Evaluation of a dental unit designed to prevent retraction of oral fluids.

J J Crawford, C Broderius.   

Abstract

Most older units that drive high-speed dental handpieces include a retraction device designed to prevent dripping of fluids from the water-spray line after the unit has been turned off. Oral fluids and debris may be drawn, along with water, into the water line. Bacteria and debris, unless deliberately flushed out of the water line, may be transmitted to the next patient. Most new units now have mechanisms to prevent retraction. A new dental control unit, designed to eliminate mechanical aspiration of oral fluids, was tested. A kit to test the level of retraction in the water-spray line was also examined. The new unit was found to be effective in protecting the unit water from contamination by oral microorganisms, thereby protecting patients from cross-contamination. However, the control of retraction did not prevent the colonization of the water line by aquatic bacteria. Both internal and external sources of microorganisms must be controlled.

Entities:  

Mesh:

Year:  1990        PMID: 2197655

Source DB:  PubMed          Journal:  Quintessence Int        ISSN: 0033-6572            Impact factor:   1.677


  2 in total

1.  Cross-infection risks associated with current procedures for using high-speed dental handpieces.

Authors:  D L Lewis; R K Boe
Journal:  J Clin Microbiol       Date:  1992-02       Impact factor: 5.948

2.  Efficacy of a Low Dose of Hydrogen Peroxide (Peroxy Ag⁺) for Continuous Treatment of Dental Unit Water Lines: Challenge Test with Legionella pneumophila Serogroup 1 in a Simulated Dental Unit Waterline.

Authors:  Savina Ditommaso; Monica Giacomuzzi; Elisa Ricciardi; Carla M Zotti
Journal:  Int J Environ Res Public Health       Date:  2016-07-22       Impact factor: 3.390

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.