Literature DB >> 21987213

MRI-guided biopsy and aspiration in the head and neck: evaluation of 77 patients.

Yubo Lü1, Ming Liu, Chengli Li, Lebin Wu, Jan Fritz.   

Abstract

OBJECTIVES: To evaluate the efficacy and safety of MRI-guided percutaneous biopsy procedures of head and neck lesions using 0.23T open MRI with optical tracking.
METHODS: A retrospective analysis of 77 patients (51 male, 26 female; mean age, 43 years; range, 11-88 years) who underwent MRI-guided percutaneous biopsy of a head and neck lesion was performed. Mean lesion diameter was 3 cm (range, 1-7.8 cm). Rapid gradient echo sequences were used for image guidance. 23/77 lesions were biopsied after intravenous gadolinium. Tissue sampling techniques included needle aspiration (n = 19) and core needle biopsy (n = 58). Outcome variables included technical success, diagnostic accuracy, procedure time and complications.
RESULTS: In all patients, a sufficient amount of tissue for pathological analysis was obtained. Pathological analysis diagnosed 41 malignant lesions and 36 benign lesions. In 42 cases, surgical correlation was available. In 35 cases, the final diagnosis was confirmed by imaging and clinical follow-up. MR-guided biopsy had a sensitivity, specificity, positive predictive value, negative predictive value and accuracy of 93.2%, 100%, 100%, 91.7%, and 96%, respectively. Procedure time was 29 min (range, 15-47 min). No major complications occurred.
CONCLUSIONS: MRI-guided biopsy of head and neck lesions has a high diagnostic performance and is safe in clinical practice. KEY POINTS: • MRI-guided biopsy helps clinicians to assess patients with head&neck masses. • Differention of malignant and benign lesions is possible with 96% accuracy. • The safety profile of MRI-guided biopsy of head&neck lesions is favorable. • MRI guidance enables accurate biopsy without the use of ionizing radiation.

Entities:  

Mesh:

Year:  2011        PMID: 21987213     DOI: 10.1007/s00330-011-2270-8

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  34 in total

1.  [Use of open MRI in head-neck surgery].

Authors:  F Bootz; S Keiner; A Weber; T Schulz; B Scheffler
Journal:  HNO       Date:  1999-12       Impact factor: 1.284

2.  Magnetic resonance imaging guided bone biopsies in the iPath-200 system.

Authors:  Lebin Wu; Chuanting Li; Liguang Chen; Chengli Li; Xiuling Qiu
Journal:  Chin Med J (Engl)       Date:  2003-06       Impact factor: 2.628

3.  MR-guided aspiration cytology in the head and neck at high field strength.

Authors:  G Hathout; R B Lufkin; B Jabour; J Andrews; D Castro
Journal:  J Magn Reson Imaging       Date:  1992 Jan-Feb       Impact factor: 4.813

4.  CT-guided aspirations in the head and neck: assessment of the first 216 cases.

Authors:  Paul M Sherman; David M Yousem; Laurie A Loevner
Journal:  AJNR Am J Neuroradiol       Date:  2004-10       Impact factor: 3.825

5.  Evaluation of MR imaging guided steroid injection of the sacroiliac joints for the treatment of children with refractory enthesitis-related arthritis.

Authors:  J Fritz; N Tzaribachev; C Thomas; J A Carrino; C D Claussen; J S Lewin; P L Pereira
Journal:  Eur Radiol       Date:  2010-10-29       Impact factor: 5.315

Review 6.  [Image-guided diagnostic percutaneous puncture biopsy].

Authors:  D Negru; Corina Ursulescu; V Fotea
Journal:  Rev Med Chir Soc Med Nat Iasi       Date:  2007 Oct-Dec

7.  A new needle and technique for MRI-guided aspiration cytology of the head and neck.

Authors:  T Trapp; R Lufkin; E Abemayor; L Layfield; W Hanafee; P Ward
Journal:  Laryngoscope       Date:  1989-01       Impact factor: 3.325

8.  Magnetic resonance navigation for head and neck lesions.

Authors:  S P Davis; V K Anand; G Dhillon
Journal:  Laryngoscope       Date:  1999-06       Impact factor: 3.325

9.  Percutaneous magnetic resonance image-guided biopsy and aspiration in the head and neck.

Authors:  E M Merkle; J S Lewin; A J Aschoff; D W Stepnick; J L Duerk; C F Lanzieri; M Strauss
Journal:  Laryngoscope       Date:  2000-03       Impact factor: 3.325

10.  [Preoperative marking and biopsy of nonpalpable breast lesions with a guidance system for the open Magnetom].

Authors:  H Sittek; E Linsmeier; C Perlet; P Schneider; C Baudrexel; M Untch; M Reiser
Journal:  Radiologe       Date:  2000-11       Impact factor: 0.635

View more
  4 in total

1.  Assessment of a needle-tracking system for bipolar radiofrequency ablation.

Authors:  Masashi Hirooka; Yohei Koizumi; Yusuke Imai; Hironori Ochi; Yoshiko Nakamura; Osamu Yoshida; Atsushi Hiraoka; Yoshio Tokumoto; Masanori Abe; Yoichi Hiasa
Journal:  J Med Ultrason (2001)       Date:  2015-10-26       Impact factor: 1.314

2.  Percutaneous Image-Guided Cryoablation of Head and Neck Tumors for Local Control, Preservation of Functional Status, and Pain Relief.

Authors:  Jeffrey P Guenette; Kemal Tuncali; Nathan Himes; Paul B Shyn; Thomas C Lee
Journal:  AJR Am J Roentgenol       Date:  2016-11-15       Impact factor: 3.959

3.  Volume navigation with contrast enhanced ultrasound and image fusion for percutaneous interventions: first results.

Authors:  Ernst Michael Jung; Chris Friedrich; Patrick Hoffstetter; Lena Marie Dendl; Frank Klebl; Ayman Agha; Phillipp Wiggermann; Christian Stroszcynski; Andreas Georg Schreyer
Journal:  PLoS One       Date:  2012-03-20       Impact factor: 3.240

4.  Unexpected esophageal diseases appeared in thyroid resections.

Authors:  Liu Ye-huan; Lyu Shi-xu; Zhou Yi-li; Wang Ou-chen; Zhang Xiao-hua
Journal:  World J Surg Oncol       Date:  2015-03-30       Impact factor: 2.754

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.