automatic low glucose suspend (LGS) was released in 2009. Data from 7 months of real-world use of the system were analyzed to assess usage patterns and effectiveness of LGS.
METHOD: Data from 935 patients totaling 49,867 patient days were collected; the LGS feature was on for 82% of these days. A subset of 278 subjects who used the pump for ≥3 months was analyzed separately; these subjects provided 28,401 patient days of data, with LGS used for 92% of the time.
RESULTS: The LGS threshold was most commonly set between 50 and 60 mg/dl. A total of 27,216 LGS events occurred, and 60% began in the afternoon or evening. The median duration of LGS events was 9.87 min, 45% lasted for <5 min, and 11% lasted for >115 min (equivalent to the full extent of the LGS event between 115 and 120 min). Among the episodes lasting for >115 min, the mean sensor glucose (SG) was 58.8 ± 12.4 mg/dl at LGS activation (time 0), rose to 102.2 ± 52.8 mg/dl by the end of the LGS episode (when insulin delivery was automatically resumed), and was 150.1 ± 68.6 mg/dl at 240 min. In the 278-subject subgroup, LGS usage significantly reduced the number of SG readings <50 mg/dl (p = 0.001) and >300 mg/dl (p = 0.001).
CONCLUSIONS: The LGS feature was on for most of the patient days in the study. Most LGS episodes lasted for <10 min. Use of the LGS feature significantly reduced exposure to hypoglycemia. Profound hyperglycemia resulting from LGS episodes lasting >115 min was not observed.
© 2011 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22027306      PMCID: PMC3208869          DOI: 10.1177/193229681100500514

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  7 in total

1.  Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes.

Authors:  Richard M Bergenstal; William V Tamborlane; Andrew Ahmann; John B Buse; George Dailey; Stephen N Davis; Carol Joyce; Tim Peoples; Bruce A Perkins; John B Welsh; Steven M Willi; Michael A Wood
Journal:  N Engl J Med       Date:  2010-06-29       Impact factor: 91.245

2.  Parental attitudes towards overnight closed-loop glucose control in children with type 1 diabetes.

Authors:  Daniela Elleri; Carlo L Acerini; Janet M Allen; Josephine Hayes; Claire Pesterfield; Malgorzata E Wilinska; David B Dunger; Roman Hovorka
Journal:  Diabetes Technol Ther       Date:  2010-01       Impact factor: 6.118

3.  Confirmation of hypoglycemia in the "dead-in-bed" syndrome, as captured by a retrospective continuous glucose monitoring system.

Authors:  Robert J Tanenberg; Christopher A Newton; Almond J Drake
Journal:  Endocr Pract       Date:  2010 Mar-Apr       Impact factor: 3.443

4.  Prolonged nocturnal hypoglycemia is common during 12 months of continuous glucose monitoring in children and adults with type 1 diabetes.

Authors: 
Journal:  Diabetes Care       Date:  2010-03-03       Impact factor: 19.112

5.  Preventing hypoglycemia using predictive alarm algorithms and insulin pump suspension.

Authors:  Bruce Buckingham; Erin Cobry; Paula Clinton; Victoria Gage; Kimberly Caswell; Elizabeth Kunselman; Fraser Cameron; H Peter Chase
Journal:  Diabetes Technol Ther       Date:  2009-02       Impact factor: 6.118

6.  Is an automatic pump suspension feature safe for children with type 1 diabetes? An exploratory analysis with a closed-loop system.

Authors:  Eda Cengiz; Karena L Swan; William V Tamborlane; Garry M Steil; Amy T Steffen; Stuart A Weinzimer
Journal:  Diabetes Technol Ther       Date:  2009-04       Impact factor: 6.118

7.  Insulin pump therapy with automated insulin suspension in response to hypoglycemia: reduction in nocturnal hypoglycemia in those at greatest risk.

Authors:  Pratik Choudhary; John Shin; Yongyin Wang; Mark L Evans; Peter J Hammond; David Kerr; James A M Shaw; John C Pickup; Stephanie A Amiel
Journal:  Diabetes Care       Date:  2011-09       Impact factor: 19.112

  7 in total
  30 in total

Review 1.  Sensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes.

Authors:  Isabelle Steineck; Ajenthen Ranjan; Kirsten Nørgaard; Signe Schmidt
Journal:  J Diabetes Sci Technol       Date:  2016-10-06

2.  Putting brakes on insulin pump infusion to prevent hypoglycemia.

Authors:  Eda Cengiz
Journal:  J Diabetes Sci Technol       Date:  2011-09-01

3.  Improved outcomes from diabetes monitoring: the benefits of better adherence, therapy adjustments, patient education, and telemedicine support.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2012-05-01

4.  Hypoglycemia early alarm systems based on recursive autoregressive partial least squares models.

Authors:  Elif Seyma Bayrak; Kamuran Turksoy; Ali Cinar; Lauretta Quinn; Elizabeth Littlejohn; Derrick Rollins
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

Review 5.  Hypo- and Hyperglycemic Alarms: Devices and Algorithms.

Authors:  Daniel Howsmon; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2015-04-30

Review 6.  Multivariable Adaptive Artificial Pancreas System in Type 1 Diabetes.

Authors:  Ali Cinar
Journal:  Curr Diab Rep       Date:  2017-08-15       Impact factor: 4.810

7.  Where Do We Stand with Closed-Loop Systems and Their Challenges?

Authors:  Melanie Jackson; Jessica R Castle
Journal:  Diabetes Technol Ther       Date:  2020-05-22       Impact factor: 6.118

8.  Inpatient studies of a Kalman-filter-based predictive pump shutoff algorithm.

Authors:  Fraser Cameron; Darrell M Wilson; Bruce A Buckingham; Hasmik Arzumanyan; Paula Clinton; H Peter Chase; John Lum; David M Maahs; Peter M Calhoun; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

9.  Hypoglycemia Detection and Carbohydrate Suggestion in an Artificial Pancreas.

Authors:  Kamuran Turksoy; Jennifer Kilkus; Iman Hajizadeh; Sediqeh Samadi; Jianyuan Feng; Mert Sevil; Caterina Lazaro; Nicole Frantz; Elizabeth Littlejohn; Ali Cinar
Journal:  J Diabetes Sci Technol       Date:  2016-11-01

10.  Evaluation of a novel continuous glucose monitoring-based method for mealtime insulin dosing--the iBolus--in subjects with type 1 diabetes using continuous subcutaneous insulin infusion therapy: a randomized controlled trial.

Authors:  Paolo Rossetti; F Javier Ampudia-Blasco; Alejandro Laguna; Ana Revert; Josep Vehì; Juan F Ascaso; Jorge Bondia
Journal:  Diabetes Technol Ther       Date:  2012-09-24       Impact factor: 6.118

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.