BACKGROUND: Reprocessing of endoscopes generally requires labour-intensive manual cleaning followed by high-level disinfection in an automated endoscope reprocessor (AER). EVOTECH Endoscope Cleaner and Reprocessor (ECR) is approved for fully automated cleaning and disinfection whereas AERs require manual cleaning prior to the high-level disinfection procedure. The purpose of this economic evaluation was to determine the cost-efficiency of the ECR versus AER methods of endoscopy reprocessing in an actual practice setting.Entities:
Mesh:
Year: 2011 PMID: 21967345 PMCID: PMC3198958 DOI: 10.1186/1471-230X-11-105
Source DB: PubMed Journal: BMC Gastroenterol ISSN: 1471-230X Impact factor: 3.067
Steps observed in the reprocessing of endoscopes
| EVOTECH ECR | Manual cleaning plus AER* | |
|---|---|---|
| Fill sink with water | x | x |
| Clean valves | x | x |
| Manual wet leak test** | x | x |
| Add enzymatic detergent to sink | x | |
| Brush internal channels three times each | x | |
| Connect cleaning adapters/tubing to scope | x | |
| Flush interior of scope with enzymatic detergent using hospital-developed automated flushing system | x | |
| Drain sink of enzymatic detergent | x | |
| Rinse interior of scope with tap water using hospital-developed automated flushing system | x | |
| Rinse exterior of scope with tap water spray | x | |
| Drain sink and remove scope from sink | x | x |
| Dry exterior of scope with a blue wipe | x | |
| Walk from sink to reprocessor with scope | x | x |
| Place scope in an unoccupied basin of reprocessor | x | x |
| Connect colour coded connecting tubing to scope according to specific scope connection diagram | x | |
| Enter identification data for operator, physician, scope type, procedure and patient*** | x | |
| Connect reprocessing adapter/tubing to connection port in basin | x | |
| Place lid closure on reprocessor | x | x |
| Program and start cycle | x | x |
| Enter operator identification | x | x |
| Open reprocessor lid | x | x |
| Disconnect color coded tubing from scope | x | |
| Disconnect reprocessing adapter/tubing from connection port in basin with manual check to ensure connections were properly placed | x | |
| Return accessories and scope to Cleanascope® transport tray for next procedure or to the storage cupboard if the scope is not required in the short term | x | x |
*Included all steps involved in preparing scope for reprocessing and then starting the reprocessing cycle.
**The EVOTECH ECR performs an automated wet and dry leak test. A manual leak test according to the endoscope manufacturer's instructions is also recommended. Thus, although it may not be required with EVOTECH ECR, a manual wet leak test was incorporated in the analysis.
***Although the Medivators DSD-201 was capable of recording the same information, this procedure was not in place at the study site.
#Included all steps required to remove scope from reprocessor and replace scope in storage.
Mean total reprocessing time and time savings with the EVOTECH ECR
| Colonoscope | Gastroscope | Bronchoscope | |
|---|---|---|---|
| Number of scopes | 21 | 9 | 4 |
| Median time (mins) per scope | 37.42 | 37.22 | 36.04 |
| [inter-quartile range] | [36.72, 37.63] | [36.65, 37.38] | [35.69, 36.58] |
| Number of scopes | 23 | 16 | 8 |
| Median time (mins) per scope | 49.88 | 44.27 | 42.43 |
| [inter-quartile range] | [49.42, 51.12] | [42.71, 45.37] | [41.70, 43.81] |
| p-value* | < 0.0001 | < 0.0001 | 0.0040 |
*Wilcoxon rank sum test
Labour only time involved in reprocessing with each reprocessor type
| Colonoscope | Gastroscope | Bronchoscope | |
|---|---|---|---|
| Median time (mins) per scope | 7.27 | 7.07 | 5.89 |
| [interquartile range] | [6.57, 7.48] | [6.50, 7.23] | [5.54, 6.44] |
| Median time (mins) per scope | 9.88 | 9.27 | 7.43 |
| [interquartile range] | [9.42, 11.12] | [7.71, 10.37] | [6.70, 8.81] |
| p-value* | < 0.0001 | 0.00002 | 0.0485 |
*Wilcoxon rank sum test
Average daily total reprocessing and personnel labour times
| Colonoscopes | Gastroscopes | Bronchoscopes | |
|---|---|---|---|
| Typical average number of scopes per day | 22 | 14 | 4 |
| Average time per scope (mins) | 38.22 | 37.92 | 36.99 |
| Total daily reprocessing time (mins) | 839 | 518 | 151 |
| Average time per scope (mins) | 50.29 | 44.24 | 42.64 |
| Total daily reprocessing time (mins) | 1104 | 604 | 174 |
| Savings in total reprocessing time with EVOTECH ECR (mins) | 265 | 86 | 23 |
| Total daily time savings for reprocessing all scopes | |||
| Average time per scope (mins) | 7.22 | 6.92 | 5.99 |
| Total daily labour time (mins) | 159 | 95 | 24 |
| Average time per scope (mins) | 10.29 | 9.24 | 7.64 |
| Total daily labour time (mins) | 226 | 126 | 31 |
| Savings in total reprocessing time with EVOTECH ECR (mins) | |||
| Total daily time savings for reprocessing all scopes | |||
Cost of consumables involved in reprocessing endoscopes with manual cleaning plus AER
| Cost per Scope | Cost per Year | |
|---|---|---|
| Multiuse brushes | $1.11 (broncho) | $15, 860.21* |
| Sterilization of multiuse brushes | $0.14 | $1, 443.00 |
| Test strips | $1.21 | $12, 465.17 |
| Fibertech Enzymatic soap (4L) | $0.70 | $7, 207.53 |
| Cidex OPA** in AER for automated endoscope cleaning | $4.15 | $42, 822.86 |
| Labour to change Cidex after every 45 cycles | $0.09 | $933.02 |
| 70% alcohol for alcohol flush | $0.006 | $62.05 |
| Cidex OPA in ultrasonic device to clean multiuse brushes | $0.03 | $284.58 |
| Blue wipes to dry endoscopes after manual cleaning | $0.29 | $3, 037.03 |
| Blue wraps to cover brushes to be autoclaved at the end of each day | $0.004 | $38.27 |
| Small filter (6 per year) | $0.009 | $90.00 |
| Ultra filter replacement every 2 years | $0.04 | $425.00 |
| Ultra filter cleaning 3 times per year - labour | $0.03 | $360.00 |
| Ultra filter cleaning 3 times per year - Cidex | $0.07 | $700.50 |
*Annual cost of disposable brushes would be much higher at $82, 528.
**A less costly disinfectant such as glutaraldehyde may be used; if glutaraldehyde was used, the disinfectant cost would be lower.
Cost of consumables involved in reprocessing endoscopes in EVOTECH ECR
| Cost per Scope | Cost per Year | |
|---|---|---|
| Cidex OPA for automated, single-use cleaning of each endoscope | $7.75 | $79, 912.68 |
| Labour to change the Cidex | $0.02 | $168.10 |
| 70% alcohol for internal flush of each endoscope | $0.03 | $310.27 |
| Cidezyme for automated daily disinfection of the EVOTECH ECR | $0.89 | $9, 174.31 |
| Labour to perform auto disinfection cycle | $0.003 | $28.86 |
| 1 μm External Pre Filter | $0.09 | $225.00 |
| 0.2 μm Internal Evotech Machine Filter | $0.33 | $825.00 |
| 0.2 μm External Hybrid Carbon Filter | $0.16 | $400.00 |
| 0.2 μm External Membrane Filter | $0.32 | $800.00 |
| Labour to change filters | $0.02 | $49.95 |
Total per cycle and annual costs of labour plus consumables
| EVOTECH ECR | Manual cleaning plus AER | |
|---|---|---|
| Annual cost of consumables | $91, 894.17 | $85, 729.23 |
| Annual cost of labour* | $26, 695.03 | $36, 854.22 |
| Total annual cost** | $118, 589.20 | $122, 583.45 |
*Does not include personnel labour time elapsed during cycle processing as employees could be available to work on other tasks while a scope is in the reprocessor.
**Does not include acquisition cost of either processor