Literature DB >> 21846935

A practical approach to estimate the weighted CT dose index over an infinite integration length.

Xinhua Li1, Da Zhang, Bob Liu.   

Abstract

The CT scanner-displayed radiation dose information is based on CT dose index (CTDI) over an integration length of 100 mm (CTDI(100)), which is lower than the CTDI over an infinite integration length (CTDI(∞)). In an adult or a pediatric body CT scan, the limiting equilibrium dose can be established near the central scan plane, and CTDI(∞) more closely indicates the accumulated dose than CTDI(100). The aim of this study was to (a) evaluate CTDI efficiencies, ϵ(CTDI(100)) = CTDI(100)/CTDI(∞), for a multi-detector CT (MDCT) scanner, (b) examine the dependences of ϵ(CTDI(100)) on kV, beam width, phantom diameter, phantom length and position in phantom and (c) investigate how to estimate CTDI(∞) based on the CT scanner-displayed information. We performed a comprehensive Geant4-based simulation study of a clinical CT scanner, and calculated ϵ(CTDI(100)) for a range of parameters. The results were compared with the ϵ(CTDI(100)) data of previous studies. Differences in the ϵ(CTDI(100)) values of these studies were assessed. A broad analysis of the ϵ(CTDI(100)) variations with the above-mentioned parameters was presented. Based on the results, we proposed a practical approach to obtain the weighted CTDI(∞) using the CT scanner-displayed information. A reference combination of 120 kV and a beam width close to 20 mm can be selected to determine the efficiencies of the weighted CTDI by using either phantom measurements or computer simulations. The results can be applied to estimate the weighted CTDI(∞) for 80-140 kV and the beam widths within 40 mm. Errors in the weighted CTDI(∞) due to the variations of kV and beam width can be 5% or less for the MDCT scanners.

Mesh:

Year:  2011        PMID: 21846935     DOI: 10.1088/0031-9155/56/18/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

Authors:  Xinhua Li; Jim Q Shi; Da Zhang; Sarabjeet Singh; Atul Padole; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery.

Authors:  Asger Greval Petersen; Søren Eiskjær; Jon Kaspersen
Journal:  Pediatr Radiol       Date:  2012-06-06

3.  Evaluation of cumulative dose for cone-beam computed tomography (CBCT) scans within phantoms made from different compositions using Monte Carlo simulations.

Authors:  Abdullah Abuhaimed; Colin J Martin; Marimuthu Sankaralingam; Kurian Oomen; David J Gentle
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

4.  Development of a method to create uniform phantoms for task-based assessment of CT image quality.

Authors:  Juliane Conzelmann; Felix Benjamin Schwarz; Bernd Hamm; Michael Scheel; Paul Jahnke
Journal:  J Appl Clin Med Phys       Date:  2020-07-28       Impact factor: 2.102

  4 in total