PURPOSE: Distal rectal stapling is often challenging because of limited space and visibility. We compared two stapling devices in the distal rectum in a cadaver study: the iDrive™ right angle linear cutter (RALC) (Covidien, New Haven, CT) and the CONTOUR® curved cutter (CC) (Ethicon Endo-Surgery, Cincinnati, OH).Entities:
Year: 2011 PMID: 21871120 PMCID: PMC3189175 DOI: 10.1186/1750-1164-5-7
Source DB: PubMed Journal: Ann Surg Innov Res ISSN: 1750-1164
Figure 1Illustration of coronal placement of the RALC.
Figure 2Illustration of sagittal placement of the RALC.
Figure 3Distal placement of the RALC on the rectum.
Distance of stapler from pelvic floor
| Stapler | Stapler used | Distance (cm) from PF median (range) | Distance (cm) from PF mean (std dev) | Mean Effect | P-value |
|---|---|---|---|---|---|
| Coronal | RALC | 1.0 (0-4.0) | 1.4 (1.3) | -0.54 | 0.003 |
| Sagittal | RALC | 1.6 (0-3.5) | 1.5 (1.2) | -1.70 | < 0.0001 |
PF = pelvic floor
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Covariates included investigator and cadaver indicators.
How many times did you have to readjust the stapler?
| Stapler | Stapler used | 0 | 1 | 2 | 3 or more | Odds Ratio (for fewer times) | P-value |
|---|---|---|---|---|---|---|---|
| Coronal | RALC | 18 | 23 | 4 | 2 | 2.75 | 0.01 |
| Sagittal | RALC | 19 | 21 | 6 | 1 | 3.53 | 0.002 |
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Stapler Performance: Yes/No questions, Coronal placement
| Question | Stapler used | YES | NO | Odds Ratio (for yes) | P Value |
|---|---|---|---|---|---|
| Was there interference of the pubic symphysis limiting placement of the device? | RALC | 1 (2%) | 47 | 0.054 | 0.006 |
| Was there readjusting of the instrument for optimal placement? | RALC | 30 (63%) | 18 | 0.47 | 0.11 |
| Did the instrument impede visibility | RALC | 1(2%) | 47 | 0.02 | 0.0003 |
| Was the whole rectum contained in the device after clamping | RALC | 45 (94%) | 3 | 1.00 | 0.96 |
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Stapler Performance: Rating questions, Coronal placement
| Rating | Stapler used | Excellent | Adequate | Poor | Odds Ratio (for poorer results) | P Value |
|---|---|---|---|---|---|---|
| Ease of placement in the pelvis | RALC | 39 (81%) | 9 (19%) | 0 | 0.18 | 0.0007 |
| Ease of placement around the colon and rectum | RALC | 36 (75%) | 11 (23%) | 1 (2%) | 0.43 | 0.08 |
| Ability to hold and retain tissue without slippage | RALC | 44 (94%) | 3 (6%) | 0 | 1.49 | 0.67 |
| Visibility | RALC | 38 (79%) | 10 (21%) | 0 | 0.39 | 0.46 |
| Pelvic access | RALC | 38 (79%) | 10 (21%) | 0 | 0.14 | < .0001 |
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Stapler Performance: Yes/No questions, Sagittal placement
| Question | Stapler used | YES | NO | Odds Ratio (for yes) | P Value |
|---|---|---|---|---|---|
| Was there interference of the pubic symphysis limiting placement of the device? | RALC | 2 (4%) | 46 | 0.064 | 0.0006 |
| Was there readjusting of the instrument for optimal placement? | RALC | 29 (60%) | 19 | 0.43 | 0.07 |
| Did the instrument impede visibility | RALC | 3 (6%) | 45 | 0.32 | 0.13 |
| Was the whole rectum contained in the device after clamping | RALC | 45 (94%) | 3 | 18.08 | 0.007 |
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Stapler Performance: Rating questions, Sagittal placement
| Rating | Stapler used | Excellent | Adequate | Poor | Odds Ratio (for poorer results) | P Value |
|---|---|---|---|---|---|---|
| Ease of placement in the pelvis | RALC | 38 (79%) | 10 (21%) | 0 | 0.08 | < .0001 |
| Ease of placement around the colon and rectum | RALC | 36 (75%) | 11 (23%) | 1 (2%) | 0.29 | 0.007 |
| Ability to hold and retain tissue without slippage | RALC | 46 (98%) | 0 | 1 (2%) | 0.04 | 0.002 |
| Visibility | RALC | 38 (79%) | 10 (21%) | 0 | 0.16 | 0.0001 |
| Pelvic access | RALC | 37 (77%) | 11 (23%) | 0 | 0.10 | < .0001 |
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Preference questions: Coronal and sagittal placement
| Question | Stapler used | CORONAL | SAGITTAL |
|---|---|---|---|
| Visually which is best? | RALC | 32 (67%) | 16 (33%) |
| What is the best placement of the device? | RALC | 35 (73%) | 13 (27%) |
RALC = Right Angle Linear Cutter
CC = Curved Cutter
Measurement of anatomic distances
| Question | Median* (range) |
|---|---|
| Umbilicus to pubic symphysis | 15 (11-18) |
| Anterior superior iliac spine (ASIS) to ASIS | 24 (20-33) |
| Pubic symphysis to sacral promontory | 12.25 (8-13.5) |
| Pelvic floor to sacral promontory | 12.5 (9-18) |
| Tip of coccyx to pubis symphysis | 11.8 (8-15) |
| Pelvic inlet** | 9.0 (7-13) |
| Anal verge to pelvic floor | 6 (5-10) |
Measurements are in cm
*12 cadavers
**Pelvic inlet measured as transverse diameter between right and left pelvic sidewall
Figure 4Mean distance from the pelvic floor to the device using coronal placement.
Figure 5Mean distance from the pelvic floor to the device using sagittal placement.