Quantitative analysis of beam delivery parameters and treatment process time for proton beam therapy.

Kazumichi Suzuki1, Michael T Gillin, Narayan Sahoo, X Ronald Zhu, Andrew K Lee, Denise Lippy.   

Abstract

PURPOSE: To evaluate patient census, equipment clinical availability, maximum daily treatment capacity, use factor for major beam delivery parameters, and treatment process time for actual treatments delivered by proton therapy systems.
METHODS: The authors have been recording all beam delivery parameters, including delivered dose, energy, range, spread-out Bragg peak widths, gantry angles, and couch angles for every treatment field in an electronic medical record system. We analyzed delivery system downtimes that had been recorded for every equipment failure and associated incidents. These data were used to evaluate the use factor of beam delivery parameters, the size of the patient census, and the equipment clinical availability of the facility. The duration of each treatment session from patient walk-in and to patient walk-out of the treatment room was measured for 82 patients with cancers at various sites.
RESULTS: The yearly average equipment clinical availability in the last 3 yrs (June 2007-August 2010) was 97%, which exceeded the target of 95%. Approximately 2200 patients had been treated as of August 2010. The major disease sites were genitourinary (49%), thoracic (25%), central nervous system (22%), and gastrointestinal (2%). Beams have been delivered in approximately 8300 treatment fields. The use factor for six beam delivery parameters was also evaluated. Analysis of the treatment process times indicated that approximately 80% of this time was spent for patient and equipment setup. The other 20% was spent waiting for beam delivery and beam on. The total treatment process time can be expressed by a quadratic polynomial of the number of fields per session. The maximum daily treatment capacity of our facility using the current treatment processes was estimated to be 133 +/- 35 patients.
CONCLUSIONS: This analysis shows that the facility has operated at a high performance level and has treated a large number of patients with a variety of diseases. The use factor of beam delivery parameters varies by disease site. Further improvements in efficiency may be realized in the equipment- and patient-related processes of treatment.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21859034     DOI: 10.1118/1.3604153

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Modelling the throughput capacity of a single-accelerator multitreatment room proton therapy centre.

Authors:  A H Aitkenhead; D Bugg; C G Rowbottom; E Smith; R I Mackay
Journal:  Br J Radiol       Date:  2012-12       Impact factor: 3.039

2.  Patient handling system for carbon ion beam scanning therapy.

Authors:  Shinichiro Mori; Toshiyuki Shirai; Yuka Takei; Takuji Furukawa; Taku Inaniwa; Yuka Matsuzaki; Motoki Kumagai; Takeshi Murakami; Koji Noda
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

3.  Impact of proton beam availability on patient treatment schedule in radiation oncology.

Authors:  Eric D Miller; Vladimir Derenchuk; Indra J Das; Peter A S Johnstone
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

4.  Conventionally fractionated large volume head and neck re-irradiation using multileaf collimator-based robotic technique: A feasibility study.

Authors:  Houda Bahig; Catherine Wang; Sweet Ping Ng; Jack Phan
Journal:  Clin Transl Radiat Oncol       Date:  2020-07-02

5.  Multiple energy extraction reduces beam delivery time for a synchrotron-based proton spot-scanning system.

Authors:  James E Younkin; Martin Bues; Terence T Sio; Wei Liu; Xiaoning Ding; Sameer R Keole; Joshua B Stoker; Jiajian Shen
Journal:  Adv Radiat Oncol       Date:  2018-02-23

Review 6.  Particle Therapy for Breast Cancer.

Authors:  Roman O Kowalchuk; Kimberly S Corbin; Rachel B Jimenez
Journal:  Cancers (Basel)       Date:  2022-02-20       Impact factor: 6.639

7.  Analysis of treatment process time for real-time-image gated-spot-scanning proton-beam therapy (RGPT) system.

Authors:  Takaaki Yoshimura; Shinichi Shimizu; Takayuki Hashimoto; Kentaro Nishioka; Norio Katoh; Tetsuya Inoue; Hiroshi Taguchi; Koichi Yasuda; Taeko Matsuura; Seishin Takao; Masaya Tamura; Yoichi M Ito; Yuto Matsuo; Hiroshi Tamura; Kenji Horita; Kikuo Umegaki; Hiroki Shirato
Journal:  J Appl Clin Med Phys       Date:  2019-12-30       Impact factor: 2.102

8.  Quantitative analysis of treatments using real-time image gated spot-scanning with synchrotron-based proton beam therapy system log data.

Authors:  Takaaki Yoshimura; Shinichi Shimizu; Takayuki Hashimoto; Kentaro Nishioka; Norio Katoh; Hiroshi Taguchi; Koichi Yasuda; Taeko Matsuura; Seishin Takao; Masaya Tamura; Sodai Tanaka; Yoichi M Ito; Yuto Matsuo; Hiroshi Tamura; Kenji Horita; Kikuo Umegaki; Hiroki Shirato
Journal:  J Appl Clin Med Phys       Date:  2020-11-05       Impact factor: 2.102

Review 9.  Future Developments in Charged Particle Therapy: Improving Beam Delivery for Efficiency and Efficacy.

Authors:  Jacinta Yap; Andrea De Franco; Suzie Sheehy
Journal:  Front Oncol       Date:  2021-12-09       Impact factor: 5.738

10.  Quantitative analysis of MRI-guided radiotherapy treatment process time for tumor real-time gating efficiency.

Authors:  Lorenzo Placidi; Davide Cusumano; Luca Boldrini; Claudio Votta; Veronica Pollutri; Marco Valerio Antonelli; Giuditta Chiloiro; Angela Romano; Viola De Luca; Francesco Catucci; Luca Indovina; Vincenzo Valentini
Journal:  J Appl Clin Med Phys       Date:  2020-10-22       Impact factor: 2.243

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.