Safety and efficacy of 24-h closed-loop insulin delivery in well-controlled pregnant women with type 1 diabetes: a randomized crossover case series.

Helen R Murphy1, Kavita Kumareswaran, Daniela Elleri, Janet M Allen, Karen Caldwell, Martina Biagioni, David Simmons, David B Dunger, Marianna Nodale, Malgorzata E Wilinska, Stephanie A Amiel, Roman Hovorka.   

Abstract

OBJECTIVE: To evaluate the safety and efficacy of closed-loop insulin delivery in well-controlled pregnant women with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII). RESEARCH DESIGN AND METHODS: A total of 12 women with type 1 diabetes (aged 32.9 years, diabetes duration 17.6 years, BMI 27.1 kg/m(2), and HbA(1c) 6.4%) were randomly allocated to closed-loop or conventional CSII. They performed normal daily activities (standardized meals, snacks, and exercise) for 24 h on two occasions at 19 and 23 weeks' gestation. Plasma glucose time in target (63-140 mg/dL) and time spent hypoglycemic were calculated.
RESULTS: Plasma glucose time in target was comparable for closed-loop and conventional CSII (median [interquartile range]: 81 [59-87] vs. 81% [54-90]; P = 0.75). Less time was spent hypoglycemic (<45 mg/dL [0.0 vs. 0.3%]; P = 0.04), with a lower low blood glucose index (2.4 [0.9-3.5] vs. 3.3 [1.9-5.1]; P = 0.03), during closed-loop insulin delivery.
CONCLUSIONS: Closed-loop insulin delivery was as effective as conventional CSII, with less time spent in extreme hypoglycemia.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22011408      PMCID: PMC3220861          DOI: 10.2337/dc11-1430

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


There is strong evidence that avoidance of hyperglycemia is key to improved pregnancy outcomes in type 1 diabetes. Currently, the price to pay for tight glucose control is increased risk of severe hypoglycemia, causing significant maternal morbidity (seizures, road-traffic accidents, and death) (1). Despite the increased use of insulin pumps and fast-acting insulin analogs, continuous glucose monitoring (CGM) highlights the prevalence of hypoglycemia exposure (3 h per day at <70 mg/dL and 1 h per day at <50 mg/dL) during pregnancy (2). Although the neonatal consequences of maternal hypoglycemia are unclear, it is accepted that the benefits of tight glycemic control must be balanced against the potential risk of hypoglycemia (3). Closed-loop systems use computerized algorithms to link insulin delivery with CGM glucose levels in real time, aiming to more closely approximate normal glucose concentrations. Overnight studies in children and adults indicate the potential for reduced nocturnal hypoglycemia (4,5). We previously documented the effectiveness of overnight closed-loop deliver in early- and late-gestation type 1 diabetic pregnancy, with 84–100% time in target without nocturnal hypoglycemia during sedentary conditions (6). The aim of this study was to evaluate the safety and efficacy of 24-h closed-loop insulin delivery, incorporating normal daily activities and exercise.

RESEARCH DESIGN AND METHODS

From April 2010 to April 2011, 12 pregnant women with type 1 diabetes from two U.K. antenatal clinics (Addenbrooke’s Hospital, Cambridge, U.K., and Kings College Hospital, London, U.K.) were recruited into a randomized crossover trial of closed-loop insulin delivery. The same study protocol comparing closed-loop and conventional continuous subcutaneous insulin infusion (CSII) was applied for two 24-h visits, separated by a 1- to 6-week interval. Study protocols were approved by the research ethics committee, and all participants provided written informed consent.

Study devices and procedures

A FreeStyle Navigator sensor (Abbott Diabetes Care, Alameda, CA) was inserted the day before each study. An intravenous sampling catheter and study pump (Animas 2020; Johnson & Johnson, New Brunswick, NJ) were inserted on arrival (1200 h). After lunch (50 g carbohydrate), an Actiheart physical activity energy expenditure (PAEE) monitor (CamNtech, Cambridge, U.K.) was attached (7). From 1400 h, venous samples were obtained every 15–30 min until the study ended at 1230 h on day 2. Plasma glucose concentrations were measured immediately (YSI 2300 STAT Plus Analyzer; Farnborough, U.K.), with plasma extracted for later insulin concentration measurements by immunochemiluminometric assay (Invitron, Monmouth, U.K.).

Daily activities

Activities included three 20-min walks (1400, 1930, and 0900 h) and two 50-min sessions of brisk treadmill walking (1500 h on day 1 and 0930 h on day 2). Meal and snack choices were consistent between visits. Pre-exercise snacks were given according to capillary glucose measurements (15 g carbohydrate >108 mg/dL and 30 g carbohydrate ≤108 mg/dL). An additional 15-g carbohydrate snack was provided at 2100 h.

Insulin delivery

During closed-loop, basal insulin infusion rates were manually adjusted at 15-min intervals according to CGM glucose levels and the control algorithm advice. During conventional CSII, the women set temporary basal rates and used correction boluses according to capillary glucose measurements (7–10 per day). During both visits, insulin boluses were calculated by women according to capillary glucose levels, aiming for the National Institute for Health and Clinical Excellence (NICE)-recommended target glucose range of 63–140 mg/dL (8).

Hypoglycemia

Hypoglycemia was defined as plasma glucose levels ≤54 mg/dL with symptoms or ≤45 mg/dL without symptoms. Episodes were treated with 15 g oral carbohydrate (90 mL Lucozade Energy Original; GlaxoSmithKline, Middlesex, England, U.K.).

Statistical analysis

The primary outcome was plasma glucose time in target (63–140 mg/dL) from 1400 h on day 1 to 1230 h on day 2. Secondary outcomes were time spent above and below target, mean glucose concentration, glucose SD, and low blood glucose index (LBGI). The Wilcoxon signed rank test was used to compare paired measurements within an individual between closed-loop and CSII visits. Values are given as medians (interquartile ranges), unless otherwise stated. Analyses were conducted using SPSS version 15 (SPSS, Chicago, IL).

RESULTS

Participants (n = 12) had a median age of 32.9 years (30.4–36.7), diabetes duration of 17.6 years (8.0–27.3), CSII duration of 2.0 years (0.8–2.0), weight of 77.0 kg (68.5–84.6), BMI of 27.1 kg/m2 (25.3–30.8), and HbA1c of 6.4% (6.1–6.6). A total of 11 women started CSII preconception, and 7 were primaparous. Studies were performed at 19 weeks’ (16–25) and 23 weeks’ (20–28) gestation, with a between-visit interval of 27 days (17–34). The primary and secondary outcome data are shown in Table 1, with details of the glycemic control achieved shown in the Supplementary Data. There was comparable median [interquartile range] plasma glucose time in target between closed-loop and CSII visits (81 [59-88] vs. 81% [54-90]; P = 0.75).
Table 1

Primary and secondary outcomes during closed-loop and conventional CSII

Closed-loop deliveryCSIIP
Primary outcome
 Plasma glucose time in target
  Percentage of time in target (63–40 mg/dL)81 (59–88)81 (54–90)0.75
Secondary outcomes
 Plasma glucose outcomes
  Mean plasma glucose (mg/dL)108 (99–121)104 (92–108)0.35
  SD of plasma glucose (mg/dL)25.2 (23.4–37.8)28.8 (25.2–41.4)0.69
 Hypoglycemia
  Percentage of time hypoglycemic <63 mg/dL6.9 (1.3–12)7.5 (3.6–18)0.48
  Percentage of time hypoglycemic ≤50 mg/dL0.6 (0.0–2.1)1.5 (0.0–2.7)0.17
  Percentage of time hypoglycemic ≤45 mg/dL0.0 (0.0–0.2)0.3 (0.0–1.5)0.04
  Symptomatic hypoglycemia817
   Episodes >45–54 mg/dL812
   Episodes >36–45 mg/dL55
   Episodes <36 mg/dL03
  LBGI*2.4 (0.9–3.5)3.3 (1.9–5.1)0.03
 Hyperglycemia
  Percentage of time hyperglycemic >140 mg/dL14.0 (6.8–28)6.7 (5.8–22)0.75
  Percentage of time hyperglycemic ≥180 mg/dL0.2 (0.0–6.0)0.0 (0.0–6.2)0.78
  High blood glucose index0.8 (0.3–1.6)0.4 (0.3–1.4)0.81
 PAEE (kJ/kg)23.4 (19.7–27.0)21.2 (19.0–22.1)0.09
Insulin
 Insulin infusion (units/h)0.7 (0.5–0.9)0.8 (0.5–1.1)0.35
 SD insulin infusion rate0.9 (0.5–1.0)0.2 (0.2–0.5)<0.001
 Plasma insulin concentration120 (101–146)107 (82–145)0.88

Data are median (interquartile range), unless otherwise indicated.

*LBGI assessed the duration and extent of hypoglycemia.

Primary and secondary outcomes during closed-loop and conventional CSII Data are median (interquartile range), unless otherwise indicated. *LBGI assessed the duration and extent of hypoglycemia. There were 13 hypoglycemic episodes (8 symptomatic and 5 asymptomatic) during closed-loop delivery versus 20 hypoglycemic episodes (17 symptomatic and 3 asymptomatic) during CSII, including 3 episodes below 36 mg/dL. During closed-loop delivery, the LBGI was lower (P = 0.03), with less time spent below 45 mg/dL (P = 0.04).

Overnight glucose control

The overnight plasma glucose time in target was strikingly high, with no differences between visits (95 [84-100] vs. 100% [64-100]). However, during conventional CSII, there was significantly less overnight time spent in target by CGM glucose measurements (98 [94-100] vs. 83% [50-100]; P = 0.03). There were no discrepancies between plasma and CGM measurements during any other time period.

CONCLUSIONS

In this cohort of pregnant women with tight glycemic control, closed-loop insulin delivery was as effective as conventional CSII but potentially safer because closed-loop delivery reduced the extent and duration of hypoglycemia. Although there were fewer episodes of hypoglycemia, decreased LBGI, and less time spent below 45 mg/dL, closed-loop delivery could not prevent exercise-related hypoglycemia. Even algorithms incorporating glucagon cannot prevent exercise-related hypoglycemia if there is a rapid glucose reduction, increased insulin on board, or sensor inaccuracy (9,10). The finding that overnight time in target measured by CGM, rather than plasma glucose, favors closed-loop delivery warrants additional consideration. Plasma glucose measurements are impractical for home studies, so establishing the optimal means of assessing overnight closed-loop delivery is important. The strengths of this study include a robust crossover design, standardized meals, and physical activity to approximate a real-life setting. Limitations are the small sample size and that the system was not fully automated. The achievement of 100% overnight time in target provides support for the further investigation of closed-loop systems in pregnancy. Home testing over multiple nights now is required to determine whether the near-optimal overnight glucose control can be translated into longer-term real-life benefits.
  9 in total

1.  Reliability and validity of the combined heart rate and movement sensor Actiheart.

Authors:  S Brage; N Brage; P W Franks; U Ekelund; N J Wareham
Journal:  Eur J Clin Nutr       Date:  2005-04       Impact factor: 4.016

2.  Factors influencing the effectiveness of glucagon for preventing hypoglycemia.

Authors:  Jessica R Castle; Julia M Engle; Joseph El Youssef; Ryan G Massoud; W Kenneth Ward
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

3.  Efficacy determinants of subcutaneous microdose glucagon during closed-loop control.

Authors:  Steven J Russell; Firas H El-Khatib; David M Nathan; Edward R Damiano
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

Review 4.  Maternal hypoglycemia during pregnancy in type 1 diabetes: maternal and fetal consequences.

Authors:  Edith W M T ter Braak; Inge M Evers; D Willem Erkelens; Gerard H A Visser
Journal:  Diabetes Metab Res Rev       Date:  2002 Mar-Apr       Impact factor: 4.876

5.  Risk indicators predictive for severe hypoglycemia during the first trimester of type 1 diabetic pregnancy.

Authors:  Inge M Evers; Edith W M T ter Braak; Harold W de Valk; Bertine van Der Schoot; Nancy Janssen; Gerard H A Visser
Journal:  Diabetes Care       Date:  2002-03       Impact factor: 19.112

6.  Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial.

Authors:  Roman Hovorka; Janet M Allen; Daniela Elleri; Ludovic J Chassin; Julie Harris; Dongyuan Xing; Craig Kollman; Tomas Hovorka; Anne Mette F Larsen; Marianna Nodale; Alessandra De Palma; Malgorzata E Wilinska; Carlo L Acerini; David B Dunger
Journal:  Lancet       Date:  2010-02-04       Impact factor: 79.321

7.  Changes in the glycemic profiles of women with type 1 and type 2 diabetes during pregnancy.

Authors:  Helen R Murphy; Gerry Rayman; Katherine Duffield; Karen S Lewis; Susan Kelly; Balroop Johal; Duncan Fowler; Rosemary C Temple
Journal:  Diabetes Care       Date:  2007-07-31       Impact factor: 19.112

8.  Closed-loop insulin delivery during pregnancy complicated by type 1 diabetes.

Authors:  Helen R Murphy; Daniela Elleri; Janet M Allen; Julie Harris; David Simmons; Gerry Rayman; Rosemary Temple; David B Dunger; Ahmad Haidar; Marianna Nodale; Malgorzata E Wilinska; Roman Hovorka
Journal:  Diabetes Care       Date:  2011-01-07       Impact factor: 19.112

9.  Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies.

Authors:  Roman Hovorka; Kavita Kumareswaran; Julie Harris; Janet M Allen; Daniela Elleri; Dongyuan Xing; Craig Kollman; Marianna Nodale; Helen R Murphy; David B Dunger; Stephanie A Amiel; Simon R Heller; Malgorzata E Wilinska; Mark L Evans
Journal:  BMJ       Date:  2011-04-13
  9 in total
  35 in total

1.  Evaluation of stochastic adjustment for glucose sensor bias during closed-loop insulin delivery.

Authors:  Craig Kollman; Peter Calhoun; John Lum; Werner Sauer; Roy W Beck
Journal:  Diabetes Technol Ther       Date:  2013-11-15       Impact factor: 6.118

2.  The Clinical Case for the Integration of a Ketone Sensor as Part of a Closed Loop Insulin Pump System.

Authors:  Melissa H Lee; Barbora Paldus; Balasubramanium Krishnamurthy; Sybil A McAuley; Rajiv Shah; Alicia J Jenkins; David N O'Neal
Journal:  J Diabetes Sci Technol       Date:  2019-01-10

Review 3.  Different insulin types and regimens for pregnant women with pre-existing diabetes.

Authors:  Sinéad M O'Neill; Louise C Kenny; Ali S Khashan; Helen M West; Rebecca Md Smyth; Patricia M Kearney
Journal:  Cochrane Database Syst Rev       Date:  2017-02-03

4.  Design and in silico evaluation of an intraperitoneal-subcutaneous (IP-SC) artificial pancreas.

Authors:  Justin J Lee; Eyal Dassau; Howard Zisser; Francis J Doyle
Journal:  Comput Chem Eng       Date:  2014-11-05       Impact factor: 3.845

5.  In silico evaluation of an artificial pancreas combining exogenous ultrafast-acting technosphere insulin with zone model predictive control.

Authors:  Justin J Lee; Eyal Dassau; Howard Zisser; Rebecca A Harvey; Lois Jovanovič; Francis J Doyle
Journal:  J Diabetes Sci Technol       Date:  2013-01-01

6.  Multivariable Artificial Pancreas for Various Exercise Types and Intensities.

Authors:  Kamuran Turksoy; Iman Hajizadeh; Nicole Hobbs; Jennifer Kilkus; Elizabeth Littlejohn; Sediqeh Samadi; Jianyuan Feng; Mert Sevil; Caterina Lazaro; Julia Ritthaler; Brooks Hibner; Nancy Devine; Laurie Quinn; Ali Cinar
Journal:  Diabetes Technol Ther       Date:  2018-09-06       Impact factor: 6.118

Review 7.  Artificial Pancreas Project at Cambridge 2013.

Authors:  R Hovorka
Journal:  Diabet Med       Date:  2015-04-15       Impact factor: 4.359

Review 8.  Novel Insulin Delivery Technologies in Women with Pregestational Type 1 Diabetes: A Review of the Literature.

Authors:  Erin Drever; Denice S Feig
Journal:  Obstet Med       Date:  2013-03-01

9.  Evaluating the accuracy and large inaccuracy of two continuous glucose monitoring systems.

Authors:  Lalantha Leelarathna; Marianna Nodale; Janet M Allen; Daniela Elleri; Kavita Kumareswaran; Ahmad Haidar; Karen Caldwell; Malgorzata E Wilinska; Carlo L Acerini; Mark L Evans; Helen R Murphy; David B Dunger; Roman Hovorka
Journal:  Diabetes Technol Ther       Date:  2012-12-20       Impact factor: 6.118

10.  Outpatient glycemic control with a bionic pancreas in type 1 diabetes.

Authors:  Steven J Russell; Firas H El-Khatib; Manasi Sinha; Kendra L Magyar; Katherine McKeon; Laura G Goergen; Courtney Balliro; Mallory A Hillard; David M Nathan; Edward R Damiano
Journal:  N Engl J Med       Date:  2014-06-15       Impact factor: 91.245

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.