Literature DB >> 21947650

Surgical removal of infected pacemaker leads without cardiopulmonary bypass after failed extraction using the Excimer Laser Sheath Extraction System.

Chiho Tokunaga1, Yoshiharu Enomoto, Fujio Sato, Shinya Kanemoto, Shonosuke Matsushita, Yuji Hiramatsu, Kazutaka Aonuma, Yuzuru Sakakibara.   

Abstract

With the growing number of cardiac pacemakers and internal cardioverter defibrillator implantations, problems with endocardial lead infection have been increasing. The newly developed Excimer Laser Sheath Lead Extraction System has been recognized as being highly useful for removing chronic infected leads. However, serious bleeding complications are a concern when this system is used. Here we report our experience with a 67-year-old man who was diagnosed with pacemaker endocarditis. Initially, lead removal was attempted using the Excimer Laser Sheath Extraction System, though this was abandoned because of severe adhesion of the leads and the junction of the supra vena cava (SVC) with the right atrium. Surgical removal of the leads was performed without using cardiopulmonary bypass and the leads were removed without any complications. During surgery, we found there was a silent perforation of the innominate vein brought about by the Excimer Laser Sheath System. Also, the junction of the SVC with the right atrium was thought to be an area potentially at high risk of perforation, because of a lack of surrounding tissue. It is our opinion that those who carry out procedures with the Excimer Laser Sheath System should understand the potential risk of perforation based on cardiac anatomy and should be prepared for lethal bleeding complications. Also, for emergent situations, we believe that close backup by a cardiovascular surgical team should be considered essential for performing the Excimer Laser Sheath Lead Extraction safely.

Entities:  

Mesh:

Year:  2011        PMID: 21947650     DOI: 10.1007/s10047-011-0610-8

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  6 in total

Review 1.  Recommendations for extraction of chronically implanted transvenous pacing and defibrillator leads: indications, facilities, training. North American Society of Pacing and Electrophysiology Lead Extraction Conference Faculty.

Authors:  C J Love; B L Wilkoff; C L Byrd; P H Belott; J A Brinker; N E Fearnot; R A Friedman; S Furman; L B Goode; D L Hayes; D T Kawanishi; V Parsonnet; C Reiser; H J Van Zandt
Journal:  Pacing Clin Electrophysiol       Date:  2000-04       Impact factor: 1.976

2.  Pacemaker and internal cardioverter defibrillator lead extraction: a safe and effective surgical approach.

Authors:  John M Kratz; John M Toole
Journal:  Ann Thorac Surg       Date:  2010-11       Impact factor: 4.330

Review 3.  Intravascular lead extraction using the excimer laser: pitfalls and tips for success.

Authors:  Robert W Feldtman
Journal:  Semin Vasc Surg       Date:  2008-03       Impact factor: 1.000

Review 4.  Update on laser lead extraction.

Authors:  Jeffrey S Snow; Sameer Parekh
Journal:  J Invasive Cardiol       Date:  2009-05       Impact factor: 2.022

5.  Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA).

Authors:  Bruce L Wilkoff; Charles J Love; Charles L Byrd; Maria Grazia Bongiorni; Roger G Carrillo; George H Crossley; Laurence M Epstein; Richard A Friedman; Charles E H Kennergren; Przemyslaw Mitkowski; Raymond H M Schaerf; Oussama M Wazni
Journal:  Heart Rhythm       Date:  2009-05-22       Impact factor: 6.343

Review 6.  Extraction of transvenous pacing and ICD leads.

Authors:  Macy C Smith; Charles J Love
Journal:  Pacing Clin Electrophysiol       Date:  2008-06       Impact factor: 1.976

  6 in total
  1 in total

Review 1.  Journal of Artificial Organs 2012: the year in review.

Authors:  Y Sawa; E Tatsumi; T Tsukiya; K Matsuda; K Fukunaga; A Kishida; T Masuzawa; G Matsumiya; A Myoui; M Nishimura; T Nishimura; T Nishinaka; E Okamoto; S Tokunaga; T Tomo; Y Yagi; T Yamaoka
Journal:  J Artif Organs       Date:  2013-02-28       Impact factor: 1.731

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.