Literature DB >> 19682838

Detection of locking bolt loosening in the stem-condyle junction of a modular femoral stem in revision total knee arthroplasty.

Jae-Min Ahn1, Jeung-Tak Suh.   

Abstract

We report one case of loosening of the locking bolt in the stem-condyle junction of a constrained modular femoral component in revision total knee arthroplasty. Early detection of loosening of the locking bolt was possible using reconstructed view of computerized tomography scans before complete disengagement and radiographic loosening of the stemmed femoral component. When using the Total Condyle III femoral component with a stem extension, surgeons should keep in mind that stress concentration at the stem-condyle junction can result in failure of the locking bolt although it is rare and that loosening or disengagement of the locking bolt may be a sign of a loose stemmed femoral component. Copyright 2010 Elsevier Inc. All rights reserved.

Mesh:

Year:  2009        PMID: 19682838     DOI: 10.1016/j.arth.2009.06.014

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  6 in total

1.  Fracture at the stem-condylar junction of a modular femoral prosthesis in a varus-valgus constrained total knee arthroplasty.

Authors:  D D Nikolopoulos; I G Polyzois; E A Magnissalis; P F Bernard; I V Michos
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-08-25       Impact factor: 4.342

2.  Ten-Year Results of Primary and Revision Condylar-Constrained Total Knee Arthroplasty in Patients with Severe Coronal Plane Instability.

Authors:  Andrea Camera; Stefano Biggi; Gabriele Cattaneo; Giovanni Brusaferri
Journal:  Open Orthop J       Date:  2015-08-31

3.  Fracture of the femoral adapter bolt and taper adapter in a modern rotating platform knee arthroplasty.

Authors:  Elexis C Baral; Alexander S McLawhorn; Timothy M Wright; Edwin P Su
Journal:  Arthroplast Today       Date:  2017-03-03

4.  Trunnion Failure in Revision Total Knee Arthroplasty.

Authors:  Samir Chihab; Bryan M Grommersch; Greg A Erens; George N Guild
Journal:  Arthroplast Today       Date:  2021-02-26

5.  Clinical Outcomes of Tibial Components with Modular Stems Used in Primary TKA.

Authors:  Nicole Durig; Thomas Pace; Brandon Broome; Obi Osuji; Melinda K Harman
Journal:  Adv Orthop       Date:  2014-02-05

6.  A Novel RFID-Based Sensing Method for Low-Cost Bolt Loosening Monitoring.

Authors:  Jian Wu; Xingmei Cui; Yunpeng Xu
Journal:  Sensors (Basel)       Date:  2016-01-28       Impact factor: 3.576

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.