Literature DB >> 21952598

Reliability of the HemoCue® hemoglobinometer in critically ill patients: a prospective observational study.

O Mimoz1, D Frasca, A Médard, L Soubiron, B Debaene, C Dahyot-Fizelier.   

Abstract

BACKGROUND: The aim of the study was to assess the reliability of hemoglobin measurements made with HemoCue®, compared with those made with the reference method in critically ill patients.
METHODS: A prospective observational study was conducted in three adult surgical intensive care units of a university hospital. One hundred and ninety-eight consecutive patients were included, and a total of 1166 hemoglobin concentrations were measured using arterial blood samples in the laboratory (HbLAB) and at bedside (HbHC.art) with a portable hemoglobinometer (HemoCue® Hb201+). Simultaneously, a capillary measurement (HbHC.cap) was performed at bedside using the same device.
RESULTS: The mean difference (bias) between HbHC.cap and HbLAB was 0.2 g/dL (95%CI, 0.1;0.3), and limits of agreement were -1.3 g/dL (95%CI, -1.4;-1.2) to 1.7 g/dL (95%CI, 1.6;1.9). The discrepancies between HbHC.cap and HbLAB were greater than 1 g/dL in 30.8% of cases. The bias between HbHC.art and HbLAB was -0.1 g/dL (95% CI, -0.2;0.2), and limits of agreement were slightly better at -1.1 g/dL (95% CI, -1.2;-1.0) and 1.0 g/dL (95% CI, 0.9;1.1). The HemoCue®'s accuracy was not affected by the hospital unit, the puncture site (finger or ear), norepinephrine administration or by hemoglobin levels below 10 g/dL or 8 g/dL.
CONCLUSION: Capillary HemoCue® is not sufficiently accurate to make a therapeutic decision such as whether a blood transfusion should be performed. The method's performance was moderately improved by the use of arterial blood.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21952598

Source DB:  PubMed          Journal:  Minerva Anestesiol        ISSN: 0375-9393            Impact factor:   3.051


  10 in total

1.  Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.

Authors:  Jan Zatloukal; Jiri Pouska; Jakub Kletecka; Richard Pradl; Jan Benes
Journal:  J Clin Monit Comput       Date:  2015-10-27       Impact factor: 2.502

Review 2.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

Review 3.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

Review 4.  Diagnosis of anaemia: old things rearranged.

Authors:  Gabriele Halwachs-Baumann
Journal:  Wien Med Wochenschr       Date:  2012-11-08

5.  Acute iatrogenic polycythemia induced by massive red blood cell transfusion during subtotal abdominal colectomy.

Authors:  David Chiapaikeo; Pejman Rohani
Journal:  Hematol Rep       Date:  2015-03-23

6.  Prevalence of Portable Point of Care Tests Used on Medical Service Trips in Latin America and the Caribbean.

Authors:  Christopher Dainton; Nikki Shah; Charlene H Chu
Journal:  Ann Glob Health       Date:  2018-11-05       Impact factor: 2.462

7.  Targeted provision of oral iron: the evolution of a practical screening option.

Authors:  Caitlin R Crowley; Noel W Solomons; Klaus Schümann
Journal:  Adv Nutr       Date:  2012-07-01       Impact factor: 8.701

8.  The impact of an algorithm-guided management of preoperative anemia in perioperative hemoglobin level and transfusion of major orthopedic surgery patients.

Authors:  Dietmar Enko; Franz Wallner; Achim von-Goedecke; Christa Hirschmugl; Vinzenz Auersperg; Gabriele Halwachs-Baumann
Journal:  Anemia       Date:  2013-03-27

9.  Accuracy of Capillary Hemoglobin Measurements for the Detection of Anemia among U.S. Low-Income Toddlers and Pregnant Women.

Authors:  Safia Boghani; Zuguo Mei; Geraldine S Perry; Gary M Brittenham; Mary E Cogswell
Journal:  Nutrients       Date:  2017-03-09       Impact factor: 5.717

10.  Comparison of haemoglobin assessments by HemoCue and two automated haematology analysers in young Laotian children.

Authors:  Guy-Marino Hinnouho; Maxwell A Barffour; K Ryan Wessells; Kenneth H Brown; Sengchanh Kounnavong; Bigphone Chanhthavong; Kethmany Ratsavong; Chidchamai Kewcharoenwong; Sonja Y Hess
Journal:  J Clin Pathol       Date:  2017-12-02       Impact factor: 3.411

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.