Literature DB >> 11147325

[Preoperative marking and biopsy of nonpalpable breast lesions with a guidance system for the open Magnetom].

H Sittek1, E Linsmeier, C Perlet, P Schneider, C Baudrexel, M Untch, M Reiser.   

Abstract

PURPOSE: MRI-guided interventional maneuvers in the breast, when closed MRI scanners are employed are relatively complex and time consuming. The purpose of our investigations was to develop a special device for the localization and biopsy of breast lesions using an open low-field MRI Scanner (Magnetom Open, Siemens, Erlangen, Germany) permitting free access to the patient.
MATERIALS AND METHODS: Due to its particular material and construction characteristics, a newly developed device for localization and biopsy of breast lesions is appropriate to remain within the magnetic field during the examination without degrading image quality. We performed 125 tests in a phantom and 31 investigations in patients including 26 wire-localizations and 5 biopsies.
RESULTS: The interventional maneuvers in the phantom could be performed very precisely. In all 125 phantom studies, the needle (14 G coaxial biopsy cannula, Bard) was positioned close to the simulated lesion. In 25 out of 26 patients the needle was positioned within a distance less than 5 mm from the lesion. When using the 14 G coaxial biopsy cannula (Bard, Karlsruhe, Germany) the needle tip was found adjacent next to the lesion in all 12 cases (7 wire-localizations, 5 biopsies).
CONCLUSIONS: Our results indicate that our device allows very precise preoperative localization of breast lesions within the Magnetom Open. Using MRI-compatible, large-core needles, biopsy under direct MRI control resulted in good results.

Entities:  

Mesh:

Year:  2000        PMID: 11147325     DOI: 10.1007/s001170050882

Source DB:  PubMed          Journal:  Radiologe        ISSN: 0033-832X            Impact factor:   0.635


  3 in total

1.  MRI-guided biopsy and aspiration in the head and neck: evaluation of 77 patients.

Authors:  Yubo Lü; Ming Liu; Chengli Li; Lebin Wu; Jan Fritz
Journal:  Eur Radiol       Date:  2011-10-11       Impact factor: 5.315

Review 2.  [Interventional MRI of the breast. Indications, technique, results and perspectives].

Authors:  K Hellerhoff; T Schlossbauer; P Herzog; M Reiser
Journal:  Radiologe       Date:  2008-04       Impact factor: 0.635

3.  Contrast-enhanced breast MR imaging of claustrophobic or oversized patients using an open low-field magnet.

Authors:  M Calabrese; D Brizzi; L Carbonaro; M Chiaramondia; M A Kirchin; F Sardanelli
Journal:  Radiol Med       Date:  2009-02-04       Impact factor: 3.469

  3 in total