Randomized controlled trial of the A.P. Advance, McGrath, and Macintosh laryngoscopes in normal and difficult intubation scenarios: a manikin study.

E Burdett1, D J Ross-Anderson, J Makepeace, P A Bassett, S G Clarke, V Mitchell.   

Abstract

BACKGROUND: Several videolaryngoscopes are available which may facilitate tracheal intubation in difficult airways. We compared the McGrath(®) Series 5 and the Venner™ A.P. Advance™ (APA) videolaryngoscopes with a Macintosh laryngoscope by studying the performance of experienced anaesthetists using manikins in normal and difficult airway scenarios.
METHODS: We recruited 48 anaesthetists into a randomized trial. Each performed tracheal intubation with each laryngoscope in one easy and one difficult laryngoscopy scenario. The primary endpoint was time to intubation. Other endpoints were time to best glottic visualization, grade of view, and number of glottic advances.
RESULTS: There were no dropouts. In the easy scenario, the time to intubation was greater using the McGrath [median time 40.7 s, inter-quartile range (IQR) 31.0, 57.4, P<0.001] than the other devices. In the difficult scenario, the time to intubation using the APA with Difficult Airway Blade (DAB) was less (median time 23.2 s, IQR 19.8, 29.0, P<0.001) than the other devices. Time to glottic visualization was reduced using the McGrath and the APA with DAB. Glottic advances were fewer using the APA with DAB.
CONCLUSIONS: Experienced anaesthetists required a longer time for intubation in a standard manikin using a McGrath compared with other laryngoscopes, but a shorter time for intubation in a difficult manikin using an APA with DAB, and with fewer glottic advances, compared with other laryngoscopes.

Entities:  

Mesh:

Year:  2011        PMID: 21940397     DOI: 10.1093/bja/aer295

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  13 in total

1.  Comparison of Macintosh, McCoy and C-MAC D-Blade video laryngoscope intubation by prehospital emergency health workers: a simulation study.

Authors:  Ahmet Yildirim; Hasan A Kiraz; İbrahim Ağaoğlu; Okhan Akdur
Journal:  Intern Emerg Med       Date:  2016-03-21       Impact factor: 3.397

2.  Does Video Laryngoscopy Offer Advantages over Direct Laryngoscopy during Cardiopulmonary Resuscitation?

Authors:  Ayten Saraçoğlu; Olgaç Bezen; Türker Şengül; Egin Hüsnü Uğur; Sibel Şener; Fisun Yüzer
Journal:  Turk J Anaesthesiol Reanim       Date:  2015-03-03

3.  Comparison of the C-MAC D-Blade, Conventional C-MAC, and Macintosh Laryngoscopes in Simulated Easy and Difficult Airways.

Authors:  Alper Kılıçaslan; Ahmet Topal; Atilla Erol; Sema Tuncer Uzun
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-08-01

4.  Tracheal intubation in patients with cervical spine immobilization: A comparison of McGrath(®) video laryngoscope and Truview EVO2(®) laryngoscope.

Authors:  Ruchi Bhola; Swaran Bhalla; Radha Gupta; Ishwar Singh; Sunil Kumar
Journal:  Indian J Anaesth       Date:  2014-05

5.  Comparison of the McGrath MAC video laryngoscope and the Pentax Airwayscope during chest compression: a manikin study.

Authors:  Atsushi Kotera; Hiroki Irie; Shinsuke Iwashita; Junichi Taniguchi; Shunji Kasaoka; Yoshihiro Kinoshita
Journal:  J Intensive Care       Date:  2014-03-06

6.  Comparing McGRATH® MAC, C-MAC®, and Macintosh Laryngoscopes Operated by Medical Students: A Randomized, Crossover, Manikin Study.

Authors:  Myungju Shin; Sun Joon Bai; Ki-Young Lee; Ein Oh; Hyun Joo Kim
Journal:  Biomed Res Int       Date:  2016-09-15       Impact factor: 3.411

7.  Comparison of McGrath(®) Series 5 video laryngoscope with Macintosh laryngoscope: A prospective, randomised trial in patients with normal airways.

Authors:  Mehmet Sargin; Mehmet Selcuk Uluer
Journal:  Pak J Med Sci       Date:  2016 Jul-Aug       Impact factor: 1.088

8.  A prospective, randomised, clinical study to compare the use of McGrath(®), Truview(®) and Macintosh laryngoscopes for endotracheal intubation by novice and experienced Anaesthesiologists.

Authors:  Sumitra G Bakshi; Vinayak S Vanjari; Jigeeshu V Divatia
Journal:  Indian J Anaesth       Date:  2015-07

9.  Comparison of the Glidescope®, flexible fibreoptic intubating bronchoscope, iPhone modified bronchoscope, and the Macintosh laryngoscope in normal and difficult airways: a manikin study.

Authors:  Adrian Langley; Gabriel Mar Fan
Journal:  BMC Anesthesiol       Date:  2014-02-28       Impact factor: 2.217

10.  Comparison of McGrath Series 3 and Macintosh Laryngoscopes for Tracheal Intubation in Patients With Normal Airway by Inexperienced Anesthetists: A Randomized Study.

Authors:  Zi-Jia Liu; Jie Yi; Wen-Juan Guo; Chao Ma; Yu-Guang Huang
Journal:  Medicine (Baltimore)       Date:  2016-01       Impact factor: 1.817