PURPOSE: To evaluate the relative safety and potential efficacy of high-intensity focused ultrasound cyclocoagulation by a miniaturized annular device containing six piezoceramic transducers in patients with refractory glaucoma. METHODS: This was a three-center prospective interventional pilot study. Twelve eyes of 12 patients with refractory glaucoma were insonified using a ring-shaped probe containing six miniaturized high-frequency transducers operating at 21 MHz. Ultrasound biomicroscopy (UBM) and a complete ophthalmic examination were performed before the procedure and at 1 day, 1 week, 1 month, and 3 months after the procedure. Additional visits were performed 6 and 12 months after the procedure. RESULTS: Intraocular pressure was significantly reduced (P < 0.01) from a mean preoperative value of 37.9 ± 10.7 mm Hg to a mean postoperative value of 27.3 ± 12.4, 25.2 ± 11.3, 25.2 ± 7.7, 24.8 ± 9.8, and 26.3 ± 5.1 mm Hg at 1 day, 1 week, 1 month, 3 months, and 6 months, respectively, and to a mean value of 24.7 ± 8.5 at the last follow-up visit. No major intraoperative or postoperative complications occurred. Minor postoperative corneal complications developed in four patients with previous corneal abnormalities: superficial punctate keratitis (n = 3) and central superficial corneal ulceration (n = 1). UBM showed cystic involution of the ciliary body in 9 of the 12 eyes and a suprachoroidal fluid space in 8 of the 12 eyes. CONCLUSIONS: Ultrasonic circular cyclocoagulation using high-intensity focused ultrasound delivered by a circular miniaturized device containing six piezoceramic transducers seems to be an effective and well-tolerated method to reduce intraocular pressure in patients with refractory glaucoma.
PURPOSE: To evaluate the relative safety and potential efficacy of high-intensity focused ultrasound cyclocoagulation by a miniaturized annular device containing six piezoceramic transducers in patients with refractory glaucoma. METHODS: This was a three-center prospective interventional pilot study. Twelve eyes of 12 patients with refractory glaucoma were insonified using a ring-shaped probe containing six miniaturized high-frequency transducers operating at 21 MHz. Ultrasound biomicroscopy (UBM) and a complete ophthalmic examination were performed before the procedure and at 1 day, 1 week, 1 month, and 3 months after the procedure. Additional visits were performed 6 and 12 months after the procedure. RESULTS: Intraocular pressure was significantly reduced (P < 0.01) from a mean preoperative value of 37.9 ± 10.7 mm Hg to a mean postoperative value of 27.3 ± 12.4, 25.2 ± 11.3, 25.2 ± 7.7, 24.8 ± 9.8, and 26.3 ± 5.1 mm Hg at 1 day, 1 week, 1 month, 3 months, and 6 months, respectively, and to a mean value of 24.7 ± 8.5 at the last follow-up visit. No major intraoperative or postoperative complications occurred. Minor postoperative corneal complications developed in four patients with previous corneal abnormalities: superficial punctate keratitis (n = 3) and central superficial corneal ulceration (n = 1). UBM showed cystic involution of the ciliary body in 9 of the 12 eyes and a suprachoroidal fluid space in 8 of the 12 eyes. CONCLUSIONS: Ultrasonic circular cyclocoagulation using high-intensity focused ultrasound delivered by a circular miniaturized device containing six piezoceramic transducers seems to be an effective and well-tolerated method to reduce intraocular pressure in patients with refractory glaucoma.
Authors: Manas R Gartia; Zhe Wu; Arun K De; Mao Ye; Santosh K Misra; Goutam Ghoshal; Corinne R Bromfield; Emery M Williams; Kuldeep Singh; Krishnarao V Tangella; Laurie Rund; Klaus Schulten; Lawrence B Schook; Partha S Ray; Everette C Burdette; Dipanjan Pan Journal: ACS Nano Date: 2015-10-12 Impact factor: 15.881
Authors: Giuseppe Giannaccare; A Vagge; C Gizzi; A Bagnis; S Sebastiani; C Del Noce; M Fresina; C E Traverso; E C Campos Journal: Graefes Arch Clin Exp Ophthalmol Date: 2016-12-03 Impact factor: 3.117